Non-invasive monitoring method for lower-leg compartment syndrome using a wireless sensor system and finite element analysis

被引:3
|
作者
Chang, Fuh-Yu [1 ]
Teng, Ping-Tun [1 ]
Chen, Liang-Chun [1 ]
机构
[1] Taiwan Univ Sci & Technol, Dept Mech Engn, 43,Sect 4,Keelung Rd, Taipei 106, Taiwan
关键词
Compartment syndrome; non-invasive pressure detect; pressure sensor; finite element method; smartphone;
D O I
10.1177/0954411920981243
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this study, a non-invasive pressure monitoring system that is portable and convenient was designed for detecting compartment syndrome. The system combines a wireless module and smartphone, which aids in the achievement of mHealth objectives, specifically, the continuous monitoring of the compartment pressure in patients. A compartment syndrome detecting method using a wireless sensor system and finite element analysis is developed and verified with an in vitro lower-leg model by rapid prototyping. The sensor system is designed to measure a five point pressure variation from the outside of the lower leg and transmit the data to a smartphone via Bluetooth. The analysis model based on the finite element method is employed to calculate the change of pressure and volume inside the four compartments of the lower leg. The in vitro experimental results show that the non-invasive detecting method can monitor the compartment pressure and provide a warning for the occurrence of compartment syndrome if the compartment pressure is higher than 30 mmHg. Furthermore, the theoretical simulation of the real lower leg shows similar trends to those of the in vitro experiments and can promptly detect the occurrence of compartment syndrome. Measured pressure values exceeding 6.3, 2.7, and 2.8 kPa for the three sensors contacting the outside centers of the superficial posterior, anterior, and lateral compartments, respectively, can indicate that each compartment contains a pressure higher than 30 mmHg. These results can provide a warning for the risk of compartment syndrome of each compartment. In addition, the measured values from the three sensors contacting the superficial posterior compartment at the outside center, close to the tibia, and close to the lateral compartment exceeding 1.8, 0.7, and 0.7 kPa, respectively, can indicate the risk of deep posterior compartment syndrome.
引用
收藏
页码:346 / 356
页数:11
相关论文
共 50 条
  • [21] Imaging and finite element analysis: A methodology for non-invasive characterization of aortic tissue
    Flamini, Vittoria
    Creane, Arthur P.
    Kerskens, Christian M.
    Lally, Caitriona
    MEDICAL ENGINEERING & PHYSICS, 2015, 37 (01) : 48 - 54
  • [22] Finite element analysis of the liver subjected to non-invasive indirect mechanical loading
    Shaabeth, Samar Ali
    Kadhem, Amina
    Lafta, Hassanain Ali
    INTERNATIONAL JOURNAL OF COMPUTATIONAL SCIENCE AND ENGINEERING, 2025, 28 (02) : 142 - 150
  • [23] Heart patient health monitoring system using invasive and non-invasive measurement
    Mastoi, Qurat-ul-Ain
    Alqahtani, Ali
    Almakdi, Sultan
    Sulaiman, Adel
    Rajab, Adel
    Shaikh, Asadullah
    Alqhtani, Samar M.
    SCIENTIFIC REPORTS, 2024, 14 (01):
  • [24] Non-Invasive Sensor Technology for the Development of a Dairy Cattle Health Monitoring System
    Awasthi, Amruta
    Awasthi, Anshul
    Riordan, Daniel
    Walsh, Joseph
    COMPUTERS, 2016, 5 (04)
  • [25] Design And Development Of An Innovative Sensor System For Non-Invasive Monitoring Of Athletic Performances
    Sabatini, Anna
    Zompanti, Alessandro
    Grasso, Simone
    Gianfelici, Antonio
    Di Castro, Andrea
    Donatucci, Bruno
    Pennazza, Giorgio
    Santonico, Marco
    2019 IEEE INTERNATIONAL WORKSHOP ON METROLOGY FOR INDUSTRY 4.0 AND INTERNET OF THINGS (METROIND4.0&IOT), 2019, : 162 - 166
  • [26] Evaluation of Mesh and Sensor Resolution for Finite Element Modeling of Non-Invasive Fetal ECG Signals
    Keenan, Emerson
    Karmakar, Chandan
    Brownfoot, Fiona C.
    Palaniswami, Marimuthu
    2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), 2021, : 4134 - 4138
  • [27] Phonomyography as a non-invasive continuous monitoring technique for muscle ischemia in an experimental model of acute compartment syndrome
    Martinez, Adriana P.
    Moser, Thomas P.
    Saran, Neil
    Paquet, Marilene
    Hemmerling, Thomas
    Berry, Greg K.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2017, 48 (11): : 2411 - 2416
  • [28] Epidural effect on maternal hemodynamics using a novel, wireless, PPG-based sensor for continuous non-invasive monitoring
    Atzmon, Yuval
    Ben-Ishai, Efrat
    Hallak, Mordechai
    Gabbay-Benziv, Rinat
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 2020, 222 (01) : S482 - S483
  • [29] A Non-Invasive Flexible Glucose Monitoring Sensor Using a Broadband Reject Filter
    Bteich, Moussa
    Hanna, Jessica
    Costantine, Joseph
    Kanj, Rouwaida
    Tawk, Youssef
    Ramadan, Ali H.
    Eid, Assaad A.
    IEEE JOURNAL OF ELECTROMAGNETICS RF AND MICROWAVES IN MEDICINE AND BIOLOGY, 2021, 5 (02): : 139 - 147
  • [30] Non-invasive measurement of muscle compartment elasticity in lower limbs to determine acute compartment syndrome: Clinical results with pressure related ultrasound
    Sellei, R. M.
    Wollnitz, J.
    Reinhardt, N.
    de la Fuente, M.
    Radermacher, K.
    Weber, C.
    Kobbe, P.
    Hildebrand, F.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2020, 51 (02): : 301 - 306