Neodymium:YAG laser damage threshold of foldable intraocular lenses

被引:28
|
作者
Trinavarat, A [1 ]
Atchaneeyasakul, LO [1 ]
Udompunturak, S [1 ]
机构
[1] Mahidol Univ, Siriraj Hosp, Dept Ophthalmol, Bangkok 10700, Thailand
来源
关键词
D O I
10.1016/S0886-3350(00)00855-5
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To determine the energy level of the neodymium:YAG (Nd,YAG) laser that induced a 50% incidence of intraocular lens (IOL) damage in 5 foldable IOL materials. Setting: Department of Ophthalmology, Siriraj Hospital Mahidol University, Bangkok, Thailand. Methods: To simulate the condition within the capsular bag, an IOL was sandwiched between 2 rubber membranes. The front membrane had a central opening mimicking the anterior capsulotomy; the back membrane acted as the posterior capsule. The model was submerged in a water-filled chamber. The Nd:YAG laser with an automatic focus 150 mum beyond the focus of the aiming beam was used to perform a posterior capsulotomy. Three IOLs of each of 5 foldable IOL materials were tested. One type of poly(methyl methacrylate) (PMMA) was studied as a reference. The incidence of IOL damage at various energy levels was recorded. linear regression analysis was used to determine the 50% incidence damage threshold. Results: The 6 materials tested included 1 silicone lens (SI-40NB, AMO), 1 hydrophobic acrylic lens (MA60BM, Alcon), 3 hydrophilic acrylic lenses (Haptibag Ang, IOLtech; ACR6D, Corneal; H60M, Bausch & Lomb), acid 1 PMMA lens (LX10BD, Alcon). The 50% incidence damage threshold values were 0.37 mJ, 0.54 mJ, 0.58 mJ, 0.52 mJ, 0.66 mJ, and 0.68 mJ, respectively. Conclusions: The 50% incidence damage threshold in all the IOLs was below the energy level normally used to perform a posterior capsulotomy in clinical practice. However, setting the laser at the lowest possible energy, focusing the laser beam beyond the posterior capsule, and performing the capsulotomy early should minimize the risk of IOL damage.
引用
收藏
页码:775 / 780
页数:6
相关论文
共 50 条
  • [31] OPTICAL RESOLUTION OF FOLDABLE INTRAOCULAR LENSES
    KULNIG, W
    SKORPIK, C
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1990, 16 (02): : 211 - 216
  • [32] Spectroscopic characterization of foldable intraocular lenses
    Bertoluzza, A
    Monti, P
    Rossi, M
    Simoni, R
    SPECTROSCOPY OF BIOLOGICAL MOLECULES: NEW DIRECTIONS, 1999, : 521 - 522
  • [33] Transsderal fixation of foldable intraocular lenses
    Kaynak, S
    Ozbek, Z
    Pasa, E
    Oner, FH
    Cingil, G
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (04): : 854 - 857
  • [34] Foldable intraocular lenses and vitreoretinal surgery
    Obstbaum, SA
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1997, 23 (04): : 457 - 457
  • [35] SMALL INCISIONS AND FOLDABLE INTRAOCULAR LENSES
    SHEPHERD, JR
    INTERNATIONAL OPHTHALMOLOGY CLINICS, 1994, 34 (02) : 103 - 112
  • [36] Foldable acrylic intraocular lenses in children
    Smith, DS
    Stager, DR
    Stager, DR
    Wilson, ME
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1997, 38 (04) : 4517 - 4517
  • [37] Sealed implantation of foldable intraocular lenses
    Nath, Dharmendra
    Blumenthal, Eytan Z.
    Sharma, Mohita
    Mimouni, Michael
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2017, 27 (05) : 555 - 558
  • [38] Incision sizes for foldable intraocular lenses
    Kohnen, T
    Lambert, RJ
    Koch, DD
    OPHTHALMOLOGY, 1997, 104 (08) : 1277 - 1286
  • [39] Influence of neodymium:YAG laser capsulotomy on ocular wavefront aberrations in pseudophakic eyes with hydrophilic and hydrophobic intraocular lenses
    Rozema, Jos J.
    Koppen, Carina
    de Groot, Veva
    Tassignon, Marie-Jose
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (11): : 1906 - 1910
  • [40] Comparison of foldable intraocular lenses in uveitis
    Rauz, S
    Stavrou, P
    Murray, PI
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1999, 40 (04) : S151 - S151