X-ray phase contrast imaging using single absorption coded aperture

被引:5
|
作者
Krejci, Frantisek [1 ,2 ]
Jakubek, Jan [1 ]
Kroupa, Martin [1 ]
机构
[1] Czech Tech Univ, Inst Expt & Appl Phys, CZ-12800 Prague 2, Czech Republic
[2] Czech Tech Univ, Fac Biomed Engn, Kladno 27201, Czech Republic
关键词
X-ray imaging; Phase contrast; Coded aperture; Phase retrivial; Medipix; RESOLUTION;
D O I
10.1016/j.nima.2010.06.161
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A new method for X-ray phase contrast imaging is proposed in which only one absorption grating (acting in fact as a coded aperture) and a table-top nano-focus X-ray source are used. This method is based on very precise determination of position of the projected pattern downstreaming from the aperture with and without the inspected object directly from the image of the pattern. The technique is successfully demonstrated on the table-top setup equipped with Medipix family detectors and a nano-focus X-ray source (source-to-detector distance < 1 m). (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:S181 / S184
页数:4
相关论文
共 50 条
  • [21] X-Ray Imaging with Phase Contrast
    Bonse, Ulrich
    Beckmann, Felix
    INTERNATIONAL CONFERENCE ON ADVANCED PHASE MEASUREMENT METHODS IN OPTICS AN IMAGING, 2010, 1236 : 189 - +
  • [22] Phase contrast X-ray imaging
    Weon, Byung Mook
    Je, Jung Ho
    Hwu, Yeukuang
    Margaritondo, Giorgio
    INTERNATIONAL JOURNAL OF NANOTECHNOLOGY, 2006, 3 (2-3) : 280 - 297
  • [23] X-ray phase contrast imaging
    Xiong, SS
    Chen, JW
    Gao, HY
    Xie, HL
    Li, RX
    Xu, ZZ
    ISTM/2003: 5TH INTERNATIONAL SYMPOSIUM ON TEST AND MEASUREMENT, VOLS 1-6, CONFERENCE PROCEEDINGS, 2003, : 3715 - 3718
  • [24] Phase contrast X-ray imaging
    Biomedical Imaging Center, Department of Materials Science and Engineering, Pohang University of Science and Technology, Pohang, 790-784, Korea, Republic of
    不详
    不详
    不详
    不详
    不详
    不详
    不详
    不详
    不详
    不详
    Int. J. Nanotechnol., 2006, 2-3 (280-297):
  • [25] StaticCodeCT: single coded aperture tensorial X-ray CT
    Cuadros, Angela P.
    Ma, Xu
    Restrepo, Carlos M.
    Arce, Gonzalo R.
    OPTICS EXPRESS, 2021, 29 (13) : 20558 - 20576
  • [26] A comparison of absorption and phase contrast for X-ray imaging of biological cells
    Nave, Colin
    JOURNAL OF SYNCHROTRON RADIATION, 2018, 25 : 1490 - 1504
  • [27] MAGNIFIED HARD X-RAY IMAGING AND TOMOGRAPHY IN ABSORPTION AND PHASE CONTRAST
    Schroer, C. G.
    Benner, B.
    Guenzler, T. F.
    Kuhlmann, M.
    Lengeler, B.
    Rau, C.
    Weitkamp, T.
    Snigirev, A.
    Snigireva, I.
    ACTA CRYSTALLOGRAPHICA A-FOUNDATION AND ADVANCES, 2002, 58 : C51 - C51
  • [28] X-ray phase contrast imaging and noise evaluation using a single phase grating interferometer
    Rizzi, J.
    Mercere, P.
    Idir, M.
    Da Silva, P.
    Vincent, G.
    Primot, Jerome
    OPTICS EXPRESS, 2013, 21 (14): : 17340 - 17351
  • [29] A coded-aperture based method allowing non-interferometric phase contrast imaging with incoherent X-ray sources
    Olivo, A.
    Ignatyev, K.
    Munro, P. R. T.
    Speller, R. D.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 648 : S28 - S31
  • [30] Modelling of a novel x-ray phase contrast imaging technique based on coded apertures
    Olivo, A.
    Speller, R.