Implant fracture of the Regenerex® modular metal tibial component: A report of three cases

被引:6
|
作者
Rytter, Soren [1 ]
Madsen, Frank [1 ]
Jepsen, Claus Fink [1 ]
Stilling, Maiken [1 ]
机构
[1] Aarhus Univ Hosp, Dept Orthoped, Aarhus, Denmark
来源
KNEE | 2019年 / 26卷 / 05期
关键词
Knee; Arthroplasty; Failure; Fracture; Regenerex (R); TOTAL KNEE ARTHROPLASTY; REPLACEMENT; FAILURE; PROSTHESES; REVISION; HIP;
D O I
10.1016/j.knee.2019.06.009
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Implant fractures are a very rare complication in primary total knee replacement (TKR) surgery and with modern implant designs and improved metals these events have nearly been eliminated. In this case series we report three cases of tibial metal baseplate fractures in uncemented Regenerex (R) TKR. Methods: Cases originated from a prospective case series of 80 patients operated between 2013 and 2016. Five patients were pilot cases and 75 were participants in a prospective randomized double-blinded clinical trial that evaluated different adjuvant bone anti-resorptive medical therapies. All patients were treated with an uncemented press-fit Regenerex (R) Porous Titanium Construct tibial tray and matching cemented (Refobacin Bone Cement R) patella and femoral components (hybrid implant). Results: We report three cases of medial side metal baseplate fractures of a modular finned tibial stem. All three baseplate fractures were in male patients. Confirmed failure of the implant occurred after 10, 12 and 23 months, in situ, with a mean follow-up of 15 months (range 10-23). Conclusions: Based on the current case series we cannot make any causal inferences. Failures may represent a multifactorial process with a cascade of events with implant failure as the result. However, like in most other case reports of metal failures in the literature, the implant fractures in this report were located on the medial side of the tibial component in male patients. (C) 2019 Elsevier B.V. All rights reserved.
引用
收藏
页码:1143 / 1151
页数:9
相关论文
共 50 条
  • [1] Posterior tibial stress fracture: A report of three cases
    van der Velde, GM
    Hsu, WS
    JOURNAL OF MANIPULATIVE AND PHYSIOLOGICAL THERAPEUTICS, 1999, 22 (05) : 341 - 346
  • [2] FRACTURE OF THE KNEE ENDOPROSTHESIS - REPORT OF 3 CASES OF TIBIAL COMPONENT FAILURE
    TJORNSTRAND, B
    LIDGREN, L
    ACTA ORTHOPAEDICA SCANDINAVICA, 1985, 56 (02): : 124 - 126
  • [3] Peri-implant fracture of the distal tibia after intra-medullary nailing of a tibial fracture: A report of two cases
    Mounasamy V.
    Desai P.
    European Journal of Orthopaedic Surgery & Traumatology, 2013, 23 (Suppl 2) : S279 - S283
  • [4] TRAUMATIC FIBULAR BOWING WITH TIBIAL FRACTURE - REPORT OF 2 CASES
    MANOLI, A
    ORTHOPEDICS, 1978, 1 (02) : 145 - 148
  • [5] Fracture and dislocation of the sternum - Report of three cases
    Holderman, HH
    ANNALS OF SURGERY, 1928, 88 : 252 - 259
  • [6] Fracture of the neck of the femur - Report of three cases
    Ayers, CE
    BOSTON MEDICAL AND SURGICAL JOURNAL, 1918, 178 : 298 - 300
  • [7] Fracture of anterior cervical plate implant - report of two cases
    Yen, CP
    Hwang, TY
    Wang, CJ
    Howng, SL
    ACTA NEUROCHIRURGICA, 2005, 147 (06) : 665 - 667
  • [8] Fracture of anterior cervical plate implant – report of two cases
    C. P. Yen
    T. Y. Hwang
    C. J. Wang
    S. L. Howng
    Acta Neurochirurgica, 2005, 147 : 665 - 667
  • [9] Component Fracture in the Kotz Modular Femoral Tibial Reconstruction System: An Under-Reported Complication
    Hauer, Tyler M.
    Houdek, Matthew T.
    Bhumbra, Rej
    Griffin, Anthony M.
    Wunder, Jay S.
    Ferguson, Peter C.
    JOURNAL OF ARTHROPLASTY, 2018, 33 (02): : 544 - 547
  • [10] First Metatarsophalangeal Joint Replacement with Modular Three-Component Press-Fit Implant. Preliminary Report
    Kolodziej, L.
    Bohatyrewicz, A.
    Zietek, P.
    ACTA CHIRURGIAE ORTHOPAEDICAE ET TRAUMATOLOGIAE CECHOSLOVACA, 2013, 80 (01) : 64 - 68