Verification of Accuracy of CyberKnife Tumor-tracking Radiation Therapy Using Patient-specific Lung Phantoms

被引:61
|
作者
Jung, Jinhong [1 ,2 ]
Song, Si Yeol [1 ]
Min, Sang [1 ]
Yoon, KyoungJun [1 ]
Choi, Wonsik [3 ]
Jeong, Seong-Yun [4 ]
Choi, Eun Kyung [1 ]
Cho, Byungchul [1 ]
机构
[1] Univ Ulsan, Coll Med, Asan Med Ctr, Dept Radiat Oncol, Seoul 138736, South Korea
[2] Kyung Hee Univ, Sch Med, Med Ctr, Dept Radiat Oncol, Seoul, South Korea
[3] Univ Ulsan, Gangneung Asan Hosp, Coll Med, Dept Radiat Oncol, Kangnung, South Korea
[4] Univ Ulsan, Coll Med, Asan Med Ctr, Asan Inst Life Sci, Seoul 138736, South Korea
基金
新加坡国家研究基金会;
关键词
MOTION COMPENSATION; MARKER PLACEMENT; SYSTEM; RADIOSURGERY; CANCER;
D O I
10.1016/j.ijrobp.2015.02.055
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To investigate the accuracy of the CyberKnife Xsight Lung Tracking System (XLTS) compared with that of a fiducial-based target tracking system (FTTS) using patient-specific lung phantoms. Methods and Materials: Three-dimensional printing technology was used to make individualized lung phantoms that closely mimicked the lung anatomy of actual patients. Based on planning computed tomographic data from 6 lung cancer patients who underwent stereotactic ablative radiation therapy using the CyberKnife, the volume above a certain Hounsfield unit (HU) was assigned as the structure to be filled uniformly with polylactic acid material by a 3-dimensional printer (3D Edison, Lokit, Korea). We evaluated the discrepancies between the measured and modeled target positions, representing the total tracking error, using 3 log files that were generated during each treatment for both the FTTS and the XLTS. We also analyzed the gamma index between the film dose measured under the FTTS and XLTS. Results: The overall mean values and standard deviations of total tracking errors for the FTTS were 0.36 +/- 0.39 mm, 0.15 +/- 0.64 mm, and 0.15 +/- 0.62 mm for the craniocaudal (CC), left-right (LR), and anteroposterior (AP) components, respectively. Those for the XLTS were 0.38 +/- 0.54 mm, 0.13 +/- 0.18 mm, and 0.14 +/- 0.37 mm for the CC, LR, and AP components, respectively. The average of gamma passing rates was 100% for the criteria of 3%, 3 mm; 99.6% for the criteria of 2%, 2 mm; and 86.8% for the criteria of 1%, 1 mm. Conclusions: The XLTS has segmentation accuracy comparable with that of the FTTS and small total tracking errors. (C) 2015 Elsevier Inc. All rights reserved.
引用
收藏
页码:745 / 753
页数:9
相关论文
共 50 条
  • [1] Patient-Specific Biomechanical Modeling of the Lung Tumor for Radiation Therapy
    Giroux, M.
    Ladjal, H.
    Beuve, M.
    Giraud, P.
    Shariat, B.
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2017, 20 : 95 - 96
  • [2] Development of patient-specific phantoms for verification of stereotactic body radiation therapy planning in patients with metallic screw fixation
    Oh, Dongryul
    Hong, Chae-Seon
    Ju, Sang Gyu
    Kim, Minkyu
    Koo, Bum Yong
    Choi, SungBack
    Park, Hee Chul
    Choi, Doo Ho
    Pyo, Hongryull
    SCIENTIFIC REPORTS, 2017, 7
  • [3] Development of patient-specific phantoms for verification of stereotactic body radiation therapy planning in patients with metallic screw fixation
    Dongryul Oh
    Chae-Seon Hong
    Sang Gyu Ju
    Minkyu Kim
    Bum Yong Koo
    SungBack Choi
    Hee Chul Park
    Doo Ho Choi
    Hongryull Pyo
    Scientific Reports, 7
  • [4] Using patient-specific phantoms to evaluate deformable image registration algorithms for adaptive radiation therapy
    Stanley, Nick
    Glide-Hurst, Carri
    Kim, Jinkoo
    Adams, Jeffrey
    Li, Shunshan
    Wen, Ning
    Chetty, Indrin J.
    Zhong, Hualiang
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (06): : 177 - 194
  • [5] Patient-Specific Quality Assurance of Respiratory-Gated Radiotherapy for Lung Using New Real-Time Tumor-Tracking Radiotherapy System
    Shiinoki, T.
    Uehara, T.
    Sera, T.
    Yuasa, Y.
    Fujimoto, K.
    Koike, M.
    Hanazawa, H.
    Shibuya, K.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [6] A Patient-Specific Model for Tracking Lung Tumor During Radiotherapy Using Surrogate Signal
    Fakhraei, S.
    Sterling, D.
    Ehler, E.
    Alaei, P.
    MEDICAL PHYSICS, 2020, 47 (06) : E572 - E572
  • [7] Patient-specific respiratory motion management using lung tumors vs fiducial markers for real-time tumor-tracking stereotactic body radiotherapy
    Tanabe, Yoshinori
    Kiritani, Michiru
    Deguchi, Tomomi
    Hira, Nanami
    Tomimoto, Syouta
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2023, 25
  • [8] Baseline Shift of Intrafractional Lung Tumor Motion in Real-Time Tumor-Tracking Radiation Therapy
    Takao, S.
    Miyamoto, N.
    Matsuura, T.
    Shimizu, S.
    Onimaru, R.
    Katoh, N.
    Inoue, T.
    Shirato, H.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (02): : S67 - S67
  • [9] Prediction accuracy of lung tumor stereotactic body radiotherapy using CyberKnife Synchrony tracking
    Chiu, Tin Lok
    Fok, Tsz Ching
    Yip, Natalie Kar Wei
    Yeung, Fu Ki
    Cheung, Chi Wai
    Yu, Siu Ki
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S4140 - S4142
  • [10] Development of patient-specific phantoms for verification of SBRT planning using 3D printer
    Hong, C. S.
    Oh, D.
    Ju, S. G.
    Kim, M.
    Koo, B. Y.
    Park, H. C.
    Choi, D. H.
    Pyo, H.
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S399 - S400