Size-specific dose estimates for pediatric head CT protocols based on the AAPM report TG-293

被引:3
|
作者
Sapignoli, Sonia [1 ]
Roggio, Antonella [1 ]
Boschini, Alessio [2 ]
Guida, Federica [1 ]
Merlo, Carlo [1 ]
Paiusco, Marta [1 ]
Zorz, Alessandra [1 ]
De Monte, Francesca [1 ,3 ]
机构
[1] IRCCS, Veneto Inst Oncol IOV, Med Phys Dept, Padua, Italy
[2] Med Phys Dept, Azienda ULSS2 Marca Trevigiana, Treviso, Italy
[3] IRCCS, Veneto Inst Oncol IOV, Med Phys Dept, Via Gattamelata 64, I-35128 Padua, Italy
关键词
SSDE; Head CT; Pediatric; Radiation dose tracking system; COMPUTED-TOMOGRAPHY; RADIATION-EXPOSURE; SUBSEQUENT RISK; PATTERNS; CHILDREN; TRENDS; SCAN;
D O I
10.1016/j.ejmp.2022.06.004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To establish the Size Specific Dose Estimate (SSDE) typical values for pediatric head CT examinations based on the AAPM report TG-293; to provide a new stratification based on the water-equivalent diameter (Dw ), given that SSDE is related to the head size. Methods: 296 Head CT scans of pediatric patients collected using a dose monitoring software were retrospectively analysed. Typical values were derived stratifying data by age in three methods: the first proposed by the European Guidelines on Diagnostic Reference Levels for Pediatric Imaging (RP185), the second by the National Istisan Report 20/22 and a local one related to the clinical protocols (LStrata). For each scan, a self-developed Matlab routine calculated the water-equivalent diameter (Dw) and related SSDE values with the conversion factors fH16and fB16provided by the AAPM reports TG-293 and TG-204, respectively. Eventually, a Dwstratification was introduced starting from a measure of the lateral dimension of the head. Results: SSDE based on TG-204 overestimatesthe dose up to 12%. Four Dwgroups were identified thanks to the good correlation between the head lateral dimension andDw: Dw < 14 cm, 14 < Dw < 16 cm, 16 < Dw < 17 cm, Dw >= 17 cm. The Dw -stratified dosimetric indices present greater variability than those grouped by age because of the large variability of the size of the infant's head. Conclusions: The variability of the SSDE metric underlines that age-optimized protocols are not when size is considered.
引用
收藏
页码:26 / 30
页数:5
相关论文
共 50 条
  • [1] Comment on the "Report of AAPM TG 204: Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations" [Report of AAPM TG 204, 2011]
    Li, Baojun
    Behrman, Richard H.
    MEDICAL PHYSICS, 2012, 39 (07) : 4613 - 4614
  • [2] Reply to "Comment on the 'Report of AAPM TG 204: Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations'" [AAPM Report 204, 2011]
    Boone, John M.
    MEDICAL PHYSICS, 2012, 39 (07) : 4615 - 4616
  • [3] Determining Size-Specific Dose Estimates for Head CT Examinations
    Hardy, A.
    Bostani, M.
    Zankl, M.
    McCollough, C.
    Cagnon, C.
    McNitt-Gray, M.
    MEDICAL PHYSICS, 2017, 44 (06) : 3314 - 3315
  • [4] Simplifying Size-Specific Radiation Dose Estimates in Pediatric CT
    Khawaja, Ranish Deedar Ali
    Singh, Sarabjeet
    Vettiyil, Beth
    Lim, Ruth
    Gee, Michael
    Westra, Sjirk
    Kalra, Mannudeep K.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 204 (01) : 167 - 176
  • [5] A Monte Carlo Study to Determine SSDE Coefficients for Pediatric Head CT (TG-293)
    Hernandez, A. M.
    Boone, J. M.
    MEDICAL PHYSICS, 2017, 44 (06) : 3314 - 3314
  • [6] THE SIZE-SPECIFIC DOSE ESTIMATE OF PAEDIATRIC HEAD CT EXAMINATIONS FOR VARIOUS PROTOCOLS
    Fahmi, Arif
    Anam, Choirul
    Ali, Mohd Hanafi
    RADIATION PROTECTION DOSIMETRY, 2020, 188 (04) : 522 - 528
  • [7] Evaluation of Size-Specific Dose Estimates for Optimizing Pediatric Chest CT Protocol
    Zarei F.
    Nasiri M.
    Etemadi Z.
    Haghighi R.R.
    Chatterjee S.
    Khaneghah P.A.
    Chatterjee V.V.
    Iranian Journal of Medical Physics, 2022, 19 (05): : 315 - 321
  • [8] Size-specific dose estimates in pediatric brain CT in relation to age and weight
    Fujii, Kaoru
    Inoue, Yusuke
    Itoh, Hiroyasu
    Sasa, Ryosuke
    Hata, Hirofumi
    Mitsui, Kohei
    RADIATION PROTECTION DOSIMETRY, 2023, 199 (10) : 1075 - 1081
  • [9] Effects of Metal On CT Size-Specific Dose Estimates
    Alsanea, M.
    Krugh, K.
    Layman, R.
    MEDICAL PHYSICS, 2016, 43 (06) : 3720 - 3720
  • [10] Improved Estimates of Trunk and Head CT Radiation Dose: Development of Size-Specific Dose Estimate
    Strauss, Keith J.
    McKenney, Sarah E.
    Brady, Samuel L.
    JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2020, 17 (04) : 560 - 562