Validation of a Brachial Cuff-Based Method for Estimating Central Systolic Blood Pressure

被引:378
|
作者
Weber, Thomas [1 ,2 ]
Wassertheurer, Siegfried [3 ,4 ]
Rammer, Martin [1 ]
Maurer, Edwin [1 ]
Hametner, Bernhard [3 ]
Mayer, Christopher C. [3 ]
Kropf, Johannes [3 ,4 ]
Eber, Bernd [1 ]
机构
[1] Klinikum Wels Grieskirchen, Dept Cardiol, A-4600 Wels, Austria
[2] Paracelsus Med Univ, Salzburg, Austria
[3] Austrian Inst Technol, Hlth & Environm Dept, Vienna, Austria
[4] Vienna Univ Technol, Inst Anal & Sci Comp, A-1040 Vienna, Austria
关键词
central systolic blood pressure; oscillometric blood pressure monitor; arterial waveforms; validation study; CENTRAL AORTIC PRESSURE; PREDICTS CARDIOVASCULAR EVENTS; GENERALIZED TRANSFER-FUNCTION; PULSE PRESSURE; ARTERIAL STIFFNESS; IMPACT; DRUGS; RISK;
D O I
10.1161/HYPERTENSIONAHA.111.176313
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
The prognostic value of central systolic blood pressure has been established recently. At present, its noninvasive assessment is limited by the need of dedicated equipment and trained operators. Moreover, ambulatory and home blood pressure monitoring of central pressures are not feasible. An algorithm enabling conventional automated oscillometric blood pressure monitors to assess central systolic pressure could be of value. We compared central systolic pressure, calculated with a transfer-function like method (ARCSolver algorithm), using waveforms recorded with a regular oscillometric cuff suitable for ambulatory measurements, with simultaneous high-fidelity invasive recordings, and with noninvasive estimations using a validated device, operating with radial tonometry and a generalized transfer function. Both studies revealed a good agreement between the oscillometric cuff-based central systolic pressure and the comparator. In the invasive study, composed of 30 patients, mean difference between oscillometric cuff/ARCSolver-based and invasive central systolic pressures was 3.0 mm Hg (SD: 6.0 mm Hg) with invasive calibration of brachial waveforms and -3.0 mm Hg (SD: 9.5 mm Hg) with noninvasive calibration of brachial waveforms. Results were similar when the reference method (radial tonometry/transfer function) was compared with invasive measurements. In the noninvasive study, composed of 111 patients, mean difference between oscillometric cuff/ARCSolver-derived and radial tonometry/transfer function-derived central systolic pressures was -0.5 mm Hg (SD: 4.7 mm Hg). In conclusion, a novel transfer function-like algorithm, using brachial cuff-based waveform recordings, is suited to provide a realistic estimation of central systolic pressure. (Hypertension. 2011;58:825-832.). Online Data Supplement
引用
收藏
页码:825 / U317
页数:11
相关论文
共 50 条
  • [41] Effect of tissue mechanical properties on cuff-based blood pressure measurements
    Lan, H.
    Al-Jumaily, A. M.
    Lowe, A.
    Hing, W.
    MEDICAL ENGINEERING & PHYSICS, 2011, 33 (10) : 1287 - 1292
  • [42] Determination of central blood pressure by a noninvasive method (brachial blood pressure and QKD interval): a noninvasive validation
    Cremer, Antoine
    Codjo, Leopold
    Butlin, Mark
    Papaioannou, Georgios
    Yeim, Sunthareth
    Jan, Emilie
    Kiat, Hosen
    Avolio, Alberto
    Gosse, Philippe
    JOURNAL OF HYPERTENSION, 2013, 31 (09) : 1847 - 1852
  • [43] Evaluation of a Brachial Cuff and Suprasystolic Waveform Algorithm Method to Noninvasively Derive Central Blood Pressure
    Costello, Benedict T.
    Schultz, Martin G.
    Black, J. Andrew
    Sharman, James E.
    AMERICAN JOURNAL OF HYPERTENSION, 2015, 28 (04) : 480 - 486
  • [44] Comparison of cuff-based and cuffless continuous blood pressure measurements in children and adolescents
    Zachwieja, Jacek
    Neyman-Bartkowiak, Anna
    Rabiega, Alina
    Wojciechowska, Marta
    Barabasz, Malgorzata
    Musielak, Anna
    Silska-Dittmar, Magdalena
    Ostalska-Nowicka, Danuta
    CLINICAL AND EXPERIMENTAL HYPERTENSION, 2020, 42 (06) : 512 - 518
  • [45] P20 Brachial and Radial Systolic Blood Pressure are not the Same: Potential Implications for Validation Protocols Including Brachial Cuff Devices and Wrist-Based Wearables
    Matthew Armstrong
    Martin Schultz
    Dean Picone
    James Sharman
    Nathan Dwyer
    Philip Roberts-Thomson
    Andrew Black
    Artery Research, 2018, 24 (1) : 85 - 85
  • [46] Cuff-Based Oscillometric Central and Brachial Blood Pressures Obtained Through ABPM are Similarly Associated with Renal Organ Damage in Arterial Hypertension
    Fernandez-Llama, Patricia
    Pareja, Julia
    Yun, Sergi
    Vazquez, Susana
    Oliveras, Anna
    Armario, Pedro
    Blanch, Pedro
    Calero, Francesca
    Sierra, Cristina
    de la Sierra, Alejandro
    KIDNEY & BLOOD PRESSURE RESEARCH, 2017, 42 (06): : 1068 - 1077
  • [47] ACQUISITION OF TIME-RESOLVED BRACHIAL PRESSURE WAVEFORMS FROM CUFF-BASED PULSE VOLUME RECORDINGS
    Phan, Timothy S.
    Dakka, Jumana
    Syed, Amer Ahmed
    Vasim, Izzah
    Oldland, Harry G.
    Kewan, Uzma
    Akers, Scott R.
    Chirinos, Julio A.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2017, 69 (11) : 1729 - 1729
  • [48] Correlation Between the Brachial Blood Pressure Values Obtained Using the Cuff Method and the Central Blood Pressure Values Obtained Invasively
    Kobayashi, Hiroshi
    Kinou, Mineko
    Takazawa, Kenji
    INTERNAL MEDICINE, 2013, 52 (15) : 1675 - 1680
  • [49] EFFECTS OF EXERCISE ON BRACHIAL SYSTOLIC BLOOD PRESSURE VERSUS CENTRAL AORTIC SYSTOLIC PRESSURE IN PATIENTS WITH HYPERTENSION
    Williams, Bryan
    Baschiera, Fabio
    Lacy, Peter S.
    Papst, Cheraz Cherif
    Brunel, Patrick
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2011, 57 (14) : E564 - E564
  • [50] Central Systolic Blood Pressure Estimated by an Osillometric Method or From an Invasive Brachial Pressure Waveform: A Direct Comparison
    Sung, Shih-Hsien
    Cheng, Hao-Min
    Wang, Kang-Ling
    Chen, Chen-Huan
    CIRCULATION, 2011, 124 (21)