Clinical introduction of a linac head-mounted 2D detector array based quality assurance system in head and neck IMRT

被引:31
|
作者
Korevaar, Erik W. [1 ]
Wauben, David J. L. [1 ]
van der Hulst, Peter C. [1 ]
Langendijk, Johannes A. [1 ]
Van't Veld, Aart A. [1 ]
机构
[1] Univ Groningen, Univ Med Ctr Groningen, Dept Radiat Oncol, NL-9700 RB Groningen, Netherlands
关键词
IMRT QA; 2D ionization chamber array; Dose reconstruction; Film dosimetry; Gamma index analysis; INTENSITY-MODULATED RADIOTHERAPY; PIXEL IONIZATION-CHAMBER; MONITOR UNIT SETTINGS; RADIATION-THERAPY; DOSE VERIFICATION; MULTILEAF COLLIMATION; SPATIAL-RESOLUTION; DYNAMIC-MODE; DOSIMETRY; DELIVERY;
D O I
10.1016/j.radonc.2011.09.007
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: IMRT QA is commonly performed in a phantom geometry but the clinical interpretation of the results in a 2D phantom plane is difficult. The main objective of our work is to move from film measurement based QA to 3D dose reconstruction in a patient CT scan. In principle, this could be achieved using a dose reconstruction method from 2D detector array measurements as available in the COMPASS system (IBA Dosimetry). The first step in the clinical introduction of this system instead of the currently used film QA procedures is to test the reliability of the dose reconstruction. In this paper we investigated the validation of the method in a homogeneous phantom with the film QA procedure as a reference. We tested whether COMPASS QA results correctly identified treatment plans that did or did not fulfil QA requirements in head and neck (H&N) IMRT. Materials and methods: A total number of 24 treatments were selected from an existing database with more than 100 film based H&N IMRT QA results. The QA results were classified as either good, just acceptable or clinically rejected (mean gamma index <0.4, 0.4-0.5 or >0.5, respectively with 3%/3 mm criteria). Film QA was repeated and compared to COMPASS QA with a MatriXX detector measurement performed on the same day. Results: Good agreement was found between COMPASS reconstructed dose and film measured dose in a phantom (mean gamma 0.83 +/- 0.09, 1SD with 1%/1 mm criteria, 0.33 +/- 0.04 with 3%/3 mm criteria). COMPASS QA results correlated well with film QA, identifying the same patients with less good QA results. Repeated measurements with film and COMPASS showed changes in delivery after a modified MLC calibration, also visible in a standard MLC check in COMPASS. The time required for QA reduced by half by using COMPASS instead of film. Conclusions: Agreement of COMPASS QA results with film based QA supports its clinical introduction for a phantom geometry. A standard MLC calibration check is sensitive to <1 mm changes that could be significant in H&N IMRT. These findings offer opportunities to further investigate the method based on a 2D detector array to 3D dose reconstruction in a patient anatomy. (C) 2011 Elsevier Ireland Ltd. All rights reserved. Radiotherapy and Oncology 100 (2011) 446-452
引用
收藏
页码:446 / 452
页数:7
相关论文
共 50 条
  • [31] A high resolution 2D array detector system for small-field MRI-linac applications
    Gargett, Maegan
    Oborn, Brad
    Alnaghy, Sarah J.
    Causer, Trent
    Petasecca, Marco
    Rosenfeld, Anatoly B.
    Metcalfe, Peter
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2018, 4 (03):
  • [32] 3D optical see-through head-mounted display based augmented reality system and its application
    Zhang, Zhenliang
    Weng, Dongdong
    Liu, Yue
    Li, Xiang
    INTERNATIONAL CONFERENCE ON OPTICAL AND PHOTONIC ENGINEERING (ICOPEN 2015), 2015, 9524
  • [33] Clinical study of 2D and 3D scan time reduction in head and neck cancer with BGO based PET CT
    Sanghera, B.
    Lowe, J.
    Lowe, G.
    Wellsted, D.
    Bammer, H.
    Wong, W.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2007, 34 : S340 - S340
  • [34] Evaluation of Angular Dependence Correction for 2D Array Detector Using for Quality Assurance of Volumetric Modulated Arc Therapy
    Karthikeyan, N.
    Ganesh, K. M.
    Vikraman, S.
    Shariff, M. H.
    MEDICAL PHYSICS, 2014, 41 (06) : 376 - 376
  • [35] RATIONALE FOR IMPLEMENTATION OF 3D VERIFICATION FOR IMRT IN HEAD AND NECK - DETERMINATION OF THE RELIABILITY OF 2D VS 3D VERIFICATION METHODS
    Rolfo, M.
    Wada, M.
    Height, R.
    Zupan, E.
    Handley, M.
    Anderson, N.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S487 - S487
  • [36] Methods for Analysis and Reporting the Patterns of Locoregional Failure in the Era of IMRT for Head and Neck Cancer: Deformable Image Registration-Based Quality Assurance Workflow
    Mohamed, A. S.
    Awan, M.
    Kocak, E.
    Beadle, B. M.
    Kantor, M. E.
    Gunn, G. B.
    Garden, A. S.
    Rosenthal, D. I.
    Fuller, C. D.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S569 - S570
  • [37] Feasibility Study for Using a 2D Array Detector for All Beam Measurements in Monthly Quality Assurance Procedure for a Uniform Scanning Proton Therapy System
    Rains, M.
    Mascia, A.
    Ramirez, E.
    Zheng, Y.
    Ding, X.
    Neuenschwander, D.
    Harris, B.
    MEDICAL PHYSICS, 2012, 39 (06) : 3732 - 3732
  • [38] See-Through Type 3D Head-Mounted Display-Based Surgical Microscope System for Microsurgery: A Feasibility Study
    Kim, Cheol-Hwan
    Ryu, Seon-Young
    Yoon, Ji-Young
    Lee, Hyoung-Kwon
    Choi, Nak-Gu
    Park, Il-Ho
    Choi, Hae-Young
    JMIR MHEALTH AND UHEALTH, 2019, 7 (03):
  • [39] A Modular Calibration Framework for 3D Interaction System Based on Optical See-Through Head-Mounted Displays in Augmented Reality
    Zhang, Zhenliang
    Weng, Dongdong
    Liu, Yue
    Wang, Yongtian
    2016 INTERNATIONAL CONFERENCE ON VIRTUAL REALITY AND VISUALIZATION (ICVRV 2016), 2016, : 393 - 400
  • [40] Scintillation imaging as a high-resolution, remote, versatile 2D detection system for MR-linac quality assurance
    Alexander, Daniel A.
    Zhang, Rongxiao
    Bruza, Petr
    Pogue, Brian W.
    Gladstone, David J.
    MEDICAL PHYSICS, 2020, 47 (09) : 3861 - 3869