Pilot Study: Comparison of Artifacts of Brackets Using Dual-Energy CT

被引:0
|
作者
Kasslatter, Manuel [1 ]
Glodny, Bernhard [2 ]
Grams, Astrid [2 ]
Crismani, Adriano [1 ]
Schieffer, Lisa [1 ]
机构
[1] Med Univ Innsbruck, Univ Klin Kieferorthopadie, Anichstr 35-G1, A-6020 Innsbruck, Austria
[2] Med Univ Innsbruck, Univ Klin Radiol & Neuroradiol, Innsbruck, Austria
关键词
brackets; dual-energy CT; artifact reduction; monochromatic reconstruction; GOVERNING BODY DONATION; COMPUTED-TOMOGRAPHY; METAL ARTIFACTS; INTRACEREBRAL HEMORRHAGE; IODINATED CONTRAST; MULTIDETECTOR CT; REDUCTION; ANGIOGRAPHY; DIFFERENTIATION; OPTIMIZATION;
D O I
10.1055/a-0966-5829
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Studies on the use of monochrome (MC) imaging after implantation of metallic materials on the extremities or spine have shown that artifacts can be significantly reduced. Due to the high atomic numbers and high density of materials used in orthodontics, maxillofacial surgery and tooth preservation, hardening artifacts present a major problem in computed tomography (CT) imaging of the teeth, and jaw. The limitations of the image quality caused by the visualization of the jaw, and conceivable possibilities for improvement have not been adequately investigated. The aim of this work was to investigate the above artifacts and reduce them by creating monochromes in the jaw area. Dual energy CTs (DECTs) were performed on 10 corpse heads with and without orthodontic appliances. Subsequently, MCs of 80 to 160 kV were produced. The artifacts were assessed qualitatively and quantitatively using established methods. On anatomical structures such as the floor of the mouth and the soft palate, which are not in the same axial plane with high artefact producing objects of high atomic number, an increase of the Hunsfield units (HU) could be observed with increasing kV number of MC. On structures in the same plane with artifact-producing objects, a decrease of the HU could be observed with increasing kV number of MC. This study shows that it is also possible to reduce artifacts in the jaw area using MC. In all investigations no significantly increased artifacts were found in the setting with orthodontic appliances compared to the setting without orthodontic appliances.
引用
收藏
页码:17 / 28
页数:12
相关论文
共 50 条
  • [31] The pilot study of dual-energy CT gemstone spectral imaging on the image quality of hand tendons
    Deng, Kai
    Li, Wei
    Wang, Jun-jun
    Wang, Guang-li
    Shi, Hao
    Zhang, Cheng-qi
    CLINICAL IMAGING, 2013, 37 (05) : 930 - 933
  • [32] Coronary stent patency: Dual-energy multidetector CT assessment in a pilot study with anthropomorphic phantom
    Boll, Daniel T.
    Merkle, Elmar M.
    Paulson, Erik K.
    Fleiter, Thorsten R.
    RADIOLOGY, 2008, 247 (03) : 687 - 695
  • [33] Improving Prostatic Delineation Using Dual-Energy CT
    Gersh, J.
    Fried, D.
    MEDICAL PHYSICS, 2014, 41 (06) : 148 - 148
  • [34] Intermanufacturer Comparison of Dual-Energy CT Iodine Quantification and Monochromatic Attenuation: A Phantom Study
    Jacobsen, Megan C.
    Schellingerhout, Dawid
    Wood, Cayla A.
    Tamm, Eric P.
    Godoy, Myrna C.
    Sun, Jia
    Cody, Dianna D.
    RADIOLOGY, 2018, 287 (01) : 224 - 234
  • [35] Lower limit of iron quantification using dual-energy CT - a phantom study
    Jiang, Xia
    Hintenlang, David E.
    White, Richard D.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (01): : 299 - 307
  • [36] Differentiation of heroin and cocaine using dual-energy CT-an experimental study
    Grimm, Jochen
    Wudy, Ramona
    Ziegeler, Edvard
    Wirth, Stefan
    Uhl, Michael
    Reiser, Maximilian F.
    Scherr, Michael
    INTERNATIONAL JOURNAL OF LEGAL MEDICINE, 2014, 128 (03) : 475 - 482
  • [37] Motion Artifacts in Kidney Stone Imaging Using Dual-Energy CT. A Phantom Study Comparing Single-Source and Dual-Source Scanners
    Ibrahim, El-Sayed
    Pooley, Robert
    Cernigliaro, Joseph
    Bridges, Mellena
    Williams, James
    Haley, William
    2015 IEEE 12th International Symposium on Biomedical Imaging (ISBI), 2015, : 1159 - 1162
  • [38] Dual-Energy CT for Diagnostic CT Colonography
    Karcaaltincaba, Musturay
    Ozdeniz, Ilknur
    RADIOGRAPHICS, 2014, 34 (03) : 847 - 847
  • [39] Dual-energy CT quantitative imaging: a comparison study between twin-beam and dual-source CT scanners
    Almeida, Isabel P.
    Schyns, Lotte E. J. R.
    Ollers, Michel C.
    van Elmpt, Wouter
    Parodi, Katia
    Landry, Guillaume
    Verhaegen, Frank
    MEDICAL PHYSICS, 2017, 44 (01) : 171 - 179
  • [40] Application and advantages of monoenergetic reconstruction images for the reduction of metallic artifacts using dual-energy CT in knee and hip prostheses
    Nicola Magarelli
    Vincenzo De Santis
    Giammaria Marziali
    Amerigo Menghi
    Aaron Burrofato
    Luigi Pedone
    Dario Del Prete
    Roberto Iezzi
    Chiara de Waure
    Marianna D’andrea
    Antonio Leone
    Cesare Colosimo
    La radiologia medica, 2018, 123 : 593 - 600