Verification of Accuracy of CyberKnife Tumor-tracking Radiation Therapy Using Patient-specific Lung Phantoms

被引:61
|
作者
Jung, Jinhong [1 ,2 ]
Song, Si Yeol [1 ]
Min, Sang [1 ]
Yoon, KyoungJun [1 ]
Choi, Wonsik [3 ]
Jeong, Seong-Yun [4 ]
Choi, Eun Kyung [1 ]
Cho, Byungchul [1 ]
机构
[1] Univ Ulsan, Coll Med, Asan Med Ctr, Dept Radiat Oncol, Seoul 138736, South Korea
[2] Kyung Hee Univ, Sch Med, Med Ctr, Dept Radiat Oncol, Seoul, South Korea
[3] Univ Ulsan, Gangneung Asan Hosp, Coll Med, Dept Radiat Oncol, Kangnung, South Korea
[4] Univ Ulsan, Coll Med, Asan Med Ctr, Asan Inst Life Sci, Seoul 138736, South Korea
基金
新加坡国家研究基金会;
关键词
MOTION COMPENSATION; MARKER PLACEMENT; SYSTEM; RADIOSURGERY; CANCER;
D O I
10.1016/j.ijrobp.2015.02.055
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To investigate the accuracy of the CyberKnife Xsight Lung Tracking System (XLTS) compared with that of a fiducial-based target tracking system (FTTS) using patient-specific lung phantoms. Methods and Materials: Three-dimensional printing technology was used to make individualized lung phantoms that closely mimicked the lung anatomy of actual patients. Based on planning computed tomographic data from 6 lung cancer patients who underwent stereotactic ablative radiation therapy using the CyberKnife, the volume above a certain Hounsfield unit (HU) was assigned as the structure to be filled uniformly with polylactic acid material by a 3-dimensional printer (3D Edison, Lokit, Korea). We evaluated the discrepancies between the measured and modeled target positions, representing the total tracking error, using 3 log files that were generated during each treatment for both the FTTS and the XLTS. We also analyzed the gamma index between the film dose measured under the FTTS and XLTS. Results: The overall mean values and standard deviations of total tracking errors for the FTTS were 0.36 +/- 0.39 mm, 0.15 +/- 0.64 mm, and 0.15 +/- 0.62 mm for the craniocaudal (CC), left-right (LR), and anteroposterior (AP) components, respectively. Those for the XLTS were 0.38 +/- 0.54 mm, 0.13 +/- 0.18 mm, and 0.14 +/- 0.37 mm for the CC, LR, and AP components, respectively. The average of gamma passing rates was 100% for the criteria of 3%, 3 mm; 99.6% for the criteria of 2%, 2 mm; and 86.8% for the criteria of 1%, 1 mm. Conclusions: The XLTS has segmentation accuracy comparable with that of the FTTS and small total tracking errors. (C) 2015 Elsevier Inc. All rights reserved.
引用
收藏
页码:745 / 753
页数:9
相关论文
共 50 条
  • [41] The Accuracy of Three-Dimensional Planned Bone Tumor Resection Using Patient-Specific Instrument
    Mueller, Daniel A.
    Stutz, Yannik
    Vlachopoulos, Lazaros
    Farshad, Mazda
    Furnstahl, Philipp
    CANCER MANAGEMENT AND RESEARCH, 2020, 12 : 6533 - 6540
  • [42] Dosimetric Evaluation of Gated Volumetric Modulated Arc Radiation Therapy (VMAT) and Tumor-Tracking Robotic Radiation Therapy for Hepatocellular Carcinoma
    Yoon, K.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (02): : E629 - E629
  • [43] Dynamic tumor-tracking stereotactic body radiation therapy for a solitary tumor in a transplanted organ: two case reports
    Okabayashi, Shun
    Matsuo, Yukinori
    Kishi, Noriko
    Hanazawa, Hideki
    Mizowaki, Takashi
    INTERNATIONAL CANCER CONFERENCE JOURNAL, 2020, 9 (04) : 221 - 226
  • [44] A Retrospective Multicenter Study of Stereotactic Body Radiation Therapy Using a Real-Time Tumor-Tracking Radiation Therapy System for Stage I Non-Small Cell Lung Cancer
    Katoh, N.
    Soda, I.
    Tamamura, H.
    Takahashi, S.
    Inoue, T.
    Onimaru, R.
    Shibuya, K.
    Hayakawa, K.
    Shirato, H.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (02): : S507 - S507
  • [45] Evaluation of the accuracy of Cyberknife tumor tracking using gafchromic EBT3 films
    Mastella, E.
    Vigorito, S.
    Rondi, E.
    Cattani, F.
    Piperno, G.
    Ferrari, A. M.
    Rozza, D.
    Strata, E.
    Orecchia, R.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S415 - S415
  • [46] Feasibility of a Patient-Specific Bolus Using the Life-Casting Method for Radiation Therapy
    Kim, Jeongho
    Park, Jeehoon
    Park, Beomjun
    Park, Byungdo
    Kim, Tae-Gyu
    APPLIED SCIENCES-BASEL, 2023, 13 (17):
  • [47] A Novel Markerless Lung Tumor-Tracking Method Using Treatment MV Beam Imaging
    Rozario, Timothy
    Chiu, Tsuicheng D.
    Chen, Mingli
    Jia, Xun
    Lu, Weiguo
    Bereg, Sergey
    Mao, Weihua
    APPLIED SCIENCES-BASEL, 2018, 8 (12):
  • [48] PixelPrint: Three-dimensional printing of realistic patient-specific lung phantoms for CT imaging
    Shapira, Nadav
    Donovan, Kevin
    Mei, Kai
    Geagan, Michael
    Roshkovan, Leonid
    Litt, Harold, I
    Gang, Grace J.
    Stayman, J. Webster
    Shinohara, Russell T.
    Noel, Peter B.
    MEDICAL IMAGING 2022: PHYSICS OF MEDICAL IMAGING, 2022, 12031
  • [49] Three-dimensional printing of patient-specific computed tomography lung phantoms: a reader study
    Shapira, Nadav
    Donovan, Kevin
    Mei, Kai
    Geagan, Michael
    Roshkovan, Leonid
    Gang, Grace J.
    Abed, Mohammed
    Linna, Nathaniel B.
    Cranston, Coulter P.
    O'Leary, Cathal N.
    Dhanaliwala, Ali H.
    Kontos, Despina
    Litt, Harold I.
    Stayman, J. Webster
    Shinohara, Russell T.
    Noel, Peter B.
    PNAS NEXUS, 2023, 2 (03):
  • [50] Feasibility study of patient-specific energy verification using a multilayer acrylic-disk radiation sensor
    Son, Jaeman
    Shin, Dongho
    Kim, Taeho
    Park, Sukwon
    Rah, Jeong-Eun
    MEDICAL PHYSICS, 2020, 47 (09) : 3789 - 3796