Dosimetric verification of biologically adapted IMRT

被引:1
|
作者
Rodal, Jan [1 ,2 ]
Waldeland, Einar [1 ]
Sovik, Aste [1 ,3 ]
Malinen, Eirik [1 ]
Rodal, Jan [1 ,2 ]
Sovik, Aste [1 ,3 ]
机构
[1] Oslo Univ Hosp, Norwegian Radium Hosp, Dept Med Phys, N-0310 Oslo, Norway
[2] Norwegian Univ Sci & Technol, Dept Phys, N-7491 Trondheim, Norway
[3] Norwegian Sch Vet Sci, Dept Compan Anim Clin Sci, N-0033 Oslo, Norway
关键词
radiation therapy; IMRT; image guidance; dosimetry; medical imaging; MODULATED RADIATION-THERAPY; KODAK EDR2 FILM; POSITRON-EMISSION-TOMOGRAPHY; COLLIMATOR LEAF WIDTH; DOSE DISTRIBUTIONS; MULTILEAF COLLIMATOR; NECK-CANCER; STEREOTACTIC RADIOSURGERY; QUANTITATIVE-EVALUATION; HETEROGENEOUS TUMORS;
D O I
10.1118/1.3581406
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate the delivery of biologically adapted high resolution intensity modulated radiotherapy (IMRT) to an anthropomorphic phantom, using dosimetric and radiobiologic measures. Methods: A compartment based 3D hypoxia map of a highly heterogeneous tumor was imported into the CT planning basis of an anthropomorphic phantom. Biologically adapted IMRT was planned according to the corresponding 3D dose prescription map with elevated dose to the hypoxic regions. Three treatment fractions were delivered to the phantom by means of a conventional linear accelerator equipped with a high resolution micro multileaf collimator (mMLC). EDR2 radiographic films were positioned in two planes of the phantom during irradiation. Software was developed to analyze dose distributions from the prescription, dose plan, and films. Dose distributions were scored within each of four radiobiologic tumor compartments. The treatment effect in each tumor compartment was estimated as the equivalent uniform dose (EUD). A conventional c analysis was utilized for quantitative comparisons of planned and delivered dose distributions. Results: The planned and delivered dose maps qualitatively resembled the prescribed dose map. The gamma analyzes showed that, on average for all films, more than 95% of the pixels within the tumor passed the 3%/2 mm criteria. For compartments with increasing degree of hypoxia and thereby increased prescribed dose, the planned and delivered EUDs were severely reduced compared to the prescription. The prescribed compartmental doses were met only for the most oxic compartment. The mean tumor dose as measured by the films was 6.6% lower than the corresponding planned dose. Conclusions: Biologically adapted radiotherapy may be delivered with high precision according to the dose plan. However, the large reduction in compartmental EUD values from prescribed to planned treatment indicated lower tumor effect than expected for such a treatment. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3581406]
引用
收藏
页码:2586 / 2594
页数:9
相关论文
共 50 条
  • [1] DOSIMETRIC AND RADIOBIOLOGIC EVALUATION OF BIOLOGICALLY ADAPTED IMRT
    Rodal, J.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S74 - S74
  • [2] IMRT DOSIMETRIC VERIFICATION PROCEDURE
    Pellejero, S.
    Lozares, S.
    Forner, A.
    Iriondo, U.
    Otal, A.
    Soto, P. M.
    Maneru, F.
    Martin-Albina, M. L.
    Miquelez, S.
    Rubio, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S512 - S512
  • [3] Quantitative dosimetric verification for IMRT
    Narita, Y
    Hatano, K
    Sakai, M
    Araki, H
    [J]. RADIOTHERAPY AND ONCOLOGY, 2003, 68 : S105 - S105
  • [4] Statistical process control for IMRT dosimetric verification
    Breen, Stephen L.
    Moseley, Douglas J.
    Zhang, Beibei
    Sharpe, Michael B.
    [J]. MEDICAL PHYSICS, 2008, 35 (10) : 4417 - 4425
  • [5] Dosimetric verification needs for conformal therapy and IMRT
    Fraass, B
    Moran, J
    [J]. MEDICAL PHYSICS, 2002, 29 (06) : 1358 - 1358
  • [6] Quantitative dosimetric verification of an IMRT planning and delivery system
    Low, DA
    Mutic, S
    Dempsey, JF
    Gerber, RL
    Bosch, WR
    Perez, CA
    Purdy, JA
    [J]. RADIOTHERAPY AND ONCOLOGY, 1998, 49 (03) : 305 - 316
  • [7] Dosimetric verification of IMRT delivery: Varian, Siemens, and Elekta
    Willcut, V
    Newman, F
    Zhu, X
    Low, D
    Percy, J
    Cain, M
    Markman, J
    Munger, T
    Nelms, B
    [J]. MEDICAL PHYSICS, 2002, 29 (06) : 1365 - 1365
  • [8] Investigation of the use of MOSFET for clinical IMRT dosimetric verification
    Chuang, CF
    Verhey, LJ
    Xia, P
    [J]. MEDICAL PHYSICS, 2002, 29 (06) : 1109 - 1115
  • [9] Implementation of IMRT treatment: verification of dosimetric and geometric accuracy
    Korevaar, EW
    Davis, JB
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2002, 178 : 119 - 119
  • [10] The use radichromic film for tornotherapy IMRT dosimetric verification
    Turian, J.
    Khelashvili, G.
    Chu, J.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2497 - 2498