A method to enhance spatial resolution of a 2D ion chamber array for quality control of MLC

被引:4
|
作者
Diaz Moreno, Rogelio [1 ]
Venencia, Daniel [2 ]
Garrigo, Edgardo [2 ]
Pipman, Yakov [3 ]
机构
[1] Inst Neurol & Neurocirugia, Dept Neurocirugia, Havana, Cuba
[2] Inst Privado Radioterapia, Cordoba, Argentina
[3] NSLIJ Hlth Syst, Dept Radiat Med, New Hyde Pk, NY USA
来源
关键词
multileaf collimator; quality control; 2D detector array; partial volume response of detector; INTENSITY-MODULATED RADIOTHERAPY; ASSURANCE; VERIFICATION; DELIVERY;
D O I
10.1120/jacmp.v12i4.3456
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This work introduces a new method for verifying MLC leaf positions with enough spatial resolution to replace film-based methods in performing QA tests. It is implemented on a 2D ion chamber array, and it is based on the principle of varying signal response of a volumetric detector to partial irradiation. A PTW 2D-ARRAY seven29 (PTW-729 2D) array was used to assess a Siemens OPTIFOCUS MLC. Partial volume response curves for chambers in the array were obtained by irradiating them with the leaves of the MLC, progressively covering varying portions of the chambers correlated with the leaf positions. The readings from the array's chambers are processed with an in-house program; it generates a reference response that translates readings into leaf positions. This principle allows discriminating errors in pairs of opposing leaves that could combine to cancel their detection with other tools. Patterns of leaf positions, similar to the Bayouth test but with different, purposefully introduced errors, were generated and used to test the effectiveness of the method. The same patterns were exposed on radiographic film and analyzed with the RIT software for validation. For four test patterns with a total of 100 errors of +/- 1 mm, +/- 2 mm and +/- 3 mm, all were correctly determined with the proposed method. The analysis of the same pattern with film using the Bayouth routine in the RIT software resulted in either somewhat low true positives combined with a large fraction of false positives, or a low true positive rate with a low false positive ratio, the results being significantly affected by the threshold selected for the analysis. This method provides an effective, easy to use tool for quantitative MLC QA assessment, with excellent spatial resolution. It can be easily applied to other 2D arrays, as long as they exhibit a partial volume detector response.
引用
收藏
页码:63 / 73
页数:11
相关论文
共 50 条
  • [31] Dosimetric Verification of the Enhanced Dynamic Wedges Using a 2D Ion Chamber Array
    Oh, S. A.
    Kim, S. K.
    Kang, M. K.
    Yea, J. W.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [32] Gantry-Angle Dependence of a 2D Ion Chamber Array for IMAT QA
    O'Daniel, J.
    Kishore, M.
    Das, S.
    Wu, Qj
    Yoo, S.
    Yin, F.
    MEDICAL PHYSICS, 2010, 37 (06) : 3237 - +
  • [33] Dosimetric Characteristics of 2D Ion Chamber Array Matrix for Verification Radiotherapy Treatments
    Sathiyan, S.
    Ravikumar, M.
    Supe, S.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [34] A FLUKA STUDY OF CROSSTALK BETWEEN MEASURING VOLUMES IN A 2D ION CHAMBER ARRAY
    Popovici, Maria Ana
    Vasilache, Radu Alin
    Buzatu, Daniela
    UNIVERSITY POLITEHNICA OF BUCHAREST SCIENTIFIC BULLETIN-SERIES A-APPLIED MATHEMATICS AND PHYSICS, 2018, 80 (04): : 301 - 308
  • [35] On-line quality assurance of rotational radiotherapy treatment delivery by means of a 2D ion chamber array and the Octavius phantom
    Van Esch, Ann
    Clermont, Christian
    Devillers, Magali
    Iori, Mauro
    Huyskens, Dominique P.
    MEDICAL PHYSICS, 2007, 34 (10) : 3825 - 3837
  • [36] Dosimetric verification of gated delivery of electron beams using a 2D ion chamber array
    Yoganathan, S. A.
    Das, K. J. Maria
    Raj, D. Gowtham
    Kumar, Shaleen
    JOURNAL OF MEDICAL PHYSICS, 2015, 40 (02) : 68 - 73
  • [37] Diagnostic X-Ray Beam Profiles Measured with a 2D Ion Chamber Array
    Li, Texin
    Gong, Jason
    Tong, Shidong
    Fountain, Betty
    Duffy, Edward
    MEDICAL PHYSICS, 2010, 37 (06)
  • [38] A 2D Ion Chamber Array Detector as a QA Device for Spot Scanning Proton Beams
    Arjomandy, B.
    Gillin, M.
    Zhu, X.
    Sahoo, N.
    Ciangaru, G.
    Bues, M.
    Poenisch, F.
    Ding, X.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [39] A study of enhanced dynamic wedge dosimetry using a 2D ion chamber array detector
    Kumar, S.
    Aparna, P.
    Anjana, P.
    Aswathi, C.
    Sitha, G.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2015, 14 (04) : 394 - 402
  • [40] Portal dosimetry for pretreatment verification of IMRT plan: a comparison with 2D ion chamber array
    Sharma, Dayananda Shamurailatpam
    Mhatre, Vaibav
    Heigrujam, Malhotra
    Talapatra, Kaustav
    Mallik, Suman
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2010, 11 (04): : 238 - 248