Implant fracture of the Regenerex® modular metal tibial component: A report of three cases

被引:6
|
作者
Rytter, Soren [1 ]
Madsen, Frank [1 ]
Jepsen, Claus Fink [1 ]
Stilling, Maiken [1 ]
机构
[1] Aarhus Univ Hosp, Dept Orthoped, Aarhus, Denmark
来源
KNEE | 2019年 / 26卷 / 05期
关键词
Knee; Arthroplasty; Failure; Fracture; Regenerex (R); TOTAL KNEE ARTHROPLASTY; REPLACEMENT; FAILURE; PROSTHESES; REVISION; HIP;
D O I
10.1016/j.knee.2019.06.009
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Implant fractures are a very rare complication in primary total knee replacement (TKR) surgery and with modern implant designs and improved metals these events have nearly been eliminated. In this case series we report three cases of tibial metal baseplate fractures in uncemented Regenerex (R) TKR. Methods: Cases originated from a prospective case series of 80 patients operated between 2013 and 2016. Five patients were pilot cases and 75 were participants in a prospective randomized double-blinded clinical trial that evaluated different adjuvant bone anti-resorptive medical therapies. All patients were treated with an uncemented press-fit Regenerex (R) Porous Titanium Construct tibial tray and matching cemented (Refobacin Bone Cement R) patella and femoral components (hybrid implant). Results: We report three cases of medial side metal baseplate fractures of a modular finned tibial stem. All three baseplate fractures were in male patients. Confirmed failure of the implant occurred after 10, 12 and 23 months, in situ, with a mean follow-up of 15 months (range 10-23). Conclusions: Based on the current case series we cannot make any causal inferences. Failures may represent a multifactorial process with a cascade of events with implant failure as the result. However, like in most other case reports of metal failures in the literature, the implant fractures in this report were located on the medial side of the tibial component in male patients. (C) 2019 Elsevier B.V. All rights reserved.
引用
收藏
页码:1143 / 1151
页数:9
相关论文
共 50 条
  • [21] FRACTURE OF THE TIBIAL TRAY FOLLOWING PCA KNEE REPLACEMENT - A REPORT OF 2 CASES
    MARUYAMA, M
    TERAYAMA, K
    SUNOHARA, H
    ADACHI, T
    SUZUKI, S
    FUKUZAWA, T
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1994, 113 (06) : 330 - 333
  • [22] AVULSION FRACTURE OF THE TIBIAL TUBERCLE WITH AVULSION OF THE PATELLAR LIGAMENT - REPORT OF 2 CASES
    FRANKL, U
    WASILEWSKI, SA
    HEALY, WL
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (09): : 1411 - 1413
  • [23] Posterior tibial tendon tenosynovectomy for rheumatoid arthritis: A report of three cases
    Tokunaga, Daisaku
    Hojo, Tatsuya
    Takatori, Ryota
    Ikoma, Kazuya
    Nagasawa, Koji
    Takamiya, Hisatake
    Ishida, Masashi
    Kubo, Toshikazu
    FOOT & ANKLE INTERNATIONAL, 2006, 27 (06) : 465 - 468
  • [24] Case Report: Composite pheochromocytoma with ganglioneuroma component: A report of three cases
    Araujo, Paula B.
    Carvallo, Mirna S.
    Vidal, Ana P.
    Nascimento, Joao B.
    Wo, Julia M.
    Naliato, Erika O.
    Neto, Silvio H. Cunha
    Conceicao, Flavia L.
    Fontes, Rosita
    de Lima, Vinicius V.
    Carvalho, Denise P.
    Soares, Paula
    Lima, Jorge
    Lourenco Jr, Delmar M.
    Violante, Alice Helena D.
    FRONTIERS IN ENDOCRINOLOGY, 2022, 13
  • [25] Total Talar Replacement with Ceramic Implant in Combination with Tibial Component of Ankle Endoprosthesis: A Case Report
    Kuznetsov, Vasilii V.
    Gudi, Sergei M.
    Skuratova, Liliya K.
    Pakhomov, Igor A.
    TRAVMATOLOGIYA I ORTOPEDIYA ROSSII, 2021, 27 (04): : 111 - 119
  • [26] Gas gangrene following implant removal after the union of a tibial plateau fracture: a case report
    Wang, Shanxi
    Liu, Lei
    BMC MUSCULOSKELETAL DISORDERS, 2018, 19
  • [27] Gas gangrene following implant removal after the union of a tibial plateau fracture: a case report
    Shanxi Wang
    Lei Liu
    BMC Musculoskeletal Disorders, 19
  • [28] Implant Installation With Simultaneous Ridge Augmentation. Report of Three Cases
    Park, Jun-Beom
    JOURNAL OF ORAL IMPLANTOLOGY, 2011, 37 (05) : 595 - 603
  • [29] Hoffa fracture associated with femoral shaft and proximal tibial fractures: report of two cases
    Jain, Anuj
    Agrawal, Prakash
    Chadha, Manish
    Pankaj, Amite
    CHINESE JOURNAL OF TRAUMATOLOGY, 2012, 15 (06) : 367 - 369
  • [30] Hoffa fracture associated with femoral shaft and proximal tibial fractures:report of two cases
    Anuj Jain
    Prakash Agrawal
    Manish Chadha
    Amite Pankaj
    中华创伤杂志(英文版), 2012, (06) : 367 - 369