PATIENT RADIATION DOSES IN CARDIAC COMPUTED TOMOGRAPHY: COMPARISON OF PUBLISHED RESULTS WITH PROSPECTIVE AND RETROSPECTIVE ACQUISITION

被引:11
|
作者
Efstathopoulos, E. P. [1 ]
Pantos, I. [1 ,2 ]
Thalassinou, S. [1 ,3 ]
Argentos, S. [1 ]
Kelekis, N. L. [1 ]
Zografos, T. [2 ]
Panayiotakis, G. [3 ]
Katritsis, D. G. [2 ]
机构
[1] Univ Athens, Gen Univ Hosp ATTIKON, Sch Med, Dept Radiol 2, GR-12462 Athens, Greece
[2] Athens Euroclin, Athens, Greece
[3] Univ Patras, Dept Med Phys, Patras, Greece
关键词
CT CORONARY-ANGIOGRAPHY; TUBE CURRENT MODULATION; IMAGE QUALITY; DUAL-SOURCE; DIAGNOSTIC-ACCURACY; HEART-RATE; SPIRAL CT; EXPOSURE; REDUCTION; ARTERIES;
D O I
10.1093/rpd/ncq602
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Prospective ECG triggering has the potential of reducing radiation exposure while maintaining diagnostic accuracy of cardiac computed tomography (CT). The aim of this study is to review patient radiation doses associated with coronary artery calcium scoring (CACS) and CT coronary angiography (CTCA) and to compare results between prospective and retrospective acquisition schemes. Patient radiation doses from CACS and CTCA were extracted from 67 relevant studies. Mean effective dose for CACS and CTCA with prospective ECG triggering is significantly lower than retrospective acquisition, 0.9 +/- 0.4 vs. 3.1 +/- 1.4 mSv, p < 0.001, and 3.4 +/- 1.4 vs. 11.1 +/- 5.4 mSv, p < 0.001, respectively. In both cardiac CT examinations, application of dose modulation techniques result in significantly lower doses in retrospective schemes, however, even with dose modulation, retrospective acquisition is associated with significantly higher doses than prospective acquisition. The number of slices acquired per rotation and the number of X-ray sources of the CT scanner (single or dual source) do not have a significant effect on patient dose.
引用
收藏
页码:83 / 91
页数:9
相关论文
共 50 条
  • [31] PATIENT DOSES USING MULTIDETECTOR COMPUTED TOMOGRAPHY SCANNERS IN KENYA
    Korir, G. K.
    Wambani, J. S.
    Korir, I. K.
    RADIATION PROTECTION DOSIMETRY, 2012, 151 (02) : 267 - 271
  • [32] Helical prospective ECG-gating in cardiac computed tomography: radiation dose and image quality
    Tony DeFrance
    Eric Dubois
    Dan Gebow
    Alex Ramirez
    Florian Wolf
    Gudrun M. Feuchtner
    The International Journal of Cardiovascular Imaging, 2010, 26 : 99 - 107
  • [33] Comparison of Diagnostic Accuracy and Radiation Dose Between Prospective Triggering and Retrospective Gated Coronary Angiography by Dual-Source Computed Tomography
    Lu, Bin
    Lu, Jin-Guo
    Sun, Ming-Li
    Hou, Zhi-Hui
    Chen, Xiong-Biao
    Tang, Xiang
    Wu, Run-Ze
    Johnson, Laura
    Qiao, Shu-Bin
    Yang, Yue-Jin
    Jiang, Shi-Liang
    AMERICAN JOURNAL OF CARDIOLOGY, 2011, 107 (09): : 1278 - 1284
  • [34] Diagnostic Accuracy of Coronary Computed Tomography Angiography A Comparison Between Prospective and Retrospective Electrocardiogram Triggering
    Pontone, Gianluca
    Andreini, Daniele
    Bartorelli, Antonio L.
    Cortinovis, Sarah
    Mushtaq, Saima
    Bertella, Erika
    Annoni, Andrea
    Formenti, Alberto
    Nobili, Enrica
    Trabattoni, Daniela
    Montorsi, Piero
    Ballerini, Giovanni
    Agostoni, Piergiuseppe
    Pepi, Mauro
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2009, 54 (04) : 346 - 355
  • [35] Image Quality and Radiation Exposure Comparison of a Double High-Pitch Acquisition for Coronary Computed Tomography Angiography Versus Standard Retrospective Spiral Acquisition in Patients With Atrial Fibrillation
    Elias dos Prazeres, Carlos Eduardo
    Magalhaes, Tiago Augusto
    de Castro Carneiro, Adriano Camargo
    Cury, Roberto Caldeira
    Moreira, Valeria de Melo
    Silva Matsumoto Bello, Juliana Hiromi
    Rochitte, Carlos Eduardo
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2018, 42 (01) : 45 - 53
  • [36] Radiation Exposure From Cardiac Computed Tomography REPLY
    Ho, Kheng Thye
    Chua, Kia-Chong
    Klotz, Ernst
    Panknin, Christoph
    JACC-CARDIOVASCULAR IMAGING, 2010, 3 (11) : 1202 - 1202
  • [37] Cardiac computed tomography and radiation: balancing benefit and risk
    Almeida, Ana G.
    REVISTA PORTUGUESA DE CARDIOLOGIA, 2010, 29 (11) : 1677 - 1682
  • [38] Radiation Dose at Cardiac Computed Tomography Facts and Fiction
    Huda, Walter
    Rowlett, W. Taylor
    Schoepf, U. Joseph
    JOURNAL OF THORACIC IMAGING, 2010, 25 (03) : 204 - 212
  • [39] Computed tomography radiation doses for common computed tomography examinations: a nationwide dose survey in United Arab Emirates
    Abuzaid, Mohamed M.
    Elshami, Wiam
    Tekin, Huseyin Ozan
    Ghonim, Hatem
    Shawki, Mona
    Salama, Dina H.
    INSIGHTS INTO IMAGING, 2020, 11 (01)
  • [40] Computed tomography radiation doses for common computed tomography examinations: a nationwide dose survey in United Arab Emirates
    Mohamed M. Abuzaid
    Wiam Elshami
    Huseyin Ozan Tekin
    Hatem Ghonim
    Mona Shawki
    Dina H. Salama
    Insights into Imaging, 11