Analytical evaluation of the new Beckman Coulter Access high sensitivity cardiac troponin I immunoassay

被引:5
|
作者
Lippi, Giuseppe [3 ]
Ferrari, Anna [1 ]
Gandini, Giorgio [2 ]
Gelati, Matteo [1 ]
Lo Cascio, Claudia [1 ]
Salvagno, Gian Luca [1 ]
机构
[1] Univ Verona, Sect Clin Biochem, Verona, Italy
[2] Univ Hosp Verona, Serv Transfus Med, Verona, Italy
[3] Univ Hosp Verona, Sect Clin Biochem, Ple Scuro 10, I-37314 Verona, Italy
关键词
immunoassay; method evaluation; myocardial infarction; troponin; ACUTE CORONARY SYNDROME; 99TH PERCENTILE; ASSAY; GENERATION; IMPRECISION; PERFORMANCE; ELEVATION; COHORT;
D O I
10.1515/cclm-2017-0350
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: This study was aimed to evaluate the analytical performance of the novel chemiluminescent and fully-automated Beckman Coulter Access hsTnI high-sensitivity immunoassay for measurement of cardiac troponin I (cTnI). Methods: The study, using lithium heparin samples, included assessment of limit of blank (LOB), limit of detection (LOD), functional sensitivity, linearity, imprecision (within run, between-run and total), calculation of 99th percentile upper reference limit (URL) in 175 healthy blood donors (mean age, 36 +/- 12 years; 47% women) and comparison with two other commercial cTnI immunoassays. Results: The LOB, LOD and functional sensitivity of Access hsTnI were 0.14, 0.34 and 1.35 ng/L, respectively. The within-run, between-run and total imprecision was 2.2%-2.9%, 4.6%-5.4%, and 5.4%-6.1%, respectively. The linearity was excellent in the range of cTnI values between 0.95 and 4195 ng/L (r = 1.00). The 99th percentile URL was 15.8 ng/L. Measurable cTnI values were found in 173/175 healthy subjects (98.9%). Good agreement of cTnI values was found with AccuTnI+3 (r = 0.97; mean bias, -9.3%), whereas less satisfactory agreement was found with Siemens Dimension Vista cTnI (r = 0.95; mean bias, -55%). Conclusions: The results of our evaluation of the Beckman Coulter Access hsTnI indicate that the analytical performance of this fully-automated immunoassay is excellent.
引用
收藏
页码:157 / 161
页数:5
相关论文
共 50 条
  • [31] Evaluation of a new Electrochemiluminescence Immunoassay System for Measuring High-Sensitivity Cardiac Troponin T
    Wang, Jinying
    Meng, Qianlong
    Zhao, Yi
    Li, Zhiyuan
    Ling, Lu
    Xu, Zhenzhen
    Li, Zhen
    Li, Xiaoyang
    CLINICAL LABORATORY, 2024, 70 (03) : 467 - 474
  • [32] Development of a whole blood analytical method for TSH using the Beckman Coulter access immunoassay system.
    Maggiore, J. A.
    Warner, C. N.
    Quaka, M. M.
    Roth, E.
    CLINICAL CHEMISTRY, 2007, 53 (06) : A191 - A191
  • [33] Evaluation of the Beckman Coulter Access® Testosterone Assay
    Dnistrian, A. M.
    Gonzalez-Espinosa, R.
    Celucia, F.
    Mathew, A.
    Fleisher, M.
    CLINICAL CHEMISTRY, 2009, 55 (06) : A180 - A180
  • [34] Analytical validation, sample stability, and clinical evaluation of a new high-sensitivity cardiac troponin I immunoassay for use in dogs, with comparison to a previous ultrasensitive assay
    Wesselowski, Sonya
    Lidbury, Jonathan
    Saunders, Ashley B.
    Gordon, Sonya G.
    Suchodolski, Jan S.
    Steiner, Joerg M.
    PLOS ONE, 2023, 18 (07):
  • [35] Evaluation of Kryptor cardiac troponin I immunoassay
    Lefèvre, G
    Graïne, H
    Bakkouch, K
    Garbarz, E
    Marlin, G
    Michel, PL
    Capeau, J
    ANNALS OF CLINICAL BIOCHEMISTRY, 2002, 39 : 145 - 147
  • [36] RACE-IT - Rapid Acute Coronary Syndrome Exclusion using the Beckman Coulter Access high-sensitivity cardiac troponin I: A stepped-wedge cluster randomized trial
    Miller, Joseph
    Cook, Bernard
    Singh-Kucukarslan, Gulmohar
    Tang, Amy
    Danagoulian, Shooshan
    Heath, Gerard
    Khalifa, Ziad
    Levy, Phillip
    Mahler, Simon A.
    Mills, Nicholas
    McCord, James
    CONTEMPORARY CLINICAL TRIALS COMMUNICATIONS, 2021, 22
  • [37] Evaluation of a modified lateral flow immunoassay for detection of high-sensitivity cardiac troponin I and myoglobin
    Zhu, Jimin
    Zou, Nengli
    Mao, Hongju
    Wang, Ping
    Zhu, Danian
    Ji, Huoyan
    Cong, Hui
    Sun, Changjiang
    Wang, Huimin
    Zhang, Feng
    Qian, Juying
    Jin, Qinghui
    Zhao, Jianlong
    BIOSENSORS & BIOELECTRONICS, 2013, 42 : 522 - 525
  • [38] A critical evaluation of the Beckman Coulter Access hsTnI: Analytical performance, reference interval and concordance
    Pretorius, Carel J.
    Tate, Jillian R.
    Wilgen, Urs
    Cullen, Louise
    Ungerer, Jacobus P. J.
    CLINICAL BIOCHEMISTRY, 2018, 55 : 49 - 55
  • [39] Clinical performance of Beckman Coulter's Access® AccuTnI™ (troponin I) in a multicenter clinical trial.
    Jevans, AW
    Apple, FS
    Wu, AH
    Venge, P
    Uettwiller-Geiger, D
    Darte, C
    Olson, M
    Southwick, PC
    CLINICAL CHEMISTRY, 2001, 47 (06) : A205 - A205
  • [40] Evaluation of Beckman Coulter UniCel DxI 800 immunoassay analyser
    Lam, C. W.
    Soh, K. P.
    Lim, S. F.
    CLINICAL CHEMISTRY, 2006, 52 (06) : A195 - A195