A measurement of the fast-neutron sensitivity of a Geiger-Muller detector in the pulsed neutron beam from a superconducting cyclotron

被引:9
|
作者
Maughan, RL [1 ]
Yudelev, M [1 ]
Kota, C [1 ]
机构
[1] WAYNE STATE UNIV, DETROIT, MI 48230 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 1996年 / 41卷 / 08期
关键词
D O I
10.1088/0031-9155/41/8/007
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The k(U) value of a commercially available miniature energy compensated Geiger-Muller (GM) detector has been determined using the modified lead attenuation method of Hough. The measurements were made in a d(48.5)-Be neutron beam produced by the superconducting cyclotron based neutron therapy facility at Harper Hospital. The unique problems associated with making measurements in a 2 ms duration pulsed beam with a 20% duty cycle are discussed. The beam monitoring system, which allows the beam pulse shape at low beam intensities to be measured, is described. By gating the GM output with a discriminator pulse derived from the beam pulse shape, the gamma-ray count rates and dead-time corrections within the 2 ms pulse and between pulses can be measured separately. The k(U) value of (0.0245 +/- 0.0015) determined for this GM detector is consistent with the values measured by other workers with identical and similar detectors in neutron beams with comparable, but not identical, neutron spectra.
引用
收藏
页码:1341 / 1351
页数:11
相关论文
共 48 条
  • [1] COMPARISON OF THE FAST-NEUTRON SENSITIVITY OF A GEIGER-MULLER COUNTER USING DIFFERENT TECHNIQUES
    MIJNHEER, BJ
    GULDBAKKE, S
    LEWIS, VE
    BROERSE, JJ
    PHYSICS IN MEDICINE AND BIOLOGY, 1982, 27 (01): : 91 - 96
  • [2] FAST-NEUTRON SENSITIVITY OF A GEIGER-MULLER COUNTER PHOTON DOSIMETER BY THE TIME-OF-FLIGHT TECHNIQUE
    KLEIN, H
    GULDBAKKE, S
    JAHR, R
    LESIECKI, H
    PHYSICS IN MEDICINE AND BIOLOGY, 1979, 24 (04): : 748 - 755
  • [3] COMPARISON OF THE FAST-NEUTRON SENSITIVITY OF A GEIGER-MULLER COUNTER USING DIFFERENT LEAD FILTRATION METHODS
    MIJNHEER, BJ
    VISSER, PA
    STRAHLENTHERAPIE, 1984, 160 (02) : 124 - 125
  • [4] THE NEUTRON SENSITIVITY OF A GEIGER-MULLER COUNTER BETWEEN 0.5 AND 8 MEV
    JONES, DTL
    MCMURRAY, WR
    HEALTH PHYSICS, 1983, 45 (03): : 659 - 663
  • [5] A FAST-NEUTRON ACTIVATION DETECTOR FOR PULSED SOURCES
    ROWLAND, MS
    ROBERTSON, JC
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1984, 224 (1-2): : 322 - 323
  • [6] DOSIMETRY OF FAST-NEUTRON BEAM PRODUCED BY MRC CYCLOTRON
    PARNELL, CJ
    BEWLEY, DK
    BRITISH JOURNAL OF RADIOLOGY, 1976, 49 (579): : 298 - 298
  • [7] FAST-NEUTRON BEAM RADIOTHERAPY INSTALLATION AT TAMVEC CYCLOTRON
    SMATHERS, JB
    OTTE, V
    ALMOND, PR
    SMITH, AR
    EUROPEAN JOURNAL OF CANCER, 1974, 10 (04) : 264 - 264
  • [8] Medium-type pulsed fast-neutron detector
    Ouyang, XP
    Li, ZF
    Huo, YK
    Zhang, ZB
    Shen, HL
    Zhang, QM
    HIGH ENERGY PHYSICS AND NUCLEAR PHYSICS-CHINESE EDITION, 2001, 25 (11): : 1100 - 1105
  • [9] Medium-type pulsed fast-neutron detector
    Ouyang, X.P.
    Li, Z.F.
    Huo, Y.K.
    Zhang, Z.B.
    Shen, H.L.
    Zhang, Q.M.
    Kao Neng Wu Li Yu Ho Wu Li/High Energy Physics and Nuclear Physics, 2001, 25 (11):
  • [10] A FAST-NEUTRON ACTIVATION DETECTOR FOR 14-MEV PULSED NEUTRON SOURCES
    RUBY, L
    RECHEN, JB
    NUCLEAR INSTRUMENTS & METHODS, 1962, 15 (01): : 74 - 76