Fast 3D in vivo swept-source optical coherence tomography using a two-axis MEMS scanning micromirror

被引:32
|
作者
Kumar, Karthik [1 ]
Condit, Jonathan C. [2 ]
McElroy, Austin [3 ]
Kemp, Nate J. [3 ]
Hoshino, Kazunori [2 ]
Milner, Thomas E. [2 ]
Zhang, Xiaojing [2 ]
机构
[1] Univ Texas Austin, Dept Elect & Comp Engn, Austin, TX 78712 USA
[2] Univ Texas Austin, Dept Biomed Engn, Austin, TX 78712 USA
[3] CardioSpectra Inc, San Antonio, TX 78229 USA
来源
关键词
optical coherence tomography (OCT); swept-source; MEMS; scanning micromirror; high-speed 3D imaging;
D O I
10.1088/1464-4258/10/4/044013
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We report on a fibre-based forward-imaging swept-source optical coherence tomography system using a high-reflectivity two-axis microelectromechanical scanning mirror for high-speed 3D in vivo visualization of cellular-scale architecture of biological specimens. The scanning micromirrors, based on electrostatic staggered vertical comb drive actuators, can provide +/- 9 degrees of optical deflection on both rotation axes and uniform reflectivity of greater than 90% over the range of imaging wavelengths (1260-1360 nm), allowing for imaging turbid samples with good signal-to-noise ratio. The wavelength-swept laser, scanning over 100 nm spectrum at 20 kHz rate, enables fast image acquisition at 10.2 million voxels s(-1) (for 3D imaging) or 40 frames s-1 (for 2D imaging with 500 transverse pixels per image) with 8.6 mu m axial resolution. Lateral resolution of 12.5 mu m over 3 mm field of view in each lateral direction is obtained using ZEMAX optical simulations for the lateral beam scanning system across the scanning angle range of the 500 mu m x 700 mu m micromirror. We successfully acquired en face and tomographic images of rigid structures (scanning micromirror), in vitro biological samples (onion peels and pickle slices) and in vivo images of human epidermis over 2 x 1 x 4 mm(3) imaging volume in real time at faster-than-video 2D frame rates. The results indicate that our system framework may be suitable for image-guided minimally invasive examination of various diseased tissues.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Swept-source common-path optical coherence tomography with a MEMS endoscopic imaging probe
    Duan, Can
    Wang, Donglin
    Zhou, Zhengwei
    Liang, Peng
    Samuelson, Sean R.
    Pozzi, Antonio
    Xie, Huikai
    OPTICAL COHERENCE TOMOGRAPHY AND COHERENCE DOMAIN OPTICAL METHODS IN BIOMEDICINE XVIII, 2014, 8934
  • [22] The Premacular Bursa's Shape Revealed In Vivo by Swept-Source Optical Coherence Tomography
    Schaal, Karen B.
    Pang, Claudine E.
    Carolina Pozzoni, M.
    Engelbert, Michael
    OPHTHALMOLOGY, 2014, 121 (05) : 1020 - 1028
  • [23] In Vivo Fascicle Bifurcation Imaging of Rat Sciatic Nerve Using Swept-Source Optical Coherence Tomography
    Choi, Daeyoung
    Lee, Jaeyul
    Jeon, Mansik
    Kim, Jeehyun
    IEEE ACCESS, 2018, 6 : 7713 - 7718
  • [24] Handheld ultrahigh speed swept source optical coherence tomography instrument using a MEMS scanning mirror
    Lu, Chen D.
    Kraus, Martin F.
    Potsaid, Benjamin
    Liu, Jonathan J.
    Choi, WooJhon
    Jayaraman, Vijaysekhar
    Cable, Alex E.
    Hornegger, Joachim
    Duker, Jay S.
    Fujimoto, James G.
    BIOMEDICAL OPTICS EXPRESS, 2014, 5 (01): : 293 - 311
  • [25] Evaluation of Ocular Diameter Parameters Using Swept-Source Optical Coherence Tomography
    Dong, Jing
    Yao, Jinhan
    Chang, Shuimiao
    Kanclerz, Piotr
    Khoramnia, Ramin
    Wang, Xiaogang
    MEDICINA-LITHUANIA, 2023, 59 (05):
  • [26] Defect inspection of actuator lenses using swept-source optical coherence tomography
    Jaeyul Lee
    Muhammad Faizan Shirazi
    Kibeom Park
    Mansik Jeon
    Jeehyun Kim
    Optical Review, 2018, 25 : 403 - 409
  • [27] Paper Insulation Aging Estimation Using Swept-Source Optical Coherence Tomography
    Mezgebo, Biniyam
    Sherif, Sherif S.
    Fernando, Namal
    Kordi, Behzad
    IEEE TRANSACTIONS ON DIELECTRICS AND ELECTRICAL INSULATION, 2022, 29 (01) : 30 - 37
  • [28] Images of Arterial Tissues Using Catheter Swept-Source Optical Coherence Tomography
    Mao, Youxin
    Flueraru, Costel
    Chang, Shoude
    Popescu, Dan P.
    Sowa, M. G.
    PHOTONICS NORTH 2010, 2010, 7750
  • [29] Laser Lens Size Measurement Using Swept-Source Optical Coherence Tomography
    Jia, Pingping
    Zhao, Hong
    Qin, Yuwei
    APPLIED SCIENCES-BASEL, 2020, 10 (14):
  • [30] En face features of polypoidalchoroidalvasculopathy using swept-source optical coherence tomography
    Won, Moon Sang
    Lee, Byung Ro
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (07)