Novel Automated Suturing Technology for Minimally Invasive Mitral Chord Implantation: A Preclinical Evaluation Study

被引:2
|
作者
Werner, Paul [1 ]
Rath, Claus [1 ]
Gross, Christoph [1 ]
Ad, Niv [2 ]
Gosev, Igor [3 ]
Amirjamshidi, Hossein [3 ]
Poschner, Thomas [1 ]
Coti, Iuliana [1 ]
Russo, Marco [4 ]
Mach, Markus [1 ]
Kocher, Alfred [1 ]
Laufer, Guenther [1 ]
Sauer, Jude [3 ]
Andreas, Martin [1 ]
机构
[1] Med Univ Vienna, Dept Cardiac Surg, Waehringer Guertel 18-20, A-1090 Vienna, Austria
[2] Univ Maryland, Dept Surg, Sch Med, Baltimore, MD USA
[3] Med Univ Rochester, Dept Surg, Rochester, NY USA
[4] San Camillo Forlanini Hosp Rome, Rome, Italy
关键词
mitral regurgitation; mitral valve repair; chordal replacement; automated ePTFE suturing; new ePTFE titanium fastener; VALVE REPAIR; LEAFLET PROLAPSE; ARTIFICIAL CHORDAE; REPLACEMENT; RESECTION; REGURGITATION; DURABILITY; LENGTH;
D O I
10.1177/15569845221133381
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective: This study evaluated the ergonomics and time requirements of using a novel automated suturing and titanium fastener deployment technology for chordal replacement in human heart specimens in open and minimally invasive cardiac surgery (MICS) simulators. Methods: Five cardiac surgeons used novel, manually poweredexpandedpolytetrafluoroethylene (ePTFE) suturing devices to automate suture placement between mitral leaflets and papillary muscles in explanted cadaver hearts, along with customized titanium fastener delivery devices to secure suture and trim suture tails. This mitral chordal replacement test was conducted using surgical models simulating open and MICS mitral repair access. The study was approved by the institutional ethical board. Results: After a brief introduction to this technique using plastic models, study surgeons performed 48 chordal replacements in human mitral valves, placing 18 in an open model and 30 in a right minithoracotomy model. The time range to complete a single chordal replacement was between 55 s and 8 min, with an overall mean duration of 3.6 +/- 1.5 min. No difference in duration of implantation was recorded for the MICS and open sternotomy simulators used. Good control of suture delivery was reported in 95.8% (n = 46) of leaflet aspect of the sutures and in 100% (N = 48) of papillary muscle sutures. Conclusions: Automated mitral chordal ePTFE suturing simulated through open and MICS access demonstrated quality handling and accurate placement of sutures in human heart specimens. A clinical trial using this technology is currently ongoing. This innovation may present an important advance facilitating enhanced minimally invasive mitral valve repair.
引用
收藏
页码:506 / 512
页数:7
相关论文
共 50 条
  • [1] Role of Automated Suturing Technology in Minimally Invasive Aortic and Mitral Valve Surgery
    Amirjamshidi, Hossein
    Sauer, Jude S.
    Knight, Peter A.
    SURGICAL TECHNOLOGY INTERNATIONAL-INTERNATIONAL DEVELOPMENTS IN SURGERY AND SURGICAL RESEARCH, 2020, 36
  • [2] Novel Automated Suturing Technology for Minimally Invasive Aortic Valve Replacements
    Wong, Joshua K.
    Melvin, Amber L.
    Siordia, Juan A.
    Joshi, Devang J.
    Sauer, Jude S.
    Knight, Peter A.
    ANNALS OF THORACIC SURGERY, 2018, 105 (02): : 645 - 649
  • [3] Evaluation of a Novel Automated Suturing Technology for Mitral Chordal Implantation: 1-Year Results
    Werner, Paul
    Poschner, Thomas
    Gross, Christoph
    Russo, Marco
    Laufer, Guenther
    Sauer, Jude
    Andreas, Martin
    ANNALS OF THORACIC SURGERY, 2024, 117 (02): : 474 - 478
  • [4] Implementation of Endoscopic Minimally Invasive Mitral Valve Replacement Surgery With Automated Suturing Technology
    Zubarevich, Alina
    Rad, Arian Arjomandi
    Beltsios, Eleftherios
    Salman, Jawad
    Pitsis, Antonios
    Popov, Aron-Frederik
    Schmack, Bastian
    Bakhtiary, Farhad
    Ruhparwar, Arjang
    Weymann, Alexander
    INNOVATIONS-TECHNOLOGY AND TECHNIQUES IN CARDIOTHORACIC AND VASCULAR SURGERY, 2024, 19 (02) : 196 - 203
  • [5] Novel automated suturing technology for synthetic mitral chordal implantation: a first-in-human study
    Werner, P.
    Poschner, T.
    Gross, C.
    Russo, M.
    Laufer, G.
    Sauer, J.
    Andreas, M.
    WIENER KLINISCHE WOCHENSCHRIFT, 2023, 135 : S403 - S403
  • [6] Pilot Study Evaluation of Standard Laparoscopic Suturing and a Novel Parenchymal Apposition Mechanism for Minimally Invasive Renal Reconstruction
    Polland, Allison R.
    Graversen, Joseph A.
    Mues, Adam C.
    Tal, Oded
    Korets, Ruslan
    Landman, Jaime
    JOURNAL OF LAPAROENDOSCOPIC & ADVANCED SURGICAL TECHNIQUES, 2011, 21 (09): : 823 - 827
  • [7] Minimally Invasive Mitral Valve Replacement and Transfemoral Aortic Valve Implantation
    Rustenbach, Christian
    Baumbach, Hardy
    Hill, Stephan
    Franke, Ulrich F. W.
    JOURNAL OF HEART VALVE DISEASE, 2015, 24 (01): : 43 - 45
  • [8] Feasibility study of a mechanical suturing device for less invasive mitral operations
    Huang, CT
    Duhaylongsod, FG
    Burfeind, WR
    Lin, CH
    Morota, T
    Kawata, T
    ANNALS OF THORACIC SURGERY, 1998, 66 (03): : 1029 - 1031
  • [9] Minimally invasive mitral valve repair with and without robotic technology in the elderly
    Tabata, Minoru
    Cohn, Lawrence H.
    AMERICAN JOURNAL OF GERIATRIC CARDIOLOGY, 2006, 15 (05): : 306 - 310
  • [10] Minimally Invasive Mitral Valve Surgery Utilizing Heart Port Technology
    Vallabhajosyula, Prashanth
    Wallen, Tyler J.
    Solometo, Lauren P.
    Fox, Jeanne
    Vernick, William J.
    Hargrove, W. Clark, III
    JOURNAL OF CARDIAC SURGERY, 2014, 29 (03) : 343 - 348