Effect of Anisotropic Electrical Conductivity Induced by Fiber Orientation on Ablation Characteristics of Pulsed Field Ablation in Atrial Fibrillation Treatment: A Computational Study

被引:8
|
作者
Zang, Lianru [1 ]
Gu, Kaihao [1 ]
Ji, Xingkai [1 ]
Zhang, Hao [1 ]
Yan, Shengjie [1 ]
Wu, Xiaomei [1 ,2 ,3 ,4 ,5 ]
机构
[1] Fudan Univ, Ctr Biomed Engn, Sch Informat Sci & Technol, Shanghai 200438, Peoples R China
[2] Fudan Univ, Acad Engn & Technol, Shanghai 200433, Peoples R China
[3] Fudan Univ, Key Lab Med Imaging Comp & Comp Assisted Interven, Shanghai 200032, Peoples R China
[4] Shanghai Engn Res Ctr Assist Devices, Shanghai 200093, Peoples R China
[5] Fudan Univ, Yiwu Res Inst, Yiwu 322000, Peoples R China
关键词
pulsed field ablation; atrial fibrillation; fiber orientation; anisotropic electrical conductivity; finite element analysis; RADIOFREQUENCY CARDIAC ABLATION; TISSUE; FLOW; CATHETER; LESIONS;
D O I
10.3390/jcdd9100319
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Pulsed field ablation (PFA) is a promising new ablation modality for the treatment of atrial fibrillation (AF); however, the effect of fiber orientation on the ablation characteristics of PFA in AF treatment is still unclear, which is likely an essential factor in influencing the ablation characteristics. This study constructed an anatomy-based left atrium (LA) model incorporating fiber orientation and selected various electrical conductivity and ablation targets to investigate the effect of anisotropic electrical conductivity (AC), compared with isotropic electrical conductivity (IC), on the ablation characteristics of PFA in AF treatment. The results show that the percentage differences in the size of the surface ablation area between AC and IC are greater than 73.71%; the maximum difference in the size of the ablation isosurface between AC and IC at different locations in the atrial wall is 3.65 mm (X-axis), 3.65 mm (Z-axis), and 4.03 mm (X-axis), respectively; and the percentage differences in the size of the ablation volume are greater than 6.9%. Under the condition of the pulse, the amplitude is 1000 V, the total PFA duration is 1 s, and the pulse train interval is 198.4 ms; the differences in the temperature increase between AC and IC in LA are less than 2.46 degrees C. Hence, this study suggests that in further exploration of the computational study of PFA in AF treatment using the same or similar conditions as those used here (myocardial electrical conductivity, pulse parameters, and electric field intensity damage threshold), to obtain more accurate computational results, it is necessary to adopt AC rather than IC to investigate the size of the surface ablation area, the size of the ablation isosurface, or the size of the ablation volume generated by PFA in LA. Moreover, if only investigating the temperature increase generated by PFA in LA, adopting IC instead of AC for simplifying the model construction process is reasonable.
引用
收藏
页数:22
相关论文
共 50 条
  • [31] Assessing recurrence following pulsed field ablation for atrial fibrillation
    Tan, Nicholas Y.
    DeSimone, Christopher V.
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2023, 66 (09) : 1961 - 1963
  • [32] Pulsed field ablation for atrial fibrillation: is it truly as good as it sounds?
    David H. Yoo
    E. Kevin Heist
    Journal of Interventional Cardiac Electrophysiology, 2024, 67 : 237 - 238
  • [33] Hemolysis after pulsed-field ablation of atrial fibrillation
    Xu, Yuankai
    Gulburak, Taalaibek Kyzy
    Lu, Yanmei
    Zhang, Jianghua
    TuErhong, Zukela
    Tang, Baopeng
    Zhou, Xianhui
    HEART RHYTHM, 2025, 22 (04) : 1103 - 1109
  • [34] Pulsed Field or Conventional Thermal Ablation for Paroxysmal Atrial Fibrillation
    Reddy, Vivek Y.
    Gerstenfeld, Edward P.
    Natale, Andrea
    Whang, William
    Cuoco, Frank A.
    Patel, Chinmay
    Mountantonakis, Stavros E.
    Gibson, Douglas N.
    Harding, John D.
    Ellis, Christopher R.
    Ellenbogen, Kenneth A.
    DeLurgio, David B.
    Osorio, Jose
    Achyutha, Anitha B.
    Schneider, Christopher W.
    Mugglin, Andrew S.
    Albrecht, Elizabeth M.
    Stein, Kenneth M.
    Lehmann, John W.
    Mansour, Moussa
    NEW ENGLAND JOURNAL OF MEDICINE, 2023, 389 (18): : 1660 - 1671
  • [35] Assessing recurrence following pulsed field ablation for atrial fibrillation
    Nicholas Y. Tan
    Christopher V. DeSimone
    Journal of Interventional Cardiac Electrophysiology, 2023, 66 : 1961 - 1963
  • [36] Pulsed Field Ablation for Pulmonary Vein Isolation in Atrial Fibrillation
    Reddy, Vivek Y.
    Neuzil, Petr
    Koruth, Jacob S.
    Petru, Jan
    Funosako, Moritoshi
    Cochet, Hubert
    Sediva, Lucie
    Chovanec, Milan
    Dukkipati, Srinivas R.
    Jais, Pierre
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2019, 74 (03) : 315 - 326
  • [37] Pulsed-field ablation: a revolution in atrial fibrillation therapy
    Maizels, Leonid
    Kalman, Jonathan M.
    NATURE REVIEWS CARDIOLOGY, 2024, 21 (08) : 519 - 520
  • [38] The Effect of Discharge Mode on the Distribution of Myocardial Pulsed Electric Field-A Simulation Study for Pulsed Field Ablation of Atrial Fibrillation
    Ji, Xingkai
    Zhang, Hao
    Zang, Lianru
    Yan, Shengjie
    Wu, Xiaomei
    JOURNAL OF CARDIOVASCULAR DEVELOPMENT AND DISEASE, 2022, 9 (04)
  • [39] A New Hope for the Treatment of Atrial Fibrillation: Application of Pulsed-Field Ablation Technology
    Wang, Zhen
    Liang, Ming
    Sun, Jingyang
    Zhang, Jie
    Han, Yaling
    JOURNAL OF CARDIOVASCULAR DEVELOPMENT AND DISEASE, 2024, 11 (06)
  • [40] Epicardial Pulsed Field Ablation for the Treatment of Paroxysmal Atrial Fibrillation During Cardiac Surgery
    Musikantow, Daniel R.
    Reddy, Vivek Y.
    Shaburishvili, Tamaz
    van Zyl, Martin
    O'Brien, Barry
    Coffey, Ken
    Reilly, John
    Asirvatham, Samuel
    de Groot, Joris R.
    JACC-CLINICAL ELECTROPHYSIOLOGY, 2024, 10 (09) : 2097 - 2099