Time-Resolved Interventional Cardiac C-arm Cone-Beam CT: An Application of the PICCS Algorithm

被引:70
|
作者
Chen, Guang-Hong [1 ]
Theriault-Lauzier, Pascal [1 ]
Tang, Jie [1 ]
Nett, Brian [1 ]
Leng, Shuai [1 ]
Zambelli, Joseph [1 ]
Qi, Zhihua [1 ]
Bevins, Nicholas [1 ]
Raval, Amish [2 ]
Reeder, Scott [3 ]
Rowley, Howard [3 ]
机构
[1] Univ Wisconsin, Dept Med Phys, Madison, WI 53792 USA
[2] Univ Wisconsin, Dept Med, Madison, WI 53792 USA
[3] Univ Wisconsin, Dept Radiol, Madison, WI 53792 USA
基金
加拿大自然科学与工程研究理事会; 美国国家卫生研究院;
关键词
Cardiac cone-beam CT; compressed sensing; high temporal resolution; interventional imaging; DUAL-ENERGY CT; COMPUTED-TOMOGRAPHY; IMAGE-RECONSTRUCTION; LEFT ATRIUM; TEMPORAL RESOLUTION; CATHETER ABLATION; PULMONARY VEINS; ANGIOGRAPHY; PERFORMANCE; ESOPHAGUS;
D O I
10.1109/TMI.2011.2172951
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Time-resolved cardiac imaging is particularly interesting in the interventional setting since it would provide both image guidance for accurate procedural planning and cardiac functional evaluations directly in the operating room. Imaging the heart in vivo using a slowly rotating C-arm system is extremely challenging due to the limitations of the data acquisition system and the high temporal resolution required to avoid motion artifacts. In this paper, a data acquisition scheme and an image reconstruction method are proposed to achieve time-resolved cardiac cone-beam computed tomography imaging with isotropic spatial resolution and high temporal resolution using a slowly rotating C-arm system. The data are acquired within 14 s using a single gantry rotation with a short scan angular range. The enabling image reconstruction method is the prior image constrained compressed sensing (PICCS) algorithm. The prior image is reconstructed from data acquired over all cardiac phases. Each cardiac phase is then reconstructed from the retrospectively gated cardiac data using the PICCS algorithm. To validate the method, several studies were performed. Both numerical simulations using a hybrid motion phantom with static background anatomy as well as physical phantom studies have been used to demonstrate that the proposed method enables accurate reconstruction of image objects with a high isotropic spatial resolution. A canine animal model scanned in vivo was used to further validate the method.
引用
收藏
页码:907 / 923
页数:17
相关论文
共 50 条
  • [21] C-arm fluoroscopic cone-beam CT for guidance of chemical thoracic sympathectomy
    Munetaka Hirose
    Mari Tabata
    Mikako Sakai
    Kenji Takeuchi
    Journal of Anesthesia, 2011, 25 : 142 - 143
  • [22] C-arm cone-beam computed tomography in interventional oncology: technical aspects and clinical applications
    Floridi, Chiara
    Radaelli, Alessandro
    Abi-Jaoudeh, Nadine
    Grass, Micheal
    De Lin, Ming
    Chiaradia, Melanie
    Geschwind, Jean-Francois
    Kobeiter, Hishman
    Squillaci, Ettore
    Maleux, Geert
    Giovagnoni, Andrea
    Brunese, Luca
    Wood, Bradford
    Carrafiello, Gianpaolo
    Rotondo, Antonio
    RADIOLOGIA MEDICA, 2014, 119 (07): : 521 - 532
  • [23] C-arm cone-beam computed tomography in interventional oncology: technical aspects and clinical applications
    Chiara Floridi
    Alessandro Radaelli
    Nadine Abi-Jaoudeh
    Micheal Grass
    Ming De Lin
    Melanie Chiaradia
    Jean-Francois Geschwind
    Hishman Kobeiter
    Ettore Squillaci
    Geert Maleux
    Andrea Giovagnoni
    Luca Brunese
    Bradford Wood
    Gianpaolo Carrafiello
    Antonio Rotondo
    La radiologia medica, 2014, 119 : 521 - 532
  • [24] Time-resolved C-arm cone beam CT angiography using SMART-RECON: Quantification of Temporal Resolution and Reconstruction Accuracy
    Garrett, John
    Li, Yinsheng
    Li, Ke
    Schafer, Sebastian
    Strother, Charles
    Chen, Guang-Hong
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [25] The Impact of the Use of C-Arm Cone-Beam CT During Chemoembolization for Hepatocellular Carcinoma
    Abdelsalam, Mohamed E.
    Figueira, Tomas M. A.
    Ensor, Joe
    Tam, Alda L.
    Avritscher, Rony
    Kaseb, Ahmed
    Gupta, Sanjay
    CURRENT MEDICAL IMAGING, 2022, 18 (04) : 372 - 380
  • [26] Algorithm-Enabled High-Performance C-arm Cone-Beam CT Angiography of Cerebral Vasculature
    Han, Xiao
    Zhang, Zheng
    Oishi, Satoru
    Satow, Tetsu
    Yokoyama, Hiromichi
    Yamada, Masanobu
    Silver, Michael D.
    Chang, Yu-Bing
    Sidky, Emil Y.
    Pan, Xiaochuan
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [27] High-Performance C-Arm Cone-Beam CT Guidance of Thoracic Surgery
    Schafer, Sebastian
    Otake, Yoshito
    Uneri, Ali
    Mirota, Daniel J.
    Nithiananthan, Sajendra
    Stayman, J. Webster
    Zbijewski, Wojciech
    Kleinszig, Gerhard
    Graumann, Rainer
    Sussman, Marc
    Siewerdsen, Jeffrey H.
    MEDICAL IMAGING 2012: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2012, 8316
  • [28] Investigation of C-arm cone-beam CT-guided surgery of the frontal recess
    Rafferty, MA
    Siewerdsen, JH
    Chan, Y
    Moseley, DJ
    Daly, MJ
    Jaffray, DA
    Irish, JC
    LARYNGOSCOPE, 2005, 115 (12): : 2138 - 2143
  • [29] GPU based parallel acceleration for fast C-arm cone-beam CT reconstruction
    Ken Chen
    Cheng Wang
    Jing Xiong
    Yaoqin Xie
    BioMedical Engineering OnLine, 17
  • [30] Development of a mobile C-arm platform for cone-beam CT guidance of surgical procedures
    Siewerdsen, J
    Moseley, D
    Kim, S
    Burch, S
    Bisland, S
    Bogaards, A
    Jaffray, D
    MEDICAL PHYSICS, 2004, 31 (06) : 1820 - 1820