Amperometric Determination of Glucose in White Grape and in Tablets as Ingredient by Screen-Printed Electrode Modified with Glucose Oxidase and Composite of Platinum and Multiwalled Carbon Nanotubes

被引:11
|
作者
Guzsvany, Valeria [1 ]
Anojcic, Jasmina [1 ]
Vajdle, Olga [1 ]
Radulovic, Emil [1 ]
Madarasz, Daniel [2 ]
Konya, Zoltan [2 ,3 ]
Kalcher, Kurt [4 ]
机构
[1] Univ Novi Sad, Dept Chem Biochem & Environm Protect, Fac Sci, Trg D Obradovica 3, Novi Sad 21000, Serbia
[2] Univ Szeged, Dept Appl & Environm Chem, Rerrich Bela Ter 1, H-6720 Szeged, Hungary
[3] MTA SZTE React Kinet & Surface Chem Res Grp, Rerrich Bela Ter 1, H-6720 Szeged, Hungary
[4] Karl Franzens Univ Graz, Inst Chem Analyt Chem, Univ Pl 1, A-8010 Graz, Austria
关键词
Screen-printed carbon electrode; Platinum and multiwalled carbon nanotube composite mediator; Glucose oxidase; Amperometry; Glucose determination; Food samples; ELECTROCHEMICAL SENSORS; BIOSENSORS; NANOPARTICLES; FOOD;
D O I
10.1007/s12161-018-1387-7
中图分类号
TS2 [食品工业];
学科分类号
0832 ;
摘要
Multiwalled carbon nanotubes (MWCNTs) and nanocomposite made from MWCNTs and Pt nanoparticles (Pt-MWCNTs) were used as bulk modifiers of screen-printed carbon electrodes (SPCEs). Cyclic voltammetric and amperometric measurements were performed by the homemade substrate electrodes, the bare SPCE and the modified ones (MWCNT-SPCE and Pt-MWCNT-SPCE), with aim to investigate their applicability for determination of H2O2 in phosphate buffer solution (0.1M; pH7.50) as supporting electrolyte. Pt-MWCNT-SPCE showed improved analytical performances including the significant decrease in overpotential of H2O2 compared to SPCE and MWCNT-SPCE. So, Pt-MWCNT-SPCE served as substrate electrode for design of a simple first-generation biosensor made from glucose oxidase (GOx) and Nafion((R)) forming the GOx/Pt-MWCNT-SPCE for reliable determination of glucose via its enzymatic by-product, the H2O2. The surface morphology characterization by scanning electron microscopy confirmed the presence of immobilized spherical shaped biosensing units at the GOx/Pt-MWCNT-SPCE. In the case of amperometry of glucose, optimal potential for the working electrode was -0.50V vs. SCE, and a satisfactory linearity was obtained in the tested concentration range from 65.8 to 260.6gmL(-1), with estimated LOQ of 35.0gmL(-1). The developed cost-effective amperometric method in combination with GOx/Pt-MWCNT-SPCE was successfully applied for the determination of glucose in selected food samples: white grapes and glucose tablets. The obtained results were in good agreement with those received by the alternatively used commercially available glucometer and by the producer declared value in the case of the tablet sample as well.
引用
收藏
页码:570 / 580
页数:11
相关论文
共 50 条
  • [1] Amperometric Determination of Glucose in White Grape and in Tablets as Ingredient by Screen-Printed Electrode Modified with Glucose Oxidase and Composite of Platinum and Multiwalled Carbon Nanotubes
    Valéria Guzsvány
    Jasmina Anojčić
    Olga Vajdle
    Emil Radulović
    Dániel Madarász
    Zoltán Kónya
    Kurt Kalcher
    Food Analytical Methods, 2019, 12 : 570 - 580
  • [2] Amperometric Determination of Glucose with FeO and Glucose Oxidase Bulk-Modified Screen-Printed Carbon Ink Biosensor
    Turkusic, E.
    Begic, S.
    Kahrovic, E.
    Kalcher, K.
    HEALTHMED, 2011, 5 (05): : 1117 - 1122
  • [3] Amperometric determination of glucose with an MnO2 and glucose oxidase bulk-modified screen-printed carbon ink biosensor
    Turkusic, E
    Kalcher, K
    Schachl, K
    Komersova, A
    Bartos, M
    Moderegger, H
    Svancara, I
    Vytras, K
    ANALYTICAL LETTERS, 2001, 34 (15) : 2633 - 2647
  • [4] Screen-printed biosensors for glucose determination in grape juice
    Gonzalo-Ruiz, Javier
    Alonso-Lomillo, M. Asuncion
    Munoz, F. Javier
    BIOSENSORS & BIOELECTRONICS, 2007, 22 (07): : 1517 - 1521
  • [5] Glucose oxidase immobilized amine terminated multiwall carbon nanotubes/reduced graphene oxide/polyaniline/gold nanoparticles modified screen-printed carbon electrode for highly sensitive amperometric glucose detection
    Maity, Debasis
    Minitha, C. R.
    Kumar, Rajendra R. T.
    MATERIALS SCIENCE AND ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2019, 105
  • [6] Amperometric determination of bonded glucose with an MnO2 and glucose oxidase bulk-modified screen-printed electrode using flow-injection analysis
    Turkusic, E
    Kalcher, J
    Kahrovic, E
    Beyene, NW
    Moderegger, H
    Sofic, E
    Begic, S
    Kalcher, K
    TALANTA, 2005, 65 (02) : 559 - 564
  • [7] Screen-printed Modified Diamond Electrode for Glucose Detection
    Kondo, Takeshi
    Horitani, Masaru
    Sakamoto, Hironori
    Shitanda, Isao
    Hoshi, Yoshinao
    Itagaki, Masayuki
    Yuasa, Makoto
    CHEMISTRY LETTERS, 2013, 42 (04) : 352 - 354
  • [8] Glucose Oxidase/Egg White Protein Microparticles with a Redox Mediator for Glucose Biosensors on a Screen-Printed Electrode and a Decomposable Electrode
    Rasitanon, Natcha
    Rattanapan, Parinthorn
    Kaewpradub, Kanyawee
    Buranachai, Chittanon
    Jeerapan, Itthipon
    BIOSENSORS-BASEL, 2023, 13 (08):
  • [9] A glucose oxidase sensor based on screen-printed carbon electrodes modified by polypyrrole
    Xu, Hui
    Li, Guang
    Wu, Jieying
    Wang, You
    Liu, Jun
    2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 1917 - 1920
  • [10] Composite glucose biosensor based on screen-printed electrodes bulk modified with Prussian blue and glucose oxidase
    Pravda, M
    O'Halloran, MP
    Kreuzer, MP
    Guilbault, GG
    ANALYTICAL LETTERS, 2002, 35 (06) : 959 - 970