The Use of Laser Guidance Reduces Fluoroscopy Time for C-Arm Cone-Beam Computed Tomography-Guided Biopsies

被引:3
|
作者
Kroes, Maarten W. [1 ]
van Strijen, Marco J. L. [2 ]
Braak, Sicco J. [2 ]
Hoogeveen, Yvonne L. [1 ]
de Lange, Frank [1 ]
Kool, Leo J. Schultze [1 ]
机构
[1] Radboud Univ Nijmegen Med Ctr, Dept Radiol & Nucl Med, POB 9101, NL-6500 HB Nijmegen, Netherlands
[2] St Antonius Hosp, Dept Radiol, POB 2500, NL-3430 EM Nieuwegein, Netherlands
关键词
C-arm; Cone-beam CT-guidance; Percutaneous biopsy; Laser guidance; Fluoroscopy time; INTERVENTIONAL RADIOLOGY; NEEDLE INTERVENTIONS; NAVIGATION SYSTEM; ACCURACY; EXPERIENCE; PUNCTURES; PHANTOM; SUITE; STAFF;
D O I
10.1007/s00270-016-1345-y
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
When using laser guidance for cone-beam computed tomography (CBCT)-guided needle interventions, planned needle paths are visualized to the operator without the need to switch between entry- and progress-view during needle placement. The current study assesses the effect of laser guidance during CBCT-guided biopsies on fluoroscopy and procedure times. Prospective data from 15 CBCT-guided biopsies of 8-65 mm thoracic and abdominal lesions assisted by a ceiling-mounted laser guidance technique were compared to retrospective data of 36 performed CBCT-guided biopsies of lesions > 20 mm using the freehand technique. Fluoroscopy time, procedure time, and number of CBCT-scans were recorded. All data are presented as median (ranges). For biopsies using the freehand technique, more fluoroscopy time was necessary to guide the needle onto the target, 165 s (83-333 s) compared to 87 s (44-190 s) for laser guidance (p < 0.001). Procedure times were shorter for freehand-guided biopsies, 24 min versus 30 min for laser guidance (p < 0.001). The use of laser guidance during CBCT-guided biopsies significantly reduces fluoroscopy time.
引用
收藏
页码:1322 / 1326
页数:5
相关论文
共 50 条
  • [1] The Use of Laser Guidance Reduces Fluoroscopy Time for C-Arm Cone-Beam Computed Tomography-Guided Biopsies
    Maarten W. Kroes
    Marco J. L. van Strijen
    Sicco J. Braak
    Yvonne L. Hoogeveen
    Frank de Lange
    Leo J. Schultze Kool
    CardioVascular and Interventional Radiology, 2016, 39 : 1322 - 1326
  • [2] Laser Guidance in C-Arm Cone-Beam CT-Guided Radiofrequency Ablation of Osteoid Osteoma Reduces Fluoroscopy Time
    Kroes, Maarten W.
    Busser, Wendy M. H.
    Hoogeveen, Yvonne L.
    de Lange, Frank
    Kool, Leo J. Schultze
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2017, 40 (05) : 728 - 734
  • [3] Laser Guidance in C-Arm Cone-Beam CT-Guided Radiofrequency Ablation of Osteoid Osteoma Reduces Fluoroscopy Time
    Maarten W. Kroes
    Wendy M. H. Busser
    Yvonne L. Hoogeveen
    Frank de Lange
    Leo J. Schultze Kool
    CardioVascular and Interventional Radiology, 2017, 40 : 728 - 734
  • [4] Assessment of Needle Guidance Devices for Their Potential to Reduce Fluoroscopy Time and Operator Hand Dose during C-Arm Cone-Beam Computed Tomography-guided Needle Interventions
    Kroes, Maarten W.
    Busser, Wendy M. H.
    Futterer, Jurgen J.
    Arntz, Mark J.
    Janssen, Caroline M. M.
    Hoogeveen, Yvonne L.
    de Lange, Frank
    Kool, Leo J. Schultze
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2013, 24 (06) : 901 - 906
  • [5] C-arm cone-beam computed tomography-guided minimally invasive open excision of an osteoid osteoma undetectable on fluoroscopy: A case report
    Higuchi, Takashi
    Yamamoto, Norio
    Hayashi, Katsuhiro
    Takeuchi, Akihiko
    Abe, Kensaku
    Taniguchi, Yuta
    Araki, Yoshihiro
    Tsuchiya, Hiroyuki
    INTERNATIONAL JOURNAL OF SURGERY CASE REPORTS, 2019, 61 : 14 - 19
  • [6] Respiratory correlated cone-beam computed tomography on an isocentric C-arm
    Kriminski, S
    Mitschke, M
    Sorensen, S
    Wink, NM
    Chow, PE
    Tenn, S
    Solberg, TD
    PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (22): : 5263 - 5280
  • [7] Cone-beam computed tomography-guided transbronchial lung biopsies: Leading the way to precision
    Gonuguntla, Hari Kishan
    Vidyasagar, Preeti
    Mrigpuri, Parul
    LUNG INDIA, 2024, 41 (06) : 480 - 482
  • [8] Use of robotic C-arm cone-beam computed tomography in surgical stabilization of rib fractures
    Yen, Ming-Hong
    Liu, Tsu-Hao
    Liu, Jung -Sen
    Yim, Shelly
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2023, 54 (12):
  • [9] C-arm cone-beam computed tomography with stereotactic needle guidance for percutaneous adrenal biopsy: initial experience
    Jiao, Dechao
    Xie, Na
    Wu, Gang
    Ren, JianZhuang
    Han, Xinwei
    ACTA RADIOLOGICA, 2017, 58 (05) : 617 - 624
  • [10] Cone-Beam Computed Tomography-Guided Percutaneous Radiologic Gastrostomy
    Moehlenbruch, Markus
    Nelles, Michael
    Thomas, Daniel
    Willinek, Winfried
    Gerstner, Andreas
    Schild, Hans H.
    Wilhelm, Kai
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2010, 33 (02) : 315 - 320