Monte Carlo evaluation of hypothetical long axial field-of-view PET scanner using GE Discovery MI PET front-end architecture

被引:5
|
作者
Tiwari, Ashok [1 ,2 ]
Merrick, Michael [1 ,3 ]
Graves, Stephen A. [1 ,3 ,4 ]
Sunderland, John [1 ,2 ,4 ]
机构
[1] Univ Iowa Hosp & Clin, Dept Radiol, 200 Hawkins Dr, Iowa City, IA 52242 USA
[2] Univ Iowa, Dept Phys & Astron, Iowa City, IA USA
[3] Univ Iowa, Dept Biomed Engn, Iowa City, IA USA
[4] Univ Iowa Hosp & Clin, Dept Radiat Oncol, Iowa City, IA 52242 USA
关键词
Discovery MI; extended axial field-of-view; GATE Monte Carlo; NEMA NU 2-2018; PET; BODY; SIMULATION; GATE;
D O I
10.1002/mp.15422
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The development of total-body PET scanners is of growing interest in the PET community. Investigation into the imaging properties of a hypothetical extended axial field-of-view (AFOV) GE Healthcare SiPM-based Discovery MI (DMI) system architecture has not yet been performed. In this work, we assessed its potential as a whole-body scanner using Monte Carlo simulations. The aim of this work was to (1) develop and validate a Monte Carlo model of a four-ring scanner and (2) extend its AFOV up to 2 m to evaluate performance gain through NEMA-based evaluation. Methods The DMI four-ring geometry and its pulse digitization scheme were modeled within the GATE Monte Carlo platform using published literature. The GATE scanner model was validated by comparing results against published NEMA performance measurements. Following the validation of the four-ring model, the model was extended to simulate 8-, 20-, 30-, and 40-ring systems. Spatial resolution, sensitivity, NECR, and scatter fraction were characterized with modified NEMA NU-2 2018 standards; however, the image quality measurements were not acquired due to computational limitations. Spatial resolutions were simulated for all scanner ring configurations using point sources to examine the effects of parallax errors. NEMA count rates were estimated using a standard 70 cm scatter phantom and an extended version of scatter phantom of length 200 cm with (1-800) MBq of F-18 for all scanners. Sensitivity was evaluated using NEMA methods with a 70 cm standard and a 200 cm long line source. Results The average FWHM of the radial/tangential/axial spatial resolution reconstructed with filtered back-projection at 1 and 10 cm from the scanner center were 3.94/4.10/4.41 mm and 5.29/4.89/5.90 mm for the four-ring scanner. Sensitivity was determined to be 14.86 cps/kBq at the center of the FOV for the four-ring scanner using a 70 cm line source. Sensitivity enhancement up to 21-fold and 60-fold were observed for 1 and 2 m AFOV scanners compared to four-ring scanner using a 200 cm long line source. Spatial resolution simulations in a 2 m AFOV scanner suggest a maximum degradation of similar to 23.8% in the axial resolution compared to the four-ring scanner. However, the transverse resolution was found to be relatively constant when increasing the axial acceptance angle up to +/- 70 degrees. The peak NECR was 212.92 kcps at 22.70 kBq/ml with a scatter fraction of 38.9% for a four-ring scanner with a 70 cm scatter phantom. Comparison of peak NECR using the 200 cm long scatter phantom relative to the four-ring scanner resulted in a NECR gain of 15 for the 20-ring and 28 for the 40-ring geometry. Spatial resolution, sensitivity, and scatter fraction showed an agreement within similar to 7% compared with published measured values. Conclusions The four-ring DMI scanner simulation was successfully validated against published NEMA measurements. Sensitivity and NECR performance of extended 1 and 2 m AFOV scanners based upon the DMI architecture were subsequently simulated. Increases in sensitivity and count-rate performance are consistent with prior simulation studies utilizing extensions of the Siemens mCT architecture and published NEMA measurements with the uEXPLORER system.
