Clinical introduction of a linac head-mounted 2D detector array based quality assurance system in head and neck IMRT

被引:31
|
作者
Korevaar, Erik W. [1 ]
Wauben, David J. L. [1 ]
van der Hulst, Peter C. [1 ]
Langendijk, Johannes A. [1 ]
Van't Veld, Aart A. [1 ]
机构
[1] Univ Groningen, Univ Med Ctr Groningen, Dept Radiat Oncol, NL-9700 RB Groningen, Netherlands
关键词
IMRT QA; 2D ionization chamber array; Dose reconstruction; Film dosimetry; Gamma index analysis; INTENSITY-MODULATED RADIOTHERAPY; PIXEL IONIZATION-CHAMBER; MONITOR UNIT SETTINGS; RADIATION-THERAPY; DOSE VERIFICATION; MULTILEAF COLLIMATION; SPATIAL-RESOLUTION; DYNAMIC-MODE; DOSIMETRY; DELIVERY;
D O I
10.1016/j.radonc.2011.09.007
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: IMRT QA is commonly performed in a phantom geometry but the clinical interpretation of the results in a 2D phantom plane is difficult. The main objective of our work is to move from film measurement based QA to 3D dose reconstruction in a patient CT scan. In principle, this could be achieved using a dose reconstruction method from 2D detector array measurements as available in the COMPASS system (IBA Dosimetry). The first step in the clinical introduction of this system instead of the currently used film QA procedures is to test the reliability of the dose reconstruction. In this paper we investigated the validation of the method in a homogeneous phantom with the film QA procedure as a reference. We tested whether COMPASS QA results correctly identified treatment plans that did or did not fulfil QA requirements in head and neck (H&N) IMRT. Materials and methods: A total number of 24 treatments were selected from an existing database with more than 100 film based H&N IMRT QA results. The QA results were classified as either good, just acceptable or clinically rejected (mean gamma index <0.4, 0.4-0.5 or >0.5, respectively with 3%/3 mm criteria). Film QA was repeated and compared to COMPASS QA with a MatriXX detector measurement performed on the same day. Results: Good agreement was found between COMPASS reconstructed dose and film measured dose in a phantom (mean gamma 0.83 +/- 0.09, 1SD with 1%/1 mm criteria, 0.33 +/- 0.04 with 3%/3 mm criteria). COMPASS QA results correlated well with film QA, identifying the same patients with less good QA results. Repeated measurements with film and COMPASS showed changes in delivery after a modified MLC calibration, also visible in a standard MLC check in COMPASS. The time required for QA reduced by half by using COMPASS instead of film. Conclusions: Agreement of COMPASS QA results with film based QA supports its clinical introduction for a phantom geometry. A standard MLC calibration check is sensitive to <1 mm changes that could be significant in H&N IMRT. These findings offer opportunities to further investigate the method based on a 2D detector array to 3D dose reconstruction in a patient anatomy. (C) 2011 Elsevier Ireland Ltd. All rights reserved. Radiotherapy and Oncology 100 (2011) 446-452
引用
收藏
页码:446 / 452
页数:7
相关论文
共 50 条
  • [21] Quality assurance of IMRT treatment plans for a 1.5 T MR-linac using a 2D ionization chamber array and a static solid phantom
    Moennich, David
    Winter, Jasmin
    Nachbar, Marcel
    Kuenzel, Luise
    Boeke, Simon
    Gani, Cihan
    Dohm, Oliver
    Zips, Daniel
    Thorwarth, Daniela
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (16):
  • [22] A study of the reliability of 2D vs. 3D verification methods for head and neck IMRT
    Rolfo, M.
    Wada, M.
    Height, R.
    Handley, M.
    Anderson, N.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01): : S414 - S415
  • [23] Correction for Directional Response of a 2D Ion Chamber Array for Quality Assurance of Arc Radiotherapy and IMRT
    Teo, B. K.
    Hristov, D.
    Mok, E.
    Luxton, G.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [24] Investigation of a 2D Detector Array for Quality Assurance with LA45 RaceTrack Microtron
    Xu, J.
    Liu, J.
    Panetta, J.
    Eldib, A.
    Lin, T.
    Price, R.
    Ma, C.
    MEDICAL PHYSICS, 2019, 46 (06) : E547 - E547
  • [25] Developing a Novel 2D electro-optic scanner for potential micro display and head-mounted display applications
    Wang, Wei-Chih
    Ghorapade, Vinayak
    Kalel, Sudarshan
    Muthuramalingam, Karthickraj
    SOFT MECHATRONICS AND WEARABLE SYSTEMS, 2024, 12948
  • [26] A Systematic Study On the Angular Dependence of 2D Ion Chamber Array for Accurate IMRT Plan Quality Assurance
    Xu, J.
    Liu, J.
    Veltchev, I.
    Ma, C.
    MEDICAL PHYSICS, 2020, 47 (06) : E691 - E691
  • [27] Development of Novel Heterogeneous Anthropomorphic Head and Neck Phantom Using 3D Printer for IMRT Patient Specific Quality Assurance
    Kadoya, N.
    Abe, K.
    Nemoto, H.
    Sato, K.
    Ieko, Y.
    Ito, K.
    Takayama, Y.
    Chiba, T.
    Dobashi, S.
    Takeda, K.
    Jingu, K.
    MEDICAL PHYSICS, 2018, 45 (06) : E595 - E595
  • [28] The clinical utility of phase-based respiratory gated PET imaging based on visual feedback with a head-mounted display system
    Mitsumoto, Takuya
    Minamimoto, Ryogo
    Sunaoka, Fumio
    Kishimoto, Seishi
    Inoue, Kazumasa
    Fukushi, Masahiro
    BRITISH JOURNAL OF RADIOLOGY, 2019, 92 (1098):
  • [29] The Use of Virtual Screens in Perceptual Studies on Electric Light: A Comparison Between 2D, Panoramic, Cardboard, and Head-Mounted Displays
    Rockcastle, Siobhan Francois
    Whalen, Galen Sollom-Brotherton
    Milgrim, Aaron
    Davis, Robert
    LEUKOS, 2024, 20 (03) : 275 - 290
  • [30] LINAC QUALITY ASSURANCE USING A 2D DIODE ARRAY DEVICE: IS IT REALLY NECESSARY THE 3D WATER PHANTOM YET?
    Rodriguez Garcia, A.
    Sanchez-Reyes Fernandez, A.
    Loscos, S.
    Vila, A.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S519 - S519