Linac reported steering error insensitive to 6 MV FFF transverse beam position deviations

被引:0
|
作者
McAloney, Jacob [1 ]
Biggs, Simon [2 ]
Sobolewski, Matthew [1 ,3 ]
机构
[1] Riverina Canc Care Ctr, Wagga Wagga, NSW, Australia
[2] Radiotherapy AI, Wagga Wagga, NSW, Australia
[3] Northern Beaches Canc Care, Frenchs Forest, Australia
关键词
Beam steering; FFF; Radiation field alignment; TECHNOLOGY;
D O I
10.1007/s13246-021-01067-2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Clinically significant beam position deviations were observed for a 6 MV FFF beam during patient specific QA on an Elekta linear accelerator. There was no significant reported transverse steering error from the machine ion chamber, and routine linac QA practices, including cardinal angle Winston-Lutz test, did not identify the deviations. Subsequent investigation using an electronic portal imaging device (EPID) revealed clinically significant beam position deviations for small steering errors. This prompted investigation into further impact and possible solutions. Testing set-points were established by adjusting transverse steering current to achieve introduced 2 T steering errors. Tests at each set-point included scanning water tank profiles and EPID images. A proposed method for adjusting the 2 T error sensitivity was tested via adjusting the 2 T loop parameter such that a reported error corresponds to specific beam position deviations. The testing set-points resulted in positional deviations of greater than 3 mm for reported errors of less than 1. A new method for improving 2 T error sensitivity was implemented. This work has shown that existing vendor protocol for establishing beam steering error for 6 MV FFF beams can lead to beam position deviations without machine interlocks or significant reported steering errors. Thus, an alternative method of establishing steering error sensitivity based on positional deviations is presented.
引用
收藏
页码:1377 / 1382
页数:6
相关论文
共 50 条
  • [21] Multicentric characterisation of lateral beam profiles generated by 6FFF beam of three 0.35 T MR-linac systems
    Sebastiano, Menna
    Randa, El Gawhary
    Lorenzo, Placidi
    Flaviovincenzo, Quaranta
    Cristian, Borrazzo
    Marica, Masi
    Matteo, Nardini
    Maria, Rago
    Luca, Indovina
    Davide, Cusumano
    Antonella, Fogliata
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2024, 119
  • [22] Multicentric characterisation of lateral beam profiles from 6 MV FFF beam of three 0.35T MR-linacs
    Fogliata, A.
    El Gawhary, R.
    Cusumano, D.
    Placidi, L.
    Quaranta, F.
    Nardini, M.
    Rago, M.
    Indovina, L.
    Menna, S.
    RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S1464 - S1465
  • [23] Simplified sigmoidal curve fitting for a 6 MV FFF photon beam of the Halcyon to determine the field size for beam commissioning and quality assurance
    Min-Geon Choi
    Martin Law
    Do-Kun Yoon
    Mikoto Tamura
    Kenji Matsumoto
    Masakazu Otsuka
    Moo-Sub Kim
    Shih-Kien Djeng
    Hajime Monzen
    Tae Suk Suh
    Radiation Oncology, 15
  • [24] Simplified sigmoidal curve fitting for a 6 MV FFF photon beam of the Halcyon to determine the field size for beam commissioning and quality assurance
    Choi, Min-Geon
    Law, Martin
    Yoon, Do-Kun
    Tamura, Mikoto
    Matsumoto, Kenji
    Otsuka, Masakazu
    Kim, Moo-Sub
    Djeng, Shih-Kien
    Monzen, Hajime
    Suh, Tae Suk
    RADIATION ONCOLOGY, 2020, 15 (01)
  • [25] Therapeutic dose simulation of a 6 MV Varian Linac photon beam using GEANT4
    Salama, E.
    Ali, A. S.
    Khaled, N. E.
    Radi, A.
    JOURNAL OF INSTRUMENTATION, 2015, 10
  • [26] Evaluation of Tissue Composition Effect On Dose Distribution in Radiotherapy with 6 MV Photon Beam of a Medical Linac
    Ghorbani, M.
    Tabatabaei, Z.
    Noghreiyan, A. Vejdani
    Meigooni, A. Soleimani
    MEDICAL PHYSICS, 2015, 42 (06) : 3428 - 3428
  • [27] Simulation of a 6 MV Elekta Precise Linac photon beam using GATE/GEANT4
    Grevillot, L.
    Frisson, T.
    Maneval, D.
    Zahra, N.
    Badel, J-N
    Sarrut, D.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (04): : 903 - 918
  • [28] On the Use of 6MV-FFF Beam in the Treatment of Hypofractionated Left Breast Patients Using DIBH-CT
    Snyder, J.
    Molloy, J.
    Bernard, M.
    Pokhrel, D.
    MEDICAL PHYSICS, 2019, 46 (06) : E420 - E420
  • [29] Effect of plan complexity on the dosimetry, delivery accuracy, and interplay effect in lung VMAT SBRT with 6 MV FFF beam
    Chao Ge
    Huidong Wang
    Kunzhi Chen
    Wuji Sun
    Huicheng Li
    Yinghua Shi
    Strahlentherapie und Onkologie, 2022, 198 : 744 - 751
  • [30] Effect of plan complexity on the dosimetry, delivery accuracy, and interplay effect in lung VMAT SBRT with 6 MV FFF beam
    Ge, Chao
    Wang, Huidong
    Chen, Kunzhi
    Sun, Wuji
    Li, Huicheng
    Shi, Yinghua
    STRAHLENTHERAPIE UND ONKOLOGIE, 2022, 198 (08) : 744 - 751