Image quality and radiation dose in planar imaging - Image quality figure of merits from the CDRAD phantom

被引:9
|
作者
Konst, Bente [1 ]
Weedon-Fekjaer, Harald [2 ]
Bath, Magnus [3 ,4 ]
机构
[1] Vestfold Hosp Trust, Dept Radiol, Tonsberg, Norway
[2] Oslo Univ Hosp, Oslo Ctr Biostat & Epidemiol, Res Support Serv, Oslo, Norway
[3] Sahlgrens Univ Hosp, Dept Med Phys & Biomed Engn, Gothenburg, Sweden
[4] Univ Gothenburg, Sahlgrenska Acad, Inst Clin Sci, Dept Radiat Phys, Gothenburg, Sweden
来源
关键词
CD-curve; contrast detail phantom; figure of merit; IQF; planar imaging; precision; LOW-CONTRAST DETAIL; DIGITAL CHEST RADIOGRAPHY; DETECTOR; SYSTEMS; PERFORMANCE; RESOLUTION;
D O I
10.1002/acm2.12649
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A contrast-detail phantom such as CDRAD is frequently used for quality assurance, optimization of image quality, and several other purposes. However, it is often used without considering the uncertainty of the results. The aim of this study was to assess two figure of merits (FOM) originating from CDRAD regarding the variations of the FOMs by dose utilized to create the x-ray image. The probability of overlapping (assessing an image acquired at a lower dose as better than an image acquired at a higher dose) was determined. Methods: The CDRAD phantom located underneath 12, 20, and 26 cm PMMA was imaged 16 times at five dose levels using an x-ray system with a flat-panel detector. All images were analyzed by CDRAD Analyser, version 1.1, which calculated the FOM inverse image quality figure (IQF(inv)) and gave contrast detail curves for each image. Inherent properties of the CDRAD phantom were used to derive a new FOM h, which describes the size of the hole with the same diameter and depth that is just visible. Data were analyzed using heteroscedastic regression of mean and variance by dose. To ease interpretation, probabilities for overlaps were calculated assuming normal distribution, with associated bootstrap confidence intervals. Results: The proportion of total variability in IQF(inv), explained by the dose (R-2), was 91%, 85%, and 93% for 12, 20, and 26 cm PMMA. Corresponding results for h were 91%, 89%, and 95%. The overlap probability for different mAs levels was 1% for 0.8 vs 1.2 mAs, 5% for 1.2 vs 1.6 mAs, 10% for 1.6 vs 2.0 mAs, and 10% for 2.0 mAs vs 2.5 mAs for 12 cm PMMA. For 20 cm PMMA, it was 0.5% for 10 vs 16 mAs, 13% for 16 vs 20 mAs, 14% for 20 vs 25 mAs, and 14% for 25 vs 32 mAs. For 26 cm PMMA, the probability varied from 0% to 6% for various mAs levels. Even though the estimated probability for overlap was small, the 95% confidence interval (CI) showed relatively large uncertainties. For 12 cm PMMA, the associated CI for 0.8 vs 1.2 mAs was 0.1-3.2%, and the CI for 1.2 vs 1.6 mAs was 2.1-7.8%. Conclusions: Inverse image quality figure and h are about equally related to dose level. The FOM h, which describes the size of a hole that should be seen in the image, may be a more intuitive FOM than IQFinv. However, considering the probabilities for overlap and their confidence intervals, the FOMs deduced from the CDRAD phantom are not sensitive to dose. Hence, CDRAD may not be an optimal phantom to differentiate between images acquired at different dose levels.
引用
收藏
页码:151 / 159
页数:9
相关论文
共 50 条
  • [41] Effect of additional filtration on radiation dose and image quality
    Gai, Xinghui
    Yang, Dewu
    Li, Xunru
    Sui, Yan
    Liu, Kang
    Wang, Junying
    Zhao, Mengjuan
    2022 IEEE 6TH ADVANCED INFORMATION TECHNOLOGY, ELECTRONIC AND AUTOMATION CONTROL CONFERENCE (IAEAC), 2022, : 1807 - 1811
  • [42] Radiation dose and image quality in head CT examinations
    Huda, W
    Chang, J
    Lieberman, KA
    Roskopf, ML
    RADIOLOGY, 2001, 221 : 160 - 160
  • [43] Radiation dose and image quality for paediatric interventional cardiology
    Vano, E.
    Ubeda, C.
    Leyton, F.
    Miranda, P.
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (15): : 4049 - 4062
  • [44] Radiation Dose Image Quality Optomization in Dental Implantalogy
    Al-Humairi, Ahmed
    Zheng, Xiaoming
    Ip, Ho Leung
    El Masoud, Bilal
    PROCEEDINGS OF THE 2016 INTERNATIONAL CONFERENCE ON APPLIED MATHEMATICS, SIMULATION AND MODELLING, 2016, 41 : 403 - 406
  • [45] Image quality and radiation dose assessment of dental CBCT
    Huang, Jinfeng
    Chen, Xiaozhao
    Wang, Hong
    ASIA-PACIFIC JOURNAL OF CLINICAL ONCOLOGY, 2022, 18 : 42 - 42
  • [46] Radiation Dose and Image Quality in Pediatric Neck CT
    Tipnis, S. V.
    Rieter, W. J.
    Patel, D.
    Stalcup, S. T.
    Matheus, M. G.
    Spampinato, M. V.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2019, 40 (06) : 1067 - 1073
  • [47] CT protocols: balancing radiation dose and image quality
    Richard Fotter
    Pediatric Radiology, 2011, 41 : 215 - 216
  • [48] Neonatal radiography: Radiation dose in association with image quality
    Dougeni, E
    Karatza, A
    Delis, H
    Kalogeropoulou, C
    Panayiotakis, G
    Mantagos, S
    PEDIATRIC RESEARCH, 2005, 58 (02) : 387 - 387
  • [49] IMAGE QUALITY FACTORS AND RADIATION-DOSE IN MAMMOGRAPHY
    HAUS, AG
    ERICKSON, L
    JOURNAL OF IMAGING TECHNOLOGY, 1984, 10 (01): : 29 - 35
  • [50] Nationwide survey of fluoroscopy: Radiation dose and image quality
    Suleiman, OH
    Conway, BJ
    Quinn, P
    Antonsen, RG
    Rueter, FG
    Slayton, RJ
    Spelic, DC
    RADIOLOGY, 1997, 203 (02) : 471 - 476