Absorbed dose beam quality correction factors kQ for the NE2571 chamber in a 5 MV and a 10 MV photon beam

被引:30
|
作者
Palmans, H
Mondelaers, W
Thierens, H
机构
[1] Univ Ghent, Dept Biomed Phys, B-9000 Ghent, Belgium
[2] Univ Ghent, Dept Subatom & Radiat Phys, B-9000 Ghent, Belgium
来源
PHYSICS IN MEDICINE AND BIOLOGY | 1999年 / 44卷 / 03期
关键词
D O I
10.1088/0031-9155/44/3/002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Dose to water(D-W) determination in clinical high-energy photon beams with ionization chambers calibrated in terms of absorbed dose to water has been proposed as an alternative to ionization chamber dosimetry based on air kerma calibrations. D-W in the clinical beam is derived using a k(Q) factor that scales the absorbed dose calibration factor in the reference beam to the absorbed dose calibration factor in the user beam. In the present study k(Q) values were determined for the NE2571 chamber in a 5 MV and a 10 MV high-energy photon beam generated at the 15 MeV high-intensity electron linac of the University of Gent. A set of three NE2571 chambers was calibrated relative to the Gent sealed water calorimeter both in Co-60 and in the linac beam at a depth of 5 cm and a source to detector distance of 100 cm. Two high-purity chemical water systems were used in the detection vessel of the calorimeter, H-2-saturated and Ar-saturated pure water, which are both supposed to give a zero heat defect. TPR1020 and %dd(10) have been evaluated as beam quality specifiers. Simulations using the BEAM/DOSXYZ Monte Carlo system were performed to evaluate potential corrections on the measured beam qualities. The average k(Q) values measured for the three NE2571 chambers in the 5 MV and 10 MV photon beams are 0.995 +/- 0.005 and 0.979 +/- 0.005 respectively. For the three chambers used, the maximum deviation of individual k(Q) values is 0.2%. The measured beam quality specifiers %dd(10) and TPR1020 are 67.0 and 0.705 for the 5 MV beam and 75.0 and 0.759 for the 10 MV beam. Although our beam design is very different from those used by other investigators for the measurement of k(Q) values, the agreement with their results is satisfactory showing a slightly better agreement when %dd(10) is used as the beam quality specifier.
引用
收藏
页码:647 / 663
页数:17
相关论文
共 50 条
  • [21] MAGNETIC ANALYSIS OF THE RADIATION COMPONENTS OF A 10 MV PHOTON-BEAM
    LING, CC
    SCHELL, MC
    RUSTGI, SN
    MEDICAL PHYSICS, 1982, 9 (01) : 20 - 26
  • [22] Monte carlo model and output factors of elekta Infinity™ 6 and 10 MV photon beam
    Yani, Sitti
    Budiansah, Indra
    Rhani, Mohamad Fahdillah
    Haryanto, Freddy
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2020, 25 (04) : 470 - 478
  • [23] Direct measurement of ion chamber correction factors, kQ and kB, in a 7 MV MRI-linac
    de Prez, Leon
    Woodings, Simon
    de Pooter, Jacco
    van Asselen, Bram
    Wolthaus, Jochem
    Jansen, Bartel
    Raaymakers, Bas
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (10):
  • [24] Measurements, Simulation, and Analytical Model of Stray Photon and Neutron Absorbed Dose from MV External Beam Radiation Therapy
    Schneider, C.
    Newhauser, W.
    MEDICAL PHYSICS, 2020, 47 (06) : E470 - E470
  • [25] Beam quality conversion factors for parallel-plate ionization chambers in MV photon beams
    Muir, B. R.
    McEwen, M. R.
    Rogers, D. W. O.
    MEDICAL PHYSICS, 2012, 39 (03) : 1618 - 1631
  • [26] MONTE CARLO SIMULATION OF 6, 10 AND 18 MV PHOTON BEAM DOSE DISTRIBUTION IN A BRAIN TUMOR
    Gonzales-Ccoscco, Alberto E.
    Guzman-Calcina, Carmen S.
    Vega-Ramirez, Jose L.
    MOMENTO-REVISTA DE FISICA, 2023, (67): : 22 - 38
  • [27] NEUTRON DOSE DELIVERED WHEN USING 18 MV UNFLATTENED PHOTON BEAM
    Vondracek, V.
    Kralik, M.
    Turek, K.
    Navratil, M.
    Kubes, J.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S476 - S476
  • [28] EFFECT OF FIELD IRREGULARITIES ON THE DOSE DISTRIBUTION OF 4 MV PHOTON-BEAM
    NIROOMANDRAD, A
    PALIWAL, BR
    RADIOTHERAPY AND ONCOLOGY, 1987, 8 (01) : 43 - 48
  • [29] EFFECT OF FIELD IRREGULARITIES ON THE DOSE DISTRIBUTION OF 4 MV PHOTON-BEAM
    NIROOMANDRAD, A
    PALIWAL, B
    MEDICAL PHYSICS, 1987, 14 (03) : 482 - 482
  • [30] Effect of various dental restorations on dose distribution of 6 MV photon beam
    Azizi, Mona
    Mowlavi, Ali Asghar
    Ghorbani, Mahdi
    Davenport, David
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2017, 13 (03) : 538 - 543