PATIENT RADIATION DOSES IN CARDIAC COMPUTED TOMOGRAPHY: COMPARISON OF PUBLISHED RESULTS WITH PROSPECTIVE AND RETROSPECTIVE ACQUISITION

被引:11
|
作者
Efstathopoulos, E. P. [1 ]
Pantos, I. [1 ,2 ]
Thalassinou, S. [1 ,3 ]
Argentos, S. [1 ]
Kelekis, N. L. [1 ]
Zografos, T. [2 ]
Panayiotakis, G. [3 ]
Katritsis, D. G. [2 ]
机构
[1] Univ Athens, Gen Univ Hosp ATTIKON, Sch Med, Dept Radiol 2, GR-12462 Athens, Greece
[2] Athens Euroclin, Athens, Greece
[3] Univ Patras, Dept Med Phys, Patras, Greece
关键词
CT CORONARY-ANGIOGRAPHY; TUBE CURRENT MODULATION; IMAGE QUALITY; DUAL-SOURCE; DIAGNOSTIC-ACCURACY; HEART-RATE; SPIRAL CT; EXPOSURE; REDUCTION; ARTERIES;
D O I
10.1093/rpd/ncq602
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Prospective ECG triggering has the potential of reducing radiation exposure while maintaining diagnostic accuracy of cardiac computed tomography (CT). The aim of this study is to review patient radiation doses associated with coronary artery calcium scoring (CACS) and CT coronary angiography (CTCA) and to compare results between prospective and retrospective acquisition schemes. Patient radiation doses from CACS and CTCA were extracted from 67 relevant studies. Mean effective dose for CACS and CTCA with prospective ECG triggering is significantly lower than retrospective acquisition, 0.9 +/- 0.4 vs. 3.1 +/- 1.4 mSv, p < 0.001, and 3.4 +/- 1.4 vs. 11.1 +/- 5.4 mSv, p < 0.001, respectively. In both cardiac CT examinations, application of dose modulation techniques result in significantly lower doses in retrospective schemes, however, even with dose modulation, retrospective acquisition is associated with significantly higher doses than prospective acquisition. The number of slices acquired per rotation and the number of X-ray sources of the CT scanner (single or dual source) do not have a significant effect on patient dose.
引用
收藏
页码:83 / 91
页数:9
相关论文
共 50 条
  • [1] Image quality and radiation dose in 256-slice cardiac computed tomography: Comparison of prospective versus retrospective image acquisition protocols
    Hosch, Waldemar
    Heye, Tobias
    Schulz, Felix
    Lehrke, Stephanie
    Schlieter, Martin
    Giannitsis, Evangelos
    Kauczor, Hans-Ulrich
    Katus, Hugo A.
    Korosoglou, Grigorios
    EUROPEAN JOURNAL OF RADIOLOGY, 2011, 80 (01) : 127 - 135
  • [2] Radiation Doses Associated With Cardiac Computed Tomography Angiography
    Daher, Iyad N.
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2009, 301 (22): : 2324 - 2324
  • [3] Radiation Doses Associated With Cardiac Computed Tomography Angiography Reply
    Hausleiter, Joerg
    Gerber, Thomas C.
    Achenbach, Stephan
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2009, 301 (22): : 2325 - 2325
  • [4] Adult patient radiation doses from non-cardiac CT examinations: a review of published results
    Pantos, I.
    Thalassinou, S.
    Argentos, S.
    Kelekis, N. L.
    Panayiotakis, G.
    Efstathopoulos, E. P.
    BRITISH JOURNAL OF RADIOLOGY, 2011, 84 (1000): : 293 - 303
  • [5] Radiation doses in cerebral perfusion computed tomography: patient and phantom study
    Arandjic, D.
    Bonutti, F.
    Biasizzo, E.
    Ciraj-Bjelac, O.
    Floreani, M.
    Giustizieri, M.
    Iaiza, F.
    Inkoom, S.
    Tommasini, G.
    Padovani, R.
    RADIATION PROTECTION DOSIMETRY, 2013, 154 (04) : 459 - 464
  • [6] RADIATION DOSES IN ADULT COMPUTED TOMOGRAPHY PRACTICE IN SERBIA: INITIAL RESULTS
    Arandjic, Danijela
    Ciraj-Bjelac, Olivera
    Hadnadjev, Darka
    Stojanovic, Sanja
    Bozovic, Predrag
    Ceklic, Sandra
    Lazarevic, Djordje
    RADIATION PROTECTION DOSIMETRY, 2014, 162 (1-2) : 135 - 138
  • [7] Radiation Doses in Cardiovascular Computed Tomography
    Kedzierski, Bartlomiej
    Macek, Piotr
    Dziadkowiec-Macek, Barbara
    Truszkiewicz, Krystian
    Poreba, Rafal
    Gac, Pawel
    LIFE-BASEL, 2023, 13 (04):
  • [8] Radiation doses in computed tomography - The increasing doses of radiation need to be controlled
    Rehani, MM
    Berry, M
    BRITISH MEDICAL JOURNAL, 2000, 320 (7235): : 593 - 594
  • [9] Cardiac computed tomography and radiation
    Sezenon, Burak
    Sahinarslan, Asife
    ANATOLIAN JOURNAL OF CARDIOLOGY, 2013, 13 (04): : 374 - 378
  • [10] Current status of patient radiation doses from computed tomography examinations in Tanzania
    Ngaile, J. E.
    Msaki, P.
    Kazema, R.
    RADIATION PROTECTION DOSIMETRY, 2006, 121 (02) : 128 - 135