引用
收藏
页码:1139 / 1152
页数:14
相关论文
共 50 条
  • [31] First-time rest-stress dynamic whole-body 82Rb-PET imaging using a long axial field-of-view PET/CT scanner
    Federico Caobelli
    Sigrid Seibel
    Korbinian Krieger
    Carola Bregenzer
    Marco Viscione
    Angela Filipa Silva Mendes
    Hasan Sari
    Lorenzo Mercolli
    Ali Afshar-Oromieh
    Axel Rominger
    European Journal of Nuclear Medicine and Molecular Imaging, 2023, 50 : 2219 - 2221
  • [32] First-time rest-stress dynamic whole-body 82Rb-PET imaging using a long axial field-of-view PET/CT scanner
    Caobelli, Federico
    Seibel, Sigrid
    Krieger, Korbinian
    Bregenzer, Carola
    Viscione, Marco
    Mendes, Angela Filipa Silva
    Sari, Hasan
    Mercolli, Lorenzo
    Afshar-Oromieh, Ali
    Rominger, Axel
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2023, 50 (07) : 2219 - 2221
  • [33] Physical Performance of a PET Scanner Prototype with Extended Axial Field of View using Sparse Detector Module Rings Configuration: A Monte Carlo simulation study
    Zein, S.
    Karakatsanis, N. A.
    Gupta, A.
    Osborne, J. R.
    Nehmeh, S. A.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2019, 46 (SUPPL 1) : S809 - S809
  • [34] Internal dosimetry study of [82Rb]Cl using a long axial field-of-view PET/CT
    Mercolli, Lorenzo
    Bregenzer, Carola
    Diemling, Markus
    Mingels, Clemens
    Rominger, Axel
    Sari, Hasan
    Seibel, Sigrid
    Sohlberg, Antti
    Viscione, Marco
    Caobelli, Federico
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2024, 51 (07) : 1869 - 1875
  • [35] System Architecture of a Fully Combined PET/CT Scanner using LabPET™ Electronics with an Upgraded Analog Front-end Optimized for PET and CT Counting Mode Operation
    Traore, Mohamed Seydou
    Thibaudeau, Christian
    Koua, Konin
    Tetrault, Marc-Andre
    Pepin, Catherine M.
    Cadorette, Jules
    Pratte, Jean-Francois
    Lecomte, Roger
    Fontaine, Rejean
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,
  • [36] Optimization of Y-90 Radioembolization Imaging for Post-Treatment Dosimetry on a Long Axial Field-of-View PET/CT Scanner
    Linder, Pia M.
    Lan, Wenhong
    Trautwein, Nils F.
    Brosch-Lenz, Julia
    von Beschwitz, Sebastian
    Kupferschlaeger, Juergen
    Reischl, Gerald
    Groezinger, Gerd
    Dittmann, Helmut
    la Fougere, Christian
    Schmidt, Fabian P.
    DIAGNOSTICS, 2023, 13 (22)
  • [37] First experiences with late acquisition [68Ga]GaPSMA-11 PET/CT using a long axial field-of-view PET/CT scanner for the diagnosis of recurrent prostate cancer
    Alberts, I.
    Prenosil, G.
    Mingels, C.
    Bohn, K.
    Viscione, M.
    Sari, H.
    Rominger, A.
    Afhsar-Oromieh, A.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2021, 48 (SUPPL 1) : S264 - S265
  • [38] Quantitative Investigation of Inter-Crystal Scatter and Penetration in the GE Discovery RX PET/CT Scanner using Monte Carlo Simulations
    Zeraatkar, Navid
    Ay, Mohammad R.
    Sarkar, Saeed
    Geramifar, Parham
    Rahmim, Arman
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 2403 - 2408
  • [39] NEMA Performance Evaluation of Long Axial Field-of-View Brain PET Insert with 7T MRI Compatibility
    Park, Haewook
    Lee, Seungeun
    Won, Jun Yeon
    Lee, Jae Sung
    JOURNAL OF NUCLEAR MEDICINE, 2020, 61
  • [40] Impact of reduced image noise on Deauville scores in patients with lymphoma scanned on a long-axial field-of-view PET/CT-scanner
    Korsholm, Kirsten
    Petersen, Nanna
    Dias, Andre
    Loft, Annika
    Andersen, Flemming
    Fischer, Barbara
    JOURNAL OF NUCLEAR MEDICINE, 2023, 64