Experimental validation of a synthetic aperture imaging technique using a therapeutic random phased array is described, demonstrating the dual nature of imaging and therapy of such an array. The transducer is capable of generating both continuous wave high intensity beams for ablating the tumor and low intensity ultrasound pulses to image the target area. Pulse-echo data is collected from the elements of the phased array to obtain B-mode images of the targets. Since therapeutic arrays are optimized for therapy only with concave apertures having low f-number and large directive elements often coarsely sampled, imaging can not be performed using conventional beamforming. We show that synthetic aperture imaging is capable of processing the acquired RF data to obtain images of the field of interest. Simulations were performed to compare different synthetic aperture imaging techniques to identify the best algorithm in terms of spatial resolution. Experimental validation was performed using a 1 MHz, 256-elements, spherical random phased array with 130 mm radius of curvature. The array was integrated with a research ultrasound scanner via custom connectors to acquire raw RF data for variety of targets. Imaging was implemented using synthetic aperture beamforming to produce images of a rib phantom and ex vivo ribs. The array was shown to resolve spherical targets within 15 mm of either side of the axis in the focal plane and obtain 3D images of the rib phantom up to 40 mm of either side of the central axis and at a depth of 3-9 cm from the array surface. The lateral and axial full width half maximum was 1.15 mm and 2.75 mm, respectively. This study was undertaken to emphasize that both therapy and image guidance with a therapeutic random phased array is possible and such a system has the potential to address some major limitations in the existing high intensity focused ultrasound (HIFU) systems. The 3D images obtained with a therapeutic array can be used to identify and locate strong scattering objects aiding to image guidance and treatment planning of the HIFU procedure.
引用
收藏
页数:16
相关论文
共 50 条
  • [21] Large Matrix Array Aperture for 3D Vascular Imaging Capture
    Colas, Quorentin
    Bantignies, Claire
    Perroteau, Marie
    Porcher, Nicolas
    Vassal, Steeven
    Guerif, Benjamin
    Kim, Taehoon
    Bosch, Johan G.
    de Jong, Nico
    Verweij, Martin D.
    Pertijs, Michiel A. P.
    Ferin, Guillaume
    Flesch, Martin
    2022 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IEEE IUS), 2022,
  • [22] A joint sparse recovery algorithm for coprime adjacent array synthetic aperture radar 3D sparse imaging
    Tian, Bokun
    Zhang, Xiaoling
    Tang, Xinxin
    Wei, Shunjun
    Shi, Jun
    INTERNATIONAL JOURNAL OF REMOTE SENSING, 2021, 42 (17) : 6560 - 6580
  • [23] 3D Terahertz Synthetic Aperture Imaging of Objects with Arbitrary Boundaries
    Kniffin, G. P.
    Zurk, L. M.
    Schecklman, S.
    Henry, S. C.
    TERAHERTZ EMITTERS, RECEIVERS, AND APPLICATIONS IV, 2013, 8846
  • [24] Ground Based Synthetic Aperture Radar with 3D Imaging Capability
    Pieraccini, Massimiliano
    Rojhani, Neda
    Miccinesi, Lapo
    2018 15TH EUROPEAN RADAR CONFERENCE (EURAD), 2018, : 206 - 209
  • [25] 3D Imaging Millimeter Wave Circular Synthetic Aperture Radar
    Zhang, Renyuan
    Cao, Siyang
    SENSORS, 2017, 17 (06)
  • [26] Phase error correction for synthetic-aperture phased-array imaging systems
    Fienup, JR
    IMAGE RECONSTRUCTION FROM INCOMPLETE DATA, 2000, 4123 : 47 - 55
  • [27] Method of improving 3D imaging resolution of APD array based on optical phased array
    Wang, Shuai
    Sun, Huayan
    Zhao, Yanzhong
    Zeng, Hairui
    Liu, Tianjian
    Hongwai yu Jiguang Gongcheng/Infrared and Laser Engineering, 2019, 48 (04):
  • [28] Phased array 3D MR spectroscopic imaging of the brain at 7 T
    Xu, Duan
    Cunningham, Charles H.
    Chen, Albert P.
    Li, Yan
    Kelley, Douglas A. C.
    Mukherjee, Pratik
    Pauly, John M.
    Nelson, Sarah J.
    Vigneron, Daniel B.
    MAGNETIC RESONANCE IMAGING, 2008, 26 (09) : 1201 - 1206
  • [29] 3-D Imaging Using Synthetic Aperture Radar With a Frequency Diverse Array
    Li, Jingjing
    Liao, Kefei
    Ouyang, Shan
    IEEE GEOSCIENCE AND REMOTE SENSING LETTERS, 2021, 18 (05) : 846 - 850
  • [30] Development and Application of spherical Ultrasonic Array probe for 3D Imaging
    Chang, Jun-jie
    Zhong, Hai-ying
    PROCEEDINGS OF 2019 FAR EAST NDT NEW TECHNOLOGY & APPLICATION FORUM (FENDT), 2019, : 178 - 182