Absorbed dose beam quality correction factors kQ for the NE2571 chamber in a 5 MV and a 10 MV photon beam

被引:30
|
作者
Palmans, H
Mondelaers, W
Thierens, H
机构
[1] Univ Ghent, Dept Biomed Phys, B-9000 Ghent, Belgium
[2] Univ Ghent, Dept Subatom & Radiat Phys, B-9000 Ghent, Belgium
来源
PHYSICS IN MEDICINE AND BIOLOGY | 1999年 / 44卷 / 03期
关键词
D O I
10.1088/0031-9155/44/3/002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Dose to water(D-W) determination in clinical high-energy photon beams with ionization chambers calibrated in terms of absorbed dose to water has been proposed as an alternative to ionization chamber dosimetry based on air kerma calibrations. D-W in the clinical beam is derived using a k(Q) factor that scales the absorbed dose calibration factor in the reference beam to the absorbed dose calibration factor in the user beam. In the present study k(Q) values were determined for the NE2571 chamber in a 5 MV and a 10 MV high-energy photon beam generated at the 15 MeV high-intensity electron linac of the University of Gent. A set of three NE2571 chambers was calibrated relative to the Gent sealed water calorimeter both in Co-60 and in the linac beam at a depth of 5 cm and a source to detector distance of 100 cm. Two high-purity chemical water systems were used in the detection vessel of the calorimeter, H-2-saturated and Ar-saturated pure water, which are both supposed to give a zero heat defect. TPR1020 and %dd(10) have been evaluated as beam quality specifiers. Simulations using the BEAM/DOSXYZ Monte Carlo system were performed to evaluate potential corrections on the measured beam qualities. The average k(Q) values measured for the three NE2571 chambers in the 5 MV and 10 MV photon beams are 0.995 +/- 0.005 and 0.979 +/- 0.005 respectively. For the three chambers used, the maximum deviation of individual k(Q) values is 0.2%. The measured beam quality specifiers %dd(10) and TPR1020 are 67.0 and 0.705 for the 5 MV beam and 75.0 and 0.759 for the 10 MV beam. Although our beam design is very different from those used by other investigators for the measurement of k(Q) values, the agreement with their results is satisfactory showing a slightly better agreement when %dd(10) is used as the beam quality specifier.
引用
收藏
页码:647 / 663
页数:17
相关论文
共 50 条
  • [2] Experimental determination of beam quality factors, kQ, for two types of Farmer chamber in a 10 MV photon and a 175 MeV proton beam
    Medin, J
    Ross, CK
    Klassen, NV
    Palmans, H
    Grusell, E
    Grindborg, JE
    PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (06): : 1503 - 1521
  • [3] Variation of Beam Quality Correction Factor with Time: Cases for Ne2571 and Ne2581 Chambers in the 6 and 10 Mv X-Rays Beams
    Samat, S. B.
    Too, S. X.
    Kadni, T.
    Dolah, M. T.
    SAINS MALAYSIANA, 2009, 38 (02): : 203 - 207
  • [4] Computation of the electron beam quality kQ,Q0 factors for the NE2571, NE2571A and NE2581A thimble ionization chambers using PENELOPE
    Erazo, F.
    Brualla, L.
    Lallena, A. M.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 38 : 76 - 80
  • [5] Calculated beam quality correction factors for ionization chambers in MV photon beams
    Tikkanen, J.
    Zink, K.
    Pimpinella, M.
    Teles, P.
    Borbinha, J.
    Ojala, J.
    Siiskonen, T.
    Goma, C.
    Pinto, M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (07):
  • [6] Calorimetric determination of kQ factors for NE 2561 and NE 2571 ionization chambers in 5 cm x 5 cm and 10 cm x 10 cm radiotherapy beams of 8 MV and 16 MV photons
    Krauss, Achim
    Kapsch, Ralf-Peter
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (20): : 6243 - 6259
  • [7] PREDICTION OF SATURNE-II + 10 MV AND 23 MV PHOTON-BEAM OUTPUT FACTORS
    SZYMCZYK, W
    GORACZKO, A
    LESIAK, J
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (03): : 789 - 793
  • [8] Photon beam quality correction factors for the NE2571A and NE2581A thimble ionization chambers using PENELOPE
    Erazo, Fabian
    Lallena, Antonio M.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (01): : 232 - 236
  • [9] Effect of density on the 10 MV photon beam penumbra
    Hoban, P.W.
    Keal, P.J.
    Round, W.H.
    Australasian Physical and Engineering Sciences in Medicine, 1992, 15 (03): : 113 - 123
  • [10] PRACTICAL DOSIMETRIC CONSIDERATIONS OF A 10 MV PHOTON BEAM
    VANDYK, J
    PHYSICS IN MEDICINE AND BIOLOGY, 1977, 22 (03): : 569 - 569