Linac reported steering error insensitive to 6 MV FFF transverse beam position deviations

被引:0
|
作者
McAloney, Jacob [1 ]
Biggs, Simon [2 ]
Sobolewski, Matthew [1 ,3 ]
机构
[1] Riverina Canc Care Ctr, Wagga Wagga, NSW, Australia
[2] Radiotherapy AI, Wagga Wagga, NSW, Australia
[3] Northern Beaches Canc Care, Frenchs Forest, Australia
关键词
Beam steering; FFF; Radiation field alignment; TECHNOLOGY;
D O I
10.1007/s13246-021-01067-2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Clinically significant beam position deviations were observed for a 6 MV FFF beam during patient specific QA on an Elekta linear accelerator. There was no significant reported transverse steering error from the machine ion chamber, and routine linac QA practices, including cardinal angle Winston-Lutz test, did not identify the deviations. Subsequent investigation using an electronic portal imaging device (EPID) revealed clinically significant beam position deviations for small steering errors. This prompted investigation into further impact and possible solutions. Testing set-points were established by adjusting transverse steering current to achieve introduced 2 T steering errors. Tests at each set-point included scanning water tank profiles and EPID images. A proposed method for adjusting the 2 T error sensitivity was tested via adjusting the 2 T loop parameter such that a reported error corresponds to specific beam position deviations. The testing set-points resulted in positional deviations of greater than 3 mm for reported errors of less than 1. A new method for improving 2 T error sensitivity was implemented. This work has shown that existing vendor protocol for establishing beam steering error for 6 MV FFF beams can lead to beam position deviations without machine interlocks or significant reported steering errors. Thus, an alternative method of establishing steering error sensitivity based on positional deviations is presented.
引用
下载
收藏
页码:1377 / 1382
页数:6
相关论文
共 50 条
  • [1] Linac reported steering error insensitive to 6 MV FFF transverse beam position deviations
    Jacob McAloney
    Simon Biggs
    Matthew Sobolewski
    Physical and Engineering Sciences in Medicine, 2021, 44 : 1377 - 1382
  • [2] Treatment planning for a 6MV FFF Halcyon linac as a replacement for a 10MV-capable Clinac
    Withers, D.
    Kidane, G.
    Farmakidis, D.
    Crees, L.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S1073 - S1073
  • [3] Dose Kernel of 6 MV FFF Beam in TrueBeam for Pencil Beam Convolution Algorithms
    Tang, G.
    Georgiev, G.
    Lim, S.
    Lovelock, D.
    LoSasso, T.
    Kirov, A.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [4] The characteristics of small field beam quality and output factor of 6 MV FFF
    Azzi, A.
    Ryangga, D.
    Pawiro, S. A.
    18TH ASIA-OCEANIA CONGRESS OF MEDICAL PHYSICS (AOCMP) & 16TH SOUTH-EAST ASIA CONGRESS OF MEDICAL PHYSICS (SEACOMP), 2019, 1248
  • [5] A dosimetric study of linac-based gated SBRT using FFF and standard 6 MV beams
    Blais, Adam
    Juhasz, Janos
    Shim, Martin
    Wierzbicki, Marcin
    Vlad, Roxana
    MEDICAL PHYSICS, 2017, 44 (08) : 4380 - 4380
  • [6] Effect of transverse magnetic fields on a simulated in-line 6 MV linac
    St Aubin, J.
    Steciw, S.
    Fallone, B. G.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (16): : 4861 - 4869
  • [7] The Effect of Transverse Magnetic Fields on a Simulated In-Line 6 MV Linac
    St Aubin, J.
    Steciw, S.
    Fallone, B.
    MEDICAL PHYSICS, 2010, 37 (06) : 3468 - +
  • [8] Verification for small fields of a 6 MV FFF beam model of a varian EDGE accelerator
    Cui, Congwu
    Liu, Zhengfeng
    Therrien, Pierre
    MEDICAL PHYSICS, 2017, 44 (08) : 4385 - 4385
  • [9] Verification of Radiation Isocenter on Linac Beam 6 MV using Computed Radiography
    Irsal, Muhammad
    Hidayanto, Eko
    Sutanto, Heri
    INTERNATIONAL CONFERENCE ON MATHEMATICS: EDUCATION, THEORY AND APPLICATION, 2017, 855
  • [10] Validation and Clinical Implementation of VMAT Liver SBRT on 6MV-FFF O-Ring Halcyon Linac
    Tackett, T.
    Webster, A.
    Bernard, M.
    Kudrimoti, M.
    Pokhrel, D.
    MEDICAL PHYSICS, 2021, 48 (06)