Automatic tube-current modulation in CT -: A comparison between different solutions

被引:23
|
作者
Althén, JN [1 ]
机构
[1] Linkoping Univ Hosp, Dept Radiat Phys, SE-58185 Linkoping, Sweden
关键词
D O I
10.1093/rpd/nch501
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
In this study, tube-current modulation systems on two different CT equipments have been evaluated: Care Dose from Siemens and Auto mA from GE Medical Systems. Care Dose modulates the tube current in the xy-plane during rotation whereas Auto mA modulates the tube currant in the z-direction. xy-Plane modulation was investigated by using an elliptic Polymethylmethacrylate phantom and a CTDI-ion chamber. To investigate modulation in the z-direction. an anthropomorphic dosimetry phantom (Atom) was used. Tests performed with and without tube-current modulation were compared with respect to absorbed dose and image quality. In the anthropomorphic phantom measurements, the dose savings were 15% using Care Dose and the photon starvation artefacts were negligible. Using Auto mA the absorbed dose depends on the chosen noise level. Image noise becomes more constant throughout the patient but photon starvation artefacts remain. We conclude that the two tube-current modulation techniques show different dose advantages, and image quality artefacts.
引用
收藏
页码:308 / 312
页数:5
相关论文
共 50 条
  • [31] INVESTIGATION OF THE INFLUENCE OF IMAGE RECONSTRUCTION FILTER AND SCAN PARAMETERS ON OPERATION OF AUTOMATIC TUBE CURRENT MODULATION SYSTEMS FOR DIFFERENT CT SCANNERS
    Sookpeng, Supawitoo
    Martin, Colin J.
    Gentle, David J.
    RADIATION PROTECTION DOSIMETRY, 2015, 163 (04) : 521 - 530
  • [32] Potential for Radiation Dose Savings in Abdominal and Chest CT Using Automatic Tube Voltage Selection in Combination With Automatic Tube Current Modulation
    Mayer, Caroline
    Meyer, Mathias
    Fink, Christian
    Schmidt, Bernhard
    Sedlmair, Martin
    Schoenberg, Stefan O.
    Henzler, Thomas
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 203 (02) : 292 - 299
  • [33] Effect of CT Scan Parameters on Adult and Pediatric CT Dose When Using Automatic Tube Current Modulation
    Winslow, J.
    Kappadath, S.
    MEDICAL PHYSICS, 2010, 37 (06) : 3102 - +
  • [34] Assessment of tube current modulation in pelvic CT
    Iball, GR
    Brettle, DS
    Moore, AC
    BRITISH JOURNAL OF RADIOLOGY, 2006, 79 (937): : 62 - 70
  • [35] A phantom study to optimise the automatic tube current modulation for chest CT in COVID-19
    Gombolevskiy, Victor
    Morozov, Sergey
    Chernina, Valeria
    Blokhin, Ivan
    Vassileva, Jenia
    EUROPEAN RADIOLOGY EXPERIMENTAL, 2021, 5 (01)
  • [36] Automatic tube current modulation for volume scan in a 320-detector row CT scanner
    Liao, Y. L.
    Lai, N. K.
    Tyan, Y. S.
    Chuang, K. S.
    Tsai, H. Y.
    RADIATION MEASUREMENTS, 2013, 56 : 328 - 332
  • [37] A phantom study to optimise the automatic tube current modulation for chest CT in COVID-19
    Victor Gombolevskiy
    Sergey Morozov
    Valeria Chernina
    Ivan Blokhin
    Jenia Vassileva
    European Radiology Experimental, 5
  • [38] Evaluating the impact of scan settings on automatic tube current modulation in CT using a novel phantom
    Merzan, Deborah
    Nowik, Patrik
    Poludniowski, Gavin
    Bujila, Robert
    BRITISH JOURNAL OF RADIOLOGY, 2017, 90 (1069):
  • [39] Evaluating Effective Doses and Image Quality for Multislice CT Using Automatic Tube Current Modulation
    Tsai, S. F.
    Tsai, H. Y.
    Tyan, Y. S.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [40] Effect of Localizer Radiography Projection on Organ Dose at Chest CT with Automatic Tube Current Modulation
    Saltybaeva, Natalia
    Krauss, Andreas
    Alkadhi, Hatem
    RADIOLOGY, 2017, 282 (03) : 842 - 849