Evaluating the impact of scan settings on automatic tube current modulation in CT using a novel phantom

被引:27
|
作者
Merzan, Deborah [1 ]
Nowik, Patrik [1 ]
Poludniowski, Gavin [1 ]
Bujila, Robert [1 ,2 ]
机构
[1] Karolinska Univ Hosp, Med Radiat Phys & Nucl Med, SE-17176 Stockholm, Sweden
[2] Royal Inst Technol, Dept Phys, SE-10691 Stockholm, Sweden
来源
BRITISH JOURNAL OF RADIOLOGY | 2017年 / 90卷 / 1069期
关键词
EXPOSURE CONTROL; DOSE REDUCTION; IMAGE QUALITY; SYSTEM; NOISE;
D O I
10.1259/bjr.20160308
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The aim of this study was to make a comprehensive evaluation of how variable scan settings can affect the performance of automatic tube current modulation (ATCM) in recent CT scanners from the four major manufacturers. Methods: A phantom was designed and manufactured for the purpose of evaluating ATCM. The phantom was scanned with four categories of systematically varied settings (scan projection radiograph, technique and reconstruction parameters and phantom miscentring). The performance of ATCM, in terms of applied tube current and noise uniformity, for the scans with varied settings was compared with a reference scan using subjective and quantitative approaches. Results: The ATCM implemented by each manufacturer is based on different principles and any affect to the performance of the ATCM, when varying scan settings, will manifest differently among the vendors. The results are summarized in four tables corresponding to the categories of varied settings. Conclusion: The developed phantom proved useful for evaluating the ATCM. It is important to understand how different implementations (vendor specific) of ATCM perform in order to make informed decisions about the selection of scan settings when designing protocols. The resulting tables can serve as a reference for understanding the different implementations of ATCM and highlight settings that should be taken into consideration when adjusting an imaging protocol. Advances in knowledge: The results from this work can serve as a reference for how changes in geometry or scan settings can affect the performance of ATCM, in terms of tube current and noise.
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Technical Note: Evaluating automatic tube current modulation in CT using the standard CTDI dosimetry phantom
    Papadakis, Antonios E.
    Damilakis, John
    MEDICAL PHYSICS, 2021, 48 (02) : 659 - 666
  • [2] Evaluation of Changes in Dose Estimation on Abdomen CT Scan with Automatic Tube Current Modulation Using In-House Phantom
    Taopik, A.
    Lubis, L. E.
    Sihono, D. S. K.
    Soejoko, D. S.
    ATOM INDONESIA, 2023, 49 (03) : 169 - 175
  • [3] A Novel Phantom to Evaluate Longitudinal and Angular Automatic Tube Current Modulation (ATCM) in CT
    Merzan, D.
    Bujila, R.
    Nowik, P.
    MEDICAL PHYSICS, 2015, 42 (06) : 3643 - 3643
  • [4] Effect of CT Scan Parameters on Adult and Pediatric CT Dose When Using Automatic Tube Current Modulation
    Winslow, J.
    Kappadath, S.
    MEDICAL PHYSICS, 2010, 37 (06) : 3102 - +
  • [5] Evaluating Effective Doses and Image Quality for Multislice CT Using Automatic Tube Current Modulation
    Tsai, S. F.
    Tsai, H. Y.
    Tyan, Y. S.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [6] A Phantom Evaluation of the Use of Ct Automatic Tube Current Modulation with Low Tube Potentials for Iodinated Contrast Studies
    Sookpeng, Supawitoo
    Martin, Colin J.
    RADIATION PROTECTION DOSIMETRY, 2022, 198 (03) : 188 - 195
  • [7] A phantom study to optimise the automatic tube current modulation for chest CT in COVID-19
    Gombolevskiy, Victor
    Morozov, Sergey
    Chernina, Valeria
    Blokhin, Ivan
    Vassileva, Jenia
    EUROPEAN RADIOLOGY EXPERIMENTAL, 2021, 5 (01)
  • [8] A phantom study to optimise the automatic tube current modulation for chest CT in COVID-19
    Victor Gombolevskiy
    Sergey Morozov
    Valeria Chernina
    Ivan Blokhin
    Jenia Vassileva
    European Radiology Experimental, 5
  • [9] Misoperation of CT Automatic Tube Current Modulation Systems with Inappropriate Patient Centering: Phantom Studies
    Matsubara, Kosuke
    Koshida, Kichiro
    Ichikawa, Katsuhiro
    Suzuki, Masayuki
    Takata, Tadanori
    Yamamoto, Tomoyuki
    Matsui, Osamu
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 192 (04) : 862 - 865
  • [10] Comparison of different phantom designs for CT scanner automatic tube current modulation system tests
    Sookpeng, S.
    Martin, C. J.
    Gentle, D. J.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2013, 33 (04) : 735 - 761