Validation of a micro-CT technique for measuring volumetric wear in retrieved acetabular liners

被引:27
|
作者
Bowden, AE [1 ]
Kurtz, SM
Edidin, AA
机构
[1] Drexel Univ, Implant Res Ctr, Philadelphia, PA 19104 USA
[2] Exponent Inc, Philadelphia, PA USA
关键词
polyethylene; wear; retrieval; microCT; registration;
D O I
10.1002/jbm.b.30318
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this study, a novel micro-CT-based technique for evaluating wear in retrieved acetabular liners was introduced and validated. Six UHMWPE acetabular components ranging in implantation time from 2.7 to 14.4 years were collected and evaluated with the use of a high-resolution micro-CT scanner. The components were scanned with a uniform volumetric resolution of 74 microns (16-bit precision) with the use of a 1024 X 1024 in-plane image matrix. Manual rigid 3D image registration of the interior hemispherical portion of the acetabular cup with geometric primitives by trained observers allowed for isolation, visualization, and measurement of the wear volume. Results for these six components indicated an average wear rate of 65 mm(3)/year. Overall scanner error was quantified gravimetrically and associated with a maximum uncertainty of 0.6%. Intra- and interobserver uncertainty analysis showed the method to be both accurate and repeatable. (c) 2005 Wiley Periodicals, Inc.
引用
收藏
页码:205 / 209
页数:5
相关论文
共 50 条
  • [41] A Nondestructive Method of Measuring Zebrafish Adipose Tissue Based on Micro-Computed Tomography (Micro-CT)
    Wang, Xin
    Wang, Guangxin
    Xiao, Yuan
    Zuo, Yanxia
    Zhou, Fang
    APPLIED SCIENCES-BASEL, 2021, 11 (22):
  • [42] Influence of Shrinkage and Viscosity of Flowable Composite Liners on Cervical Microleakage of Class II Restorations: A Micro-CT Analysis
    Nie, J.
    Yap, A. U.
    Wang, X. Y.
    OPERATIVE DENTISTRY, 2018, 43 (06) : 656 - 664
  • [43] Micro-CT as a novel technique for 3D reconstruction of molluscan anatomy
    Golding, Rosemary E.
    Jones, Allan S.
    MOLLUSCAN RESEARCH, 2007, 27 (03) : 123 - 128
  • [44] Suitability of the micro-CT technique to analyze remodeling processes in carotid arteries of mouse
    Schuermann, C.
    Yasar, Y.
    Brandes, R.
    ACTA PHYSIOLOGICA, 2014, 210 : 182 - 182
  • [45] Measuring bone strength from in vivo micro-CT using the finite element method
    Boyd, Steven
    Macdonald, H. M.
    Hanley, D. A.
    BONE, 2008, 43 : S18 - S18
  • [46] Selective Deblurring for Improved Calcification Visualization and Quantification in Carotid CT Angiography: Validation Using Micro-CT
    Rollano-Hijarrubia, Empar
    Manniesing, Rashindra
    Niessen, Wiro J.
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2009, 28 (03) : 446 - 453
  • [47] Measuring bone strength from in vivo micro-CT using the finite element method
    Boyd, Steven
    Macdonald, Heather
    Hanley, David
    BONE, 2008, 43 : S59 - S59
  • [48] From histology to micro-CT: Measuring and modeling resorption cavities and their relation to bone competence
    Vanderoost, Jef
    van Lenthe, G. Harry
    WORLD JOURNAL OF RADIOLOGY, 2014, 6 (09): : 643 - 656
  • [49] Validation of micro-CT against the section method regarding the assessment of marginal leakage of sealants
    Chen, X.
    Cuijpers, V. M. J. I.
    Fan, M. W.
    Frencken, J. E.
    AUSTRALIAN DENTAL JOURNAL, 2012, 57 (02) : 196 - 199
  • [50] Validation of HR-pQCT against micro-CT for morphometric and biomechanical analyses: A review
    Ohs, Nicholas
    Collins, Caitlyn J.
    Atkins, Penny R.
    BONE REPORTS, 2020, 13