Validation of a micro-CT technique for measuring volumetric wear in retrieved acetabular liners

被引:27
|
作者
Bowden, AE [1 ]
Kurtz, SM
Edidin, AA
机构
[1] Drexel Univ, Implant Res Ctr, Philadelphia, PA 19104 USA
[2] Exponent Inc, Philadelphia, PA USA
关键词
polyethylene; wear; retrieval; microCT; registration;
D O I
10.1002/jbm.b.30318
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this study, a novel micro-CT-based technique for evaluating wear in retrieved acetabular liners was introduced and validated. Six UHMWPE acetabular components ranging in implantation time from 2.7 to 14.4 years were collected and evaluated with the use of a high-resolution micro-CT scanner. The components were scanned with a uniform volumetric resolution of 74 microns (16-bit precision) with the use of a 1024 X 1024 in-plane image matrix. Manual rigid 3D image registration of the interior hemispherical portion of the acetabular cup with geometric primitives by trained observers allowed for isolation, visualization, and measurement of the wear volume. Results for these six components indicated an average wear rate of 65 mm(3)/year. Overall scanner error was quantified gravimetrically and associated with a maximum uncertainty of 0.6%. Intra- and interobserver uncertainty analysis showed the method to be both accurate and repeatable. (c) 2005 Wiley Periodicals, Inc.
引用
收藏
页码:205 / 209
页数:5
相关论文
共 50 条
  • [31] Micro-CT imaging technique to characterize diffusion of small- molecules
    Khazaee, Tina
    Norley, Chris J. D.
    Nikolov, Hristo N.
    Pollmann, Steven, I
    Holdsworth, David W.
    MEDICAL IMAGING 2020: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING, 2021, 11317
  • [32] Development of measuring technique for micro-scale wear performance
    Chen, Ping
    Chen, Hua-Hui
    Li, Guo-Hua
    Ning, Hua-Rong
    Jinshu Rechuli/Heat Treatment of Metals, 2007, 32 (11): : 69 - 73
  • [33] Intramyocardial capillary blood volume estimated by whole-body CT - validation by micro-CT
    Dong, Yue
    Beighley, Patricia E.
    Eaker, Diane R.
    Zamir, Mair
    Ritman, Erik L.
    MEDICAL IMAGING 2008: PHYSIOLOGY, FUNCTION, AND STRUCTURE FROM MEDICAL IMAGES, 2008, 6916
  • [34] Validation of a micro-CT system for the 3-D measurement of bone.
    Chmielewski, PA
    Dufresne, TE
    Combs, KH
    Borah, B
    Lundy, MW
    JOURNAL OF BONE AND MINERAL RESEARCH, 2001, 16 : S461 - S461
  • [35] Validation of a 3D CT method for measurement of linear wear of acetabular cups
    Jedenmalm, Anneli
    Nilsson, Fritjof
    Noz, Marilyn E.
    Green, Douglas D.
    Gedde, Ulf W.
    Clarke, Ian C.
    Stark, Andreas
    Maguire, Gerald Q., Jr.
    Zeleznik, Michael P.
    Olivecrona, Henrik
    ACTA ORTHOPAEDICA, 2011, 82 (01) : 35 - 41
  • [36] Validation of Fluorescence Molecular Tomography/Micro-CT Multimodal Imaging In Vivo in Rats
    Vonwil, Daniel
    Christensen, Jon
    Fischer, Sascha
    Ronneberger, Olaf
    Shastri, V. Prasad
    MOLECULAR IMAGING AND BIOLOGY, 2014, 16 (03) : 350 - 361
  • [37] Assessing Enamel Wear of Monolithic Ceramics With Micro-CT and Intra-oral Scanner
    Murbay, Sukhdeep
    Yeung, Shadow Kim Wai
    Yip, Chui Ying
    Pow, Edmond Ho Nang
    INTERNATIONAL DENTAL JOURNAL, 2023, 73 (04) : 496 - 502
  • [38] Validation of micro-CT techniques and 3-D architectural measures of bone
    Dufresne, TE
    Chmielewski, PA
    Borah, B
    JOURNAL OF BONE AND MINERAL RESEARCH, 2000, 15 (04) : 800 - 800
  • [39] Validation of Fluorescence Molecular Tomography/Micro-CT Multimodal Imaging In Vivo in Rats
    Daniel Vonwil
    Jon Christensen
    Sascha Fischer
    Olaf Ronneberger
    V. Prasad Shastri
    Molecular Imaging and Biology, 2014, 16 : 350 - 361
  • [40] Quantification of in vitro wear of a synthetic meniscus implant using gravimetric and micro-CT measurements
    Elsner, Jonathan J.
    Shemesh, Maoz
    Shefy-Peleg, Adaya
    Gabet, Yankel
    Zylberberg, Eyal
    Linder-Ganz, Eran
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2015, 49 : 310 - 320