A method to enhance spatial resolution of a 2D ion chamber array for quality control of MLC

被引:4
|
作者
Diaz Moreno, Rogelio [1 ]
Venencia, Daniel [2 ]
Garrigo, Edgardo [2 ]
Pipman, Yakov [3 ]
机构
[1] Inst Neurol & Neurocirugia, Dept Neurocirugia, Havana, Cuba
[2] Inst Privado Radioterapia, Cordoba, Argentina
[3] NSLIJ Hlth Syst, Dept Radiat Med, New Hyde Pk, NY USA
来源
关键词
multileaf collimator; quality control; 2D detector array; partial volume response of detector; INTENSITY-MODULATED RADIOTHERAPY; ASSURANCE; VERIFICATION; DELIVERY;
D O I
10.1120/jacmp.v12i4.3456
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This work introduces a new method for verifying MLC leaf positions with enough spatial resolution to replace film-based methods in performing QA tests. It is implemented on a 2D ion chamber array, and it is based on the principle of varying signal response of a volumetric detector to partial irradiation. A PTW 2D-ARRAY seven29 (PTW-729 2D) array was used to assess a Siemens OPTIFOCUS MLC. Partial volume response curves for chambers in the array were obtained by irradiating them with the leaves of the MLC, progressively covering varying portions of the chambers correlated with the leaf positions. The readings from the array's chambers are processed with an in-house program; it generates a reference response that translates readings into leaf positions. This principle allows discriminating errors in pairs of opposing leaves that could combine to cancel their detection with other tools. Patterns of leaf positions, similar to the Bayouth test but with different, purposefully introduced errors, were generated and used to test the effectiveness of the method. The same patterns were exposed on radiographic film and analyzed with the RIT software for validation. For four test patterns with a total of 100 errors of +/- 1 mm, +/- 2 mm and +/- 3 mm, all were correctly determined with the proposed method. The analysis of the same pattern with film using the Bayouth routine in the RIT software resulted in either somewhat low true positives combined with a large fraction of false positives, or a low true positive rate with a low false positive ratio, the results being significantly affected by the threshold selected for the analysis. This method provides an effective, easy to use tool for quantitative MLC QA assessment, with excellent spatial resolution. It can be easily applied to other 2D arrays, as long as they exhibit a partial volume detector response.
引用
收藏
页码:63 / 73
页数:11
相关论文
共 50 条
  • [1] A method to enhance 2D ion chamber array patient specific quality assurance for IMRT
    Rogelio Manuel Diaz Moreno
    Daniel Venencia
    Edgardo Garrigo
    Yakov Pipman
    Australasian Physical & Engineering Sciences in Medicine, 2017, 40 : 145 - 151
  • [2] A method to enhance 2D ion chamber array patient specific quality assurance for IMRT
    Diaz Moreno, Rogelio Manuel
    Venencia, Daniel
    Garrigo, Edgardo
    Pipman, Yakov
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2017, 40 (01) : 145 - 151
  • [3] A New Method for Quality Control of Multileaf Collimators Using a 2D Ion Chamber Array
    Moreno, R. Diaz
    Venencia, C.
    Garrigo, E.
    Pipman, Y.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [4] Single vs double resolution performance of a 2D ion chamber array
    Sevillano, D.
    Garcia-Vicente, F.
    Castro, P.
    Ferrer, C.
    Perez, L.
    Torres, J.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S195 - S195
  • [5] Study of 2D ion chamber array for angular response and QA of dynamic MLC and pretreatment IMRT plans
    Bhardwaj, Anup Kumar
    Sharma, Suresh Chander
    Rana, Bhupendra
    Shukla, Arvind
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2009, 14 (03) : 89 - 94
  • [6] Leaves position accuracy and quality assurance of MLC with an ion chamber array
    Das, I
    Szeglin, S
    Bieda, M
    Cheng, C
    MEDICAL PHYSICS, 2002, 29 (06) : 1320 - 1320
  • [7] The use of an ion chamber array for routine dynamic MLC quality assurance
    Norrlinger, B
    Islam, M
    Ramaseshan, R
    Heaton, R
    MEDICAL PHYSICS, 2002, 29 (06) : 1273 - 1273
  • [8] Statistical analysis on 2D array of ion chamber performance
    Anvari, Akbar
    Aghamiri, Seyed
    Mahdavi, Seyed
    Alaei, Parham
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2015, 14 (02) : 194 - 201
  • [9] Overtesting of the MLC calibration with a 2D array
    Wenderoth, F.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2007, 183 : 76 - 77
  • [10] FEASIBILITY STUDY OF A 2D ION CHAMBER ARRAY AND THE MULTICUBE PHANTOM FOR QUALITY ASSURANCE OF RAPIDARC
    Bocanek, J.
    Gomola, I.
    Depuydt, T.
    Van den Heuvel, F.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S446 - S446