Determination of Entrance Surface dose for Three Common Planar Diagnostic X-Ray Examinations

被引:1
|
作者
Jeyasugiththan, Jeyasingam [1 ]
Ehalagasthanna, G. P. N. T. [1 ]
Hishaam, Urshella [1 ]
Satharasinghe, D. M. [1 ]
机构
[1] Univ Colombo, Fac Sci, Dept Nucl Sci, Colombo 03, Sri Lanka
关键词
RADIOGRAPHY; QUALITY;
D O I
10.1093/rpd/ncac177
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Patient dose during diagnostic radiography procedures primarily depends on the entrance surface dose (ESD) and the exposure area. The purpose of this study was to determine ESD for five common diagnostic X-ray examinations using dose area product and X-ray tube output. The ESDs were estimated in a sample of 340 patients using normalised X-ray output from Philips-PrimaryDiagnost digital X-ray machine equipped with digital flat panel scintillator detector. The resultant mean ESDs for chest postero-anterior/lateral, lumbar spine antero-posterior (AP)/lateral and abdomen AP were 0.49, 2.00, 5.74, 12.79 and 4.62 mGy, respectively. The mean and third quartile ESD values of this study were significantly higher than the values found in the literature. This can be attributed to the lower tube potential (kV(p)) values and the higher tube current-time products (mAs). The methodology used here is a better option for low-resources countries for routine dose monitoring.
引用
收藏
页码:1361 / 1367
页数:7
相关论文
共 50 条
  • [31] Ultra low dose X-ray spinal examinations
    Fessler, P
    Prevot, G
    Hilt, B
    RESEARCH INTO SPINAL DEFORMITIES 3, 2002, 88 : 191 - 195
  • [32] Determination of the surface dose of a water phantom using a semiconductor detector for diagnostic kilovoltage x-ray beams
    Tanabe, Ryuki
    Araki, Fujio
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 84 : 198 - 204
  • [33] Assessment of entrance surface dose and health risk from common radiology examinations in Iran
    Shahbazi-Gahrouei, Daryoush
    Baradaran-Ghahfarokhi, Milad
    RADIATION PROTECTION DOSIMETRY, 2013, 154 (03) : 308 - 313
  • [34] Effective Dose Calculations in Conventional Diagnostic X-Ray Examinations for Patients in Major Sudanese Hospitals
    Suliman, I. I.
    Habbani, F. I.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6, 2007, 14 : 2087 - +
  • [35] Evaluation of dose to tooth enamel from medical diagnostic X-ray examinations at Mayak PA
    Wieser, A.
    Vasilenko, E.
    Zankl, M.
    Greiter, M.
    Ulanovsky, A.
    Sabayev, A.
    Knyazev, V.
    Zahrov, R.
    RADIATION MEASUREMENTS, 2011, 46 (09) : 808 - 812
  • [36] COMMON SENSE IN THE DIAGNOSTIC USE OF X-RAY
    MILLER, JE
    SWINDELL, GE
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1959, 170 (07): : 761 - 765
  • [37] Entrance surface dose and image quality: comparison of adult chest and abdominal X-ray examinations in general practitioner clinics, public and private hospitals in Malaysia
    Hambali, Ahmad Shariff
    Ng, Kwan-Hoong
    Abdullah, Basri Johan Jeet
    Wang, Hwee-Beng
    Jamal, Noriah
    Spelic, David C.
    Suleiman, Orhan H.
    RADIATION PROTECTION DOSIMETRY, 2009, 133 (01) : 25 - 34
  • [38] Local diagnostic reference levels in standard X-ray examinations
    Compagnone, G
    Pagan, L
    Bergamini, C
    RADIATION PROTECTION DOSIMETRY, 2005, 113 (01) : 54 - 63
  • [39] Assessment of Patient Radiation Dose during Conventional Diagnostic X-ray Examinations in Three Public Hospitals in Northern Jordan using TLDs
    Ababneh, Zaid Q.
    Ababneh, Anas M.
    Hani, Majd S. Bani
    Ababneh, Riad S.
    Aljarrah, Khaled M.
    HEALTH PHYSICS, 2020, 118 (06): : 593 - 599
  • [40] Dose loading of adult patients subjected to X-ray examinations
    Stavitskij, R.V.
    Blinov, N.N.
    Varshavskij, Yu.V.
    Barkhudarov, R.M.
    Lebedev, L.A.
    Frid, E.S.
    Sergienko, N.N.
    Meditsinskaya Radiologiya (Medical Radiology), 1998, 43 (01): : 66 - 70