A measurement of the fast-neutron sensitivity of a Geiger-Muller detector in the pulsed neutron beam from a superconducting cyclotron

被引:9
|
作者
Maughan, RL [1 ]
Yudelev, M [1 ]
Kota, C [1 ]
机构
[1] WAYNE STATE UNIV, DETROIT, MI 48230 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 1996年 / 41卷 / 08期
关键词
D O I
10.1088/0031-9155/41/8/007
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The k(U) value of a commercially available miniature energy compensated Geiger-Muller (GM) detector has been determined using the modified lead attenuation method of Hough. The measurements were made in a d(48.5)-Be neutron beam produced by the superconducting cyclotron based neutron therapy facility at Harper Hospital. The unique problems associated with making measurements in a 2 ms duration pulsed beam with a 20% duty cycle are discussed. The beam monitoring system, which allows the beam pulse shape at low beam intensities to be measured, is described. By gating the GM output with a discriminator pulse derived from the beam pulse shape, the gamma-ray count rates and dead-time corrections within the 2 ms pulse and between pulses can be measured separately. The k(U) value of (0.0245 +/- 0.0015) determined for this GM detector is consistent with the values measured by other workers with identical and similar detectors in neutron beams with comparable, but not identical, neutron spectra.