Biomechanical comparison of interfragmentary compression in transverse fractures of the olecranon

被引:73
|
作者
Wilson, J.
Bajwa, A. [1 ]
Kamath, V. [2 ]
Rangan, A. [1 ]
机构
[1] James Cook Univ Hosp, Middlesbrough TS4 3BW, Cleveland, England
[2] Victoria Hosp, Blackpool FY3 8NR, Lancashire, England
来源
关键词
FIXATION METHODS;
D O I
10.1302/0301-620X.93B2.24613
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Compression and absolute stability are important in the management of intra-articular fractures. We compared tension band wiring with plate fixation for the treatment of fractures of the olecranon by measuring compression within the fracture. Identical transverse fractures were created in models of the ulna. Tension band wires were applied to ten fractures and ten were fixed with Acumed plates. Compression was measured using a Tekscan force transducer within the fracture gap. Dynamic testing was carried out by reproducing cyclical contraction of the triceps of 20 N and of the brachialis of 10 N. Both methods were tested on each sample. Paired t-tests compared overall compression and compression at the articular side of the fracture. The mean compression for plating was 819 N (SD 602, 95% confidence interval (CI)) and for tension band wiring was 77 N (SD 19, 95% CI) (p = 0.039). The mean compression on the articular side of the fracture for plating was 343 N (SD 276, 95% CI) and for tension band wiring was 1 N (SD 2, 95% CI) (p = 0.038). During simulated movements, the mean compression was reduced in both groups, with tension band wiring at -14 N (SD 7) and for plating -173 N (SD 32). No increase in compression on the articular side was detected in the tension band wiring group. Pre-contoured plates provide significantly greater compression than tension bands in the treatment of transverse fractures of the olecranon, both over the whole fracture and specifically at the articular side of the fracture. In tension band wiring the overall compression was reduced and articular compression remained negligible during simulated contraction of the triceps, challenging the tension band principle.
引用
收藏
页码:245 / 250
页数:6
相关论文
共 50 条
  • [41] Comminuted olecranon fractures: A comparison of plating methods
    Gordon, MJ
    Budoff, JE
    Yeh, ML
    Luo, ZP
    Noble, PC
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2006, 15 (01) : 94 - 99
  • [42] Biomechanical Assessment of Locking Plate Fixation of Comminuted Proximal Olecranon Fractures
    Fantry, Amanda
    Sobel, Andrew
    Capito, Nicholas
    Hodax, Jonathan D.
    Pidgeon, Tyler
    Koruprolu, Sarath
    Rich, Ryan
    Fava, Joseph L.
    Green, Andrew
    Paxton, Scott
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2018, 32 (11) : E445 - E450
  • [43] Biomechanical Comparison of Olecranon Sled Versus Intramedullary Screw Tension Banding for Olecranon Osteotomies
    Kia, Cameron
    Dukas, Alex G.
    Marshall, Silas T.
    Voss, Andreas
    Obopilwe, Elifho
    Browner, Bruce D.
    Mazzocca, Augustus D.
    ORTHOPAEDIC JOURNAL OF SPORTS MEDICINE, 2018, 6 (12):
  • [44] Lag screw osteosynthesis of simple olecranon fractures: A biomechanical comparative study
    Wagner, Ferdinand C.
    Hamann, Alexander
    Maier, Dirk
    Ophoven, Christian
    Yilmaz, Tayfun
    Sudkamp, Norbert P.
    Jaeger, Martin
    Reising, Kilian
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2022, 236 (06) : 841 - 847
  • [45] Biomechanical Analysis of a Transverse Olecranon Fracture Model Using Tension Band Wiring
    Hammond, James
    Ruland, Robert
    Hogan, Christopher
    Rose, David
    Belkoff, Stephen
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2012, 37A (12): : 2506 - 2511
  • [46] Osteosynthesis of Phalangeal Fractures: Biomechanical Comparison of Kirschner Wires, Plates, and Compression Screws
    Rausch, Valentin
    Harbrecht, Andreas
    Kahmann, Stephanie L.
    Fenten, Thomas
    Jovanovic, Nebojsa
    Hackl, Michael
    Mueller, Lars P.
    Staat, Manfred
    Wegmann, Kilian
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2020, 45 (10): : 987.e1 - 987.e8
  • [47] Biomechanical comparison between PMMA and hydroxyapatite vertebroplasty in the treatment of osteoporotic compression fractures
    Lin, JCN
    Wu, YC
    Wang, NP
    XXXIII WORLD CONGRESS OF THE INTERNATIONAL COLLEGE OF SURGEONS - ICS 2002, 2002, : 173 - 178
  • [48] Proximal femoral fractures: a new concept of implant, stress and interfragmentary compression analysis
    Billard, Remi
    Vacher, Pierre
    Vittecoq, Eric
    Toussaint, Franck
    Devun, Laure
    Bonjour, Christian
    Arlettaz, Yvan
    SWISS MEDICAL WEEKLY, 2013, 143 : 25S - 25S
  • [49] Long-term Outcome of Displaced, Transverse, Noncomminuted Olecranon Fractures
    Flinterman, Hendrik J. A.
    Doornberg, Job N.
    Guitton, Thierry G.
    Ring, David
    Goslings, J. Carel
    Kloen, Peter
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2014, 472 (06) : 1955 - 1961
  • [50] Biomechanical comparison of straight DCP and helical plates for fixation of transverse and oblique bone fractures
    Aksakal, Bunyamin
    Gurger, Murat
    Say, Yakup
    Yilmaz, Erhan
    ACTA OF BIOENGINEERING AND BIOMECHANICS, 2014, 16 (04) : 67 - 74