MCNP Simulation of Radiation Doses Distributions in a Water Phantoms Simulating Interventional Radiology Patients

被引:0
|
作者
He, Wenjun [1 ]
Mah, Eugene [2 ]
Huda, Walter [2 ]
Selby, Bayne [2 ]
Yao, Hai [1 ]
机构
[1] Clemson Univ, Clemson MUSC Bioengn Program, Dept Bioengn, Charleston, SC 29425 USA
[2] Med Univ South Carolina, Dept Radiol & Radiol Sci, Charleston, SC 29425 USA
关键词
Interventional radiology; radiation dose; Monte Carlo; Air kerma; primary radiation; scatter radiation; energy imparted; ALARA; FLUOROSCOPICALLY GUIDED PROCEDURES; MONTE-CARLO; SKIN INJURIES; PART I; RAD-IR; TRANSMISSION; SCATTER;
D O I
10.1117/12.878271
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Purpose: To investigate the dose distributions in water cylinders simulating patients undergoing Interventional Radiological examinations. Method: The irradiation geometry consisted of an x-ray source, dose-area-product chamber, and image intensifier as currently used in Interventional Radiology. Water cylinders of diameters ranging between 17 and 30 cm were used to simulate patients weighing between 20 and 90 kg. X-ray spectra data with peak x-ray tube voltages ranging from 60 to 120 kV were generated using XCOMP3R. Radiation dose distributions inside the water cylinder (D-w) were obtained using MCNP5. The depth dose distribution along the x-ray beam central axis was normalized to free-in-air air kerma (AK) that is incident on the phantom. Scattered radiation within the water cylinders but outside the directly irradiated region was normalized to the dose at the edge of the radiation field. The total absorbed energy to the directly irradiated volume (E-p) and indirectly irradiated volume (E-s) were also determined and investigated as a function of x-ray tube voltage and phantom size. Results: At 80 kV, the average D-w/AK near the x-ray entrance point was 1.3. The ratio of D-w near the entrance point to Dw near the exit point increased from similar to 26 for the 17 cm water cylinder to similar to 290 for the 30 cm water cylinder. At 80 kV, the relative dose for a 17 cm water cylinder fell to 0.1% at 49 cm away from the central ray of the x-ray beam. For a 30 cm water cylinder, the relative dose fell to 0.1% at 53 cm away from the central ray of the x-ray beam. At a fixed x-ray tube voltage of 80 kV, increasing the water cylinder diameter from 17 to 30 cm increased the E-s/(E-p+E-s) ratio by about 50%. At a fixed water cylinder diameter of 24 cm, increasing the tube voltage from 60 kV to 120 kV increased the E-s/(E-p+E-s) ratio by about 12%. The absorbed energy from scattered radiation was between 20-30% of the total energy absorbed by the water cylinder, and was affected more by patient size than x-ray beam energy. Conclusion: MCNP offers a powerful tool to study the absorption and transmission of x-ray energy in phantoms that can be designed to represent patients undergoing Interventional Radiological procedures. This ability will permit a systematic investigation of the relationship between patient dose and diagnostic image quality, and thereby keep patient doses As Low As Reasonably Achievable (ALARA).
引用
收藏
页数:10
相关论文
共 50 条
  • [31] Mixed Reality Visualization of Radiation Dose for Health Professionals and Patients in Interventional Radiology
    Takata, Takeshi
    Nakabayashi, Susumu
    Kondo, Hiroshi
    Yamamoto, Masayoshi
    Furui, Shigeru
    Shiraishi, Kenshiro
    Kobayashi, Takenori
    Oba, Hiroshi
    Okamoto, Takahide
    Kotoku, Jun'ichi
    JOURNAL OF MEDICAL SYSTEMS, 2021, 45 (04)
  • [32] Mixed Reality Visualization of Radiation Dose for Health Professionals and Patients in Interventional Radiology
    Takeshi Takata
    Susumu Nakabayashi
    Hiroshi Kondo
    Masayoshi Yamamoto
    Shigeru Furui
    Kenshiro Shiraishi
    Takenori Kobayashi
    Hiroshi Oba
    Takahide Okamoto
    Jun’ichi Kotoku
    Journal of Medical Systems, 2021, 45
  • [33] Radiation doses in interventional radiology procedures: The RAD-IR Study - Part I: Overall measures of dose
    Miller, DL
    Balter, S
    Cole, PE
    Lu, HT
    Schueler, BA
    Geisinger, M
    Berenstein, A
    Albert, R
    Georgia, JD
    Noonan, PT
    Cardella, JF
    George, JS
    Russell, EJ
    Malisch, TW
    Vogelzang, RL
    Miller, GL
    Anderson, J
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2003, 14 (06) : 711 - 727
  • [34] Radiation Dose Management of Pregnant Patients, Pregnant Staff and Paediatric Patients Diagnostic and Interventional Radiology
    Srivastava, Ajai K.
    JOURNAL OF MEDICAL PHYSICS, 2022, 47 (02) : 212 - 214
  • [35] Calculation of Radiation Doses to Patients Undergoing Interventional Procedures by Monte Carlo Simulations
    Nam, H. W.
    Lee, M. J.
    Na, H. J.
    Kim, K. P.
    MEDICAL PHYSICS, 2020, 47 (06) : E517 - E517
  • [36] A study on patients' radiation doses from interventional cardiac procedures in Tabriz, Iran
    Mesbahi, Asghar
    Aslanabadi, Naser
    RADIATION PROTECTION DOSIMETRY, 2008, 132 (04) : 375 - 380
  • [37] Evaluation of patients' and occupational radiation risk dose during conventional and interventional radiology procedures
    Tamam, Nissren
    Salah, H.
    Almogren, Kholoud S.
    Mahgoub, Omer
    Saeed, Mohammed Khalil
    Abdullah, Yousef
    Tai, Duong Thanh
    Omer, Hiba
    Sulieman, Abdelmoneim
    Bradley, D. A.
    RADIATION PHYSICS AND CHEMISTRY, 2023, 207
  • [38] Efficiency of radiation protection equipment in interventional radiology: a systematic Monte Carlo study of eye lens and whole body doses
    Koukorava, C.
    Farah, J.
    Struelens, L.
    Clairand, I.
    Donadille, L.
    Vanhavere, F.
    Dimitriou, P.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2014, 34 (03) : 509 - 528
  • [39] Radiation protection of patients in diagnostic and interventional radiology in Asian countries: Impact of an IAEA project
    Rehani, Madan M.
    Ciraj-Bjelac, Olivera
    Al-Naemi, Huda M.
    Al-Suwaidi, Jamila Salem
    El-Nachef, Leila
    Khosravi, Hamid Reza
    Kharita, Mohammad Hassan
    Muthuvelu, Pirunthavany
    Pallewatte, Aruna S.
    Juan, Bayani Cruz San
    Shaaban, Mohamed
    Zaman, Areesha
    EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (10) : E982 - E989
  • [40] Assessment of medical radiation exposure to patients and ambient doses in several diagnostic radiology departments
    Sulieman, A.
    Elhadi, T.
    Babikir, E.
    Alkhorayef, M.
    Alnaaimi, M.
    Alduaij, M.
    Bradley, D. A.
    RADIATION PHYSICS AND CHEMISTRY, 2017, 140 : 202 - 206