Verification of Accuracy of CyberKnife Tumor-tracking Radiation Therapy Using Patient-specific Lung Phantoms

被引:61
|
作者
Jung, Jinhong [1 ,2 ]
Song, Si Yeol [1 ]
Min, Sang [1 ]
Yoon, KyoungJun [1 ]
Choi, Wonsik [3 ]
Jeong, Seong-Yun [4 ]
Choi, Eun Kyung [1 ]
Cho, Byungchul [1 ]
机构
[1] Univ Ulsan, Coll Med, Asan Med Ctr, Dept Radiat Oncol, Seoul 138736, South Korea
[2] Kyung Hee Univ, Sch Med, Med Ctr, Dept Radiat Oncol, Seoul, South Korea
[3] Univ Ulsan, Gangneung Asan Hosp, Coll Med, Dept Radiat Oncol, Kangnung, South Korea
[4] Univ Ulsan, Coll Med, Asan Med Ctr, Asan Inst Life Sci, Seoul 138736, South Korea
基金
新加坡国家研究基金会;
关键词
MOTION COMPENSATION; MARKER PLACEMENT; SYSTEM; RADIOSURGERY; CANCER;
D O I
10.1016/j.ijrobp.2015.02.055
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To investigate the accuracy of the CyberKnife Xsight Lung Tracking System (XLTS) compared with that of a fiducial-based target tracking system (FTTS) using patient-specific lung phantoms. Methods and Materials: Three-dimensional printing technology was used to make individualized lung phantoms that closely mimicked the lung anatomy of actual patients. Based on planning computed tomographic data from 6 lung cancer patients who underwent stereotactic ablative radiation therapy using the CyberKnife, the volume above a certain Hounsfield unit (HU) was assigned as the structure to be filled uniformly with polylactic acid material by a 3-dimensional printer (3D Edison, Lokit, Korea). We evaluated the discrepancies between the measured and modeled target positions, representing the total tracking error, using 3 log files that were generated during each treatment for both the FTTS and the XLTS. We also analyzed the gamma index between the film dose measured under the FTTS and XLTS. Results: The overall mean values and standard deviations of total tracking errors for the FTTS were 0.36 +/- 0.39 mm, 0.15 +/- 0.64 mm, and 0.15 +/- 0.62 mm for the craniocaudal (CC), left-right (LR), and anteroposterior (AP) components, respectively. Those for the XLTS were 0.38 +/- 0.54 mm, 0.13 +/- 0.18 mm, and 0.14 +/- 0.37 mm for the CC, LR, and AP components, respectively. The average of gamma passing rates was 100% for the criteria of 3%, 3 mm; 99.6% for the criteria of 2%, 2 mm; and 86.8% for the criteria of 1%, 1 mm. Conclusions: The XLTS has segmentation accuracy comparable with that of the FTTS and small total tracking errors. (C) 2015 Elsevier Inc. All rights reserved.
引用
收藏
页码:745 / 753
页数:9
相关论文
共 50 条
  • [31] Patient-Specific Qa for Cyberknife Using a Portable Flat-Panel Imager
    Gersh, J.
    DuBose, R.
    Shay, K.
    Fulkerson, R.
    MEDICAL PHYSICS, 2017, 44 (06) : 3061 - 3061
  • [32] Evaluation of Tumor Tracking Accuracy of CyberKnife Synchrony System Using a Plastic Scintillator
    Akino, Y.
    Sumida, I.
    Shiomi, H.
    Miyamoto, Y.
    Higashinaka, N.
    Murashima, Y.
    Ogawa, K.
    MEDICAL PHYSICS, 2015, 42 (06) : 3712 - 3712
  • [33] Accuracy of patient-specific temporal implants using PEKK
    Khashaba, Mohammed M.
    Shaheen, Hamed A.
    Ibrahim, Waleed H.
    AlDainy, Duaa Gawad
    JOURNAL OF CRANIO-MAXILLOFACIAL SURGERY, 2021, 49 (10) : 943 - 949
  • [34] Patient-Specific Quality Assurance for Monte Carlo-Calculated Lung SBRT On Cyberknife - Is It Necessary?
    Fabien, J.
    Zhang, Y.
    Brindle, J.
    Dobbins, D.
    Podder, T.
    Wessels, B.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [35] Validation of patient-specific deep learning markerless lung tumor tracking aided by 4DCBCT
    Huang, L.
    Thummerer, A.
    Papadopoulou, C., I
    Corradini, S.
    Belka, C.
    Riboldi, M.
    Kurz, C.
    Landry, G.
    PHYSICS IN MEDICINE AND BIOLOGY, 2025, 70 (06):
  • [36] Deep inspiration breath hold real-time tumor-tracking radiation therapy (DBRT) as a novel stereotactic body radiation therapy approach for lung tumors
    Tanaka, Hidekazu
    Ono, Taiki
    Ueda, Kazushi
    Karita, Masako
    Manabe, Yuki
    Kajima, Miki
    Sera, Tatsuhiro
    Fujimoto, Koya
    Yuasa, Yuki
    Shiinoki, Takehiro
    SCIENTIFIC REPORTS, 2024, 14 (01)
  • [37] Deep inspiration breath hold real-time tumor-tracking radiation therapy (DBRT) as a novel stereotactic body radiation therapy approach for lung tumors
    Hidekazu Tanaka
    Taiki Ono
    Kazushi Ueda
    Masako Karita
    Yuki Manabe
    Miki Kajima
    Tatsuhiro Sera
    Koya Fujimoto
    Yuki Yuasa
    Takehiro Shiinoki
    Scientific Reports, 14
  • [38] Development of a high-resolution two-dimensional detector-based dose verification system for tumor-tracking irradiation in the CyberKnife system
    Kawabata, Fumitaka
    Kamomae, Takeshi
    Okudaira, Kuniyasu
    Komori, Masataka
    Oguchi, Hiroshi
    Sasaki, Motoharu
    Mori, Masaki
    Kawamura, Mariko
    Abe, Shinji
    Ishihara, Shunichi
    Naganawa, Shinji
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (08):
  • [39] Evaluation of Dose Distribution and Tracking Accuracy in Dynamic Tumor-Tracking Irradiation for Liver Tumors Using a Gimbaled Linac
    Iizuka, Y.
    Ueki, N.
    Ishinara, Y.
    Akimoto, M.
    Tanabe, H.
    Takayama, K.
    Mukumoto, N.
    Nakamura, M.
    Miyabe, Y.
    Kaneko, S.
    Matsuo, Y.
    Mizowaki, T.
    Monzen, H.
    Sawada, A.
    Kokubo, M.
    Hiraoka, M.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S369 - S370
  • [40] Development of a Margin Determination Framework for Tumor-Tracking Radiation Therapy With Intraoperatively Implanted Fiducial Markers
    Kim, Jihun
    Han, Min Cheol
    Chang, Jee Suk
    Hong, Chae-Seon
    Kim, Kyung Hwan
    Byun, Hwa Kyung
    Park, Ryeong Hwang
    Koom, Woong Sub
    Park, Seong Yong
    Kim, Jin Sung
    FRONTIERS IN ONCOLOGY, 2021, 11