Total Hip Arthroplasty Using Hybrid Ceramic Bearing: A Minimum 10-Year Followup Study

被引:5
|
作者
Yoo, Jun-Il [1 ]
Ha, Yong-Chan [2 ]
Kim, Dong-Hyun [2 ]
Lee, Young-Kyun [3 ]
Koo, Kyung-Hoi [3 ]
机构
[1] Gyeongsang Natl Univ Hosp, Dept Orthopaed Surg, Jinju, South Korea
[2] Chung Ang Univ, Dept Orthopaed Surg, Coll Med, 102 Heukseok Ro, Seoul 156755, South Korea
[3] Seoul Natl Univ, Dept Orthopaed Surg, Bundang Hosp, Bundang, South Korea
关键词
Cementless; delta liner; large diameter femoral head; survival; total hip arthroplasty; HEAD; ARTICULATION; FRACTURES; FIXATION;
D O I
10.4103/ortho.IJOrtho_416_18
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Contemporary ceramic articulation in total hip arthroplasty (THA) has shown successful clinical and radiologic outcomes at mid-to-long term. Delta ceramic, which includes 24% zirconia, has been developed to reduce the rate of ceramic fracture. However, the phase transformation of the zirconia that might cause surface roughness, ceramic wear, and subsequent osteolysis remains a concern. The purpose of this study was to evaluate clinicoradiological outcomes, ceramic-related complications, and survivorship of THA with the use of hybrid ceramic bearing; BIOLOX delta liner-on-BIOLOX forte head articulation at a minimum followup of 10 years. Materials and Methods: One hundred cementless THAs (85 patients) were performed with the use of 36-mm hybrid ceramic bearing. The thickness of the delta ceramic liner was 4-6 mm. Four patients (4 hips) were lost and six patients (6 hips) died within 10 years after the operation. Thus, the remaining 75 patients (90 hips) were followed up for 10-12 years. There were 43 men and 32 women. The mean age of these patients at the index operation was 58 years (21-78 years). Results: There were no fractures of ceramic liner or head. There was no measurable ceramic wear, and neither pelvic nor femoral osteolysis was identified. All acetabular and femoral components were bone-ingrown. Three (3/90, 3.3%) patients noted noise of clicking sound. The mean Harris hip score was 87 points (range 72-99 points) at the time of final followup. THA survivorship with revision surgery as an endpoint was 97.8% (95% confidence interval: 91.4%-99.8%). Conclusions: The long term result of cementless THA with the use of a large BIOLOX forte head on the BIOLOX delta liner showed an encouraging result. Wear or osteolysis associated with composite of 24% zirconia in the fourth-generation ceramic was not observed at a minimum followup of 10 years.
引用
收藏
页码:637 / 640
页数:4
相关论文
共 50 条
  • [1] Total Hip Arthroplasty Using Hybrid Ceramic Bearing: A Minimum 10-Year Followup Study
    Jun-Il Yoo
    Yong-Chan Ha
    Dong-Hyun Kim
    Young-Kyun Lee
    Kyung-Hoi Koo
    [J]. Indian Journal of Orthopaedics, 2019, 53 : 637 - 640
  • [2] Outcomes of Ceramic- On-Ceramic Bearing Total Hip Arthroplasty: A Minimum 10-Year Follow-Up Study
    Alshammari, Malek O.
    de Petrillo, Gianni
    Epure, Laura M.
    Huk, Olga L.
    Zukor, David J.
    Antoniou, John
    [J]. JOURNAL OF ARTHROPLASTY, 2023, 38 (07): : S146 - S151
  • [3] Omnifit cementless total hip arthroplasty - A 10-year average followup
    Hellman, EJ
    Capello, WN
    Feinberg, JR
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1999, (364) : 164 - 174
  • [4] Reply to Letter to the Editor: Regarding "Outcomes of Ceramic-On-Ceramic Bearing Total Hip Arthroplasty: A Minimum 10-Year Follow-Up Study"
    Alshammari, Malek O.
    De Petrillo, Gianni
    Epure, Laura M.
    Huk, Olga L.
    Zukor, David J.
    Antoniou, John
    [J]. JOURNAL OF ARTHROPLASTY, 2023, 38 (10): : E40 - E41
  • [5] Hybrid total hip arthroplasty - 7- to 10-year results
    Berger, RA
    Kull, LR
    Rosenberg, AG
    Galante, JO
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1996, (333) : 134 - 146
  • [6] 3rd generation ceramic-on-ceramic cementless total hip arthroplasty: a minimum 10-year follow-up study
    Lau, Yeong J.
    Sarmah, Shashank
    Witt, Johan D.
    [J]. HIP INTERNATIONAL, 2018, 28 (02) : 133 - 138
  • [7] Minimum 10-Year Results of Modular Metal-On-Metal Total Hip Arthroplasty
    Wakabayashi, Hiroki
    Hasegawa, Masahiro
    Naito, Yohei
    Tone, Shine
    Sudo, Akihiro
    [J]. JOURNAL OF CLINICAL MEDICINE, 2022, 11 (21)
  • [8] Minimum 10-year results of cementless total hip arthroplasty in patients with rheumatoid arthritis
    Haraguchi, Akihisa
    Nakashima, Yasuharu
    Miyahara, Hisaaki
    Esaki, Yukio
    Okazaki, Ken
    Fukushi, Jun-ichi
    Hirata, Go
    Ikemura, Satoshi
    Kamura, Satoshi
    Sakuraba, Koji
    Fujimura, Kenjiro
    Akasaki, Yukio
    Yamada, Hisakata
    [J]. MODERN RHEUMATOLOGY, 2017, 27 (04) : 598 - 604
  • [9] POLYETHYLENE SOCKETS AND ALUMINA CERAMIC HEADS IN CEMENTED TOTAL HIP-ARTHROPLASTY - A 10-YEAR STUDY
    SUGANO, N
    NISHII, T
    NAKATA, K
    MASUHARA, K
    TAKAOKA, K
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1995, 77B (04): : 548 - 556
  • [10] Total hip arthroplasty in patients with a vascular necrosis of the femoral head - A 2- to 10-year followup
    Garino, JP
    Steinberg, ME
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1997, (334) : 108 - 115