Nationwide survey of fluoroscopy: Radiation dose and image quality

被引:18
|
作者
Suleiman, OH
Conway, BJ
Quinn, P
Antonsen, RG
Rueter, FG
Slayton, RJ
Spelic, DC
机构
[1] Ctr. Devices and Radiological Hlth., U.S. Food and Drug Administration, Rockville, MD
[2] U.S. Food and Drug Administration, Ctr. Devices and Radiological Hlth., Div. Mammography Qual. Radiat. P., Rockville, MD 20850
关键词
fluoroscopy; technology; gastrointestinal tract; radiography; phantoms; radiations; exposure to patients and personnel; measurement; test objects;
D O I
10.1148/radiology.203.2.9114107
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To determine the average abdominal entrance air kerma, low-contrast sensitivity, and spatial resolution in upper gastrointestinal tract fluoroscopy in the United States. MATERIALS AND METHODS: A random sample of fluoroscopic facilities was selected to be surveyed for the Nationwide Evaluation of X-ray Trends program. Measurements were performed by using a newly developed fluoroscopic phantom. The surveys were conducted by state radiation control personnel. RESULTS: Average air kerma rates 1 cm above the tabletop, free in air, were 43 mGy/min (n = 340). The rate increased to 64 mGy/min when a 1.6-mm-thick copper filter, which simulated the use of barium contrast medium, was added to increase attenuation. The average entrance air kerma, free in air, for radiographs was 3.4 mGy, and an average of 12 radiographs were obtained per examination. Of 352 facilities surveyed, 306 (87%) were able to resolve wire mesh with 20 or more lines per inch. Of 339 facilities for which percentage contrast could be calculated, 192 (57%) had minimum percentage contrast values of 4% or more. CONCLUSION: Spatial resolution for fluoroscopy is adequate for most of the facilities surveyed, but a substantial proportion of facilities could not visualize low-contrast test objects, which strongly suggests image quality problems.
引用
收藏
页码:471 / 476
页数:6
相关论文
共 50 条
  • [1] Nationwide survey of fluoroscopy: Radiation dose and image quality. (vol 203, pg 471, 1997)
    Suleiman, OH
    Conway, BJ
    Quinn, P
    Antonsen, RG
    Rueter, FG
    Slayton, RJ
    RADIOLOGY, 1998, 207 (01) : 278 - 278
  • [2] A multihospital survey of radiation exposure and image quality in pediatric fluoroscopy
    P. H. Brown
    P. J. Silberberg
    R. D. Thomas
    J. L. Strife
    R. B. Towbin
    Pediatric Radiology, 2000, 30 : 236 - 242
  • [3] A multihospital survey of radiation exposure and image quality in pediatric fluoroscopy
    Brown, PH
    Silberberg, PJ
    Thomas, RD
    Strife, JL
    Towbin, RB
    PEDIATRIC RADIOLOGY, 2000, 30 (04) : 236 - 242
  • [4] Radiation dose and image quality of CT fluoroscopy with partial exposure mode
    Takiguchi, Keisuke
    Urikura, Atsushi
    Yoshida, Tsukasa
    Nakaya, Yoshihiro
    Endo, Masahiro
    Aramaki, Takeshi
    DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2020, 26 (04) : 333 - 338
  • [5] Medical imaging using ionizing radiation: Optimization of dose and image quality in fluoroscopy
    Jones, A. Kyle
    Balter, Stephen
    Rauch, Phillip
    Wagner, Louis K.
    MEDICAL PHYSICS, 2014, 41 (01)
  • [6] CT Image Quality and Radiation Dose Survey in Taiwan
    Tsai, H. Y.
    Liu, H. L.
    Chen, C. C.
    Hwang, Y. S.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [7] Dose reduction and image quality in pulsed fluoroscopy
    Vetter, S
    Faulkner, K
    Strecker, EP
    Busch, HP
    RADIATION PROTECTION DOSIMETRY, 1998, 80 (1-3) : 299 - 301
  • [8] A SURVEY IN PORTUGUESE X-RAY FLUOROSCOPY EQUIPMENT: DOSE RATES AND IMAGE QUALITY
    Teixeira, Nuno G.
    Coutinho, Guilherme M.
    Trindade, Hugo R.
    Cunha, Gilda R.
    Carvoeiras, Pedro P.
    RADIATION PROTECTION DOSIMETRY, 2015, 164 (03) : 354 - 360
  • [9] Variability in Radiation Dose and Image Quality: A Comparison Across Fluoroscopy-system Vendors and Generations of Equipment
    Hill, Kevin D.
    Carboni, Michael P.
    Doyle, Thomas P.
    Idriss, Salim
    Janssen, Dana
    Nicholson, George
    Sathanandam, Shyam
    Mann, Steve
    Fleming, Gregory
    CIRCULATION, 2016, 134
  • [10] Variability in Radiation Dose and Image Quality: A Comparison Across Fluoroscopy-system Vendors and Generations of Equipment
    Hill, Kevin D.
    Carboni, Michael P.
    Doyle, Thomas P.
    Idriss, Salim
    Janssen, Dana
    Nicholson, George
    Sathanandam, Shyam
    Mann, Steve
    Fleming, Gregory
    CIRCULATION, 2016, 134