Nationwide survey of fluoroscopy: Radiation dose and image quality

被引:18
|
作者
Suleiman, OH
Conway, BJ
Quinn, P
Antonsen, RG
Rueter, FG
Slayton, RJ
Spelic, DC
机构
[1] Ctr. Devices and Radiological Hlth., U.S. Food and Drug Administration, Rockville, MD
[2] U.S. Food and Drug Administration, Ctr. Devices and Radiological Hlth., Div. Mammography Qual. Radiat. P., Rockville, MD 20850
关键词
fluoroscopy; technology; gastrointestinal tract; radiography; phantoms; radiations; exposure to patients and personnel; measurement; test objects;
D O I
10.1148/radiology.203.2.9114107
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To determine the average abdominal entrance air kerma, low-contrast sensitivity, and spatial resolution in upper gastrointestinal tract fluoroscopy in the United States. MATERIALS AND METHODS: A random sample of fluoroscopic facilities was selected to be surveyed for the Nationwide Evaluation of X-ray Trends program. Measurements were performed by using a newly developed fluoroscopic phantom. The surveys were conducted by state radiation control personnel. RESULTS: Average air kerma rates 1 cm above the tabletop, free in air, were 43 mGy/min (n = 340). The rate increased to 64 mGy/min when a 1.6-mm-thick copper filter, which simulated the use of barium contrast medium, was added to increase attenuation. The average entrance air kerma, free in air, for radiographs was 3.4 mGy, and an average of 12 radiographs were obtained per examination. Of 352 facilities surveyed, 306 (87%) were able to resolve wire mesh with 20 or more lines per inch. Of 339 facilities for which percentage contrast could be calculated, 192 (57%) had minimum percentage contrast values of 4% or more. CONCLUSION: Spatial resolution for fluoroscopy is adequate for most of the facilities surveyed, but a substantial proportion of facilities could not visualize low-contrast test objects, which strongly suggests image quality problems.
引用
收藏
页码:471 / 476
页数:6
相关论文
共 50 条
  • [21] Variability in radiation dose and image quality: A comparison across fluoroscopy-system vendors, generations of equipment and institutions
    Hill, Kevin D.
    Mann, Steve D.
    Carboni, Michael P.
    Doyle, Thomas P.
    Idriss, Salim F.
    Janssen, Dana F.
    Nicholson, George T.
    Sathanandam, Shyam
    Fleming, Greg A.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2018, 92 (07) : E471 - E477
  • [22] Regional survey of image quality and radiation dose in computed tomography examinations in Saudi Arabia
    Saeed, M. K.
    Alzoubi, A. S.
    Al-Qahtani, J. M.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2014, 37 (02) : 279 - 283
  • [23] Regional survey of image quality and radiation dose in computed tomography examinations in Saudi Arabia
    M. K. Saeed
    A. S. Alzoubi
    J. M. Al-Qahtani
    Australasian Physical & Engineering Sciences in Medicine, 2014, 37 : 279 - 283
  • [24] National Survey of Radiation Dose and Image Quality in Adult CT Head Scans in Taiwan
    Lin, Chung-Jung
    Mok, Greta S. P.
    Tsai, Mang-Fen
    Tsai, Wei-Ta
    Yang, Bang-Hung
    Tu, Chun-Yuan
    Wu, Tung-Hsin
    PLOS ONE, 2015, 10 (06):
  • [25] Optimal Scan Parameters for CT Fluoroscopy in Lung Interventional Radiologic Procedures: Relationship between Radiation Dose and Image Quality
    Yamao, Yoshikazu
    Yamakado, Koichiro
    Takaki, Haruyuki
    Yamada, Tomomi
    Murashima, Shuichi
    Uraki, Junji
    Kodama, Hiroshi
    Nagasawa, Naoki
    Takeda, Kan
    RADIOLOGY, 2010, 255 (01) : 233 - 241
  • [26] Tradeoffs in CT image quality and radiation dose
    McNitt-Gray, M
    MEDICAL PHYSICS, 2003, 30 (06) : 1410 - 1410
  • [27] INFLUENCE OF RADIATION-DOSE ON IMAGE QUALITY
    REICHMANN, S
    ASTRAND, K
    ACTA RADIOLOGICA-DIAGNOSIS, 1979, 20 (02): : 393 - 399
  • [28] A factorial experiment on image quality and radiation dose
    Norrman, E
    Persliden, J
    RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) : 246 - 252
  • [29] Radiation dose and image quality in computed tomography
    Prokop, M
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2002, 174 (05): : 631 - 636
  • [30] Radiation dose and image quality in diagnostic radiology
    Geijer, H
    ACTA RADIOLOGICA, 2002, 43 : 4 - 43