Accuracy, precision, usability, and cost of free chlorine residual testing methods

被引:27
|
作者
Murray, Anna [1 ]
Lantagne, Daniele [1 ]
机构
[1] Tufts Univ, Dept Civil & Environm Engn, Medford, MA 02155 USA
基金
美国国家科学基金会;
关键词
chlorination; developing countries; drinking water; field testing methods; free chlorine residual; household water treatment;
D O I
10.2166/wh.2014.195
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Chlorine is the most widely used disinfectant worldwide, partially because residual protection is maintained after treatment. This residual is measured using colorimetric test kits varying in accuracy, precision, training required, and cost. Seven commercially available colorimeters, color wheel and test tube comparator kits, pool test kits, and test strips were evaluated for use in low-resource settings by: (1) measuring in quintuplicate 11 samples from 0.0-4.0 mg/L free chlorine residual in laboratory and natural light settings to determine accuracy and precision; (2) conducting volunteer testing where participants used and evaluated each test kit; and (3) comparing costs. Laboratory accuracy ranged from 5.1-40.5% measurement error, with colorimeters the most accurate and test strip methods the least. Variation between laboratory and natural light readings occurred with one test strip method. Volunteer participants found test strip methods easiest and color wheel methods most difficult, and were most confident in the colorimeter and least confident in test strip methods. Costs range from 3.50-444 USD for 100 tests. Application of a decision matrix found colorimeters and test tube comparator kits were most appropriate for use in low-resource settings; it is recommended users apply the decision matrix themselves, as the appropriate kit might vary by context.
引用
收藏
页码:79 / 90
页数:12
相关论文
共 50 条
  • [41] Accuracy and precision of the MediSense Precision G blood glucose testing system.
    Zamoth, D
    Cembrowski, G
    McNeil, L
    Fineberg, SE
    Gibson, R
    CLINICAL CHEMISTRY, 1996, 42 (06) : 250 - 250
  • [42] Tolerance interval testing for assessing accuracy and precision simultaneously
    Chiang, Chieh
    Hsiao, Chin-Fu
    PLOS ONE, 2021, 16 (02):
  • [43] ACCURACY NORMALIZATION IN STANDARDS ON TESTING METHODS
    VESELKOVA, TA
    MEASUREMENT TECHNIQUES, 1974, 17 (08) : 1159 - 1161
  • [44] Testing accuracy of analytical methods by regression
    Kemeny, S.
    Deak, A.
    Banfai, B.
    JOURNAL OF CHEMOMETRICS, 2009, 23 (3-4) : 211 - 216
  • [45] Not just usability testing: Remembering and applying non-usability testing methods for learning how Web sites function
    McGovern, H
    TECHNICAL COMMUNICATION, 2005, 52 (02) : 175 - 186
  • [46] Precision and accuracy of staged free-energy perturbation methods for computing the chemical potential by molecular simulation
    Kofke, DA
    Cummings, PT
    FLUID PHASE EQUILIBRIA, 1998, 150 : 41 - 49
  • [47] BENCH TESTING OF ON-LINE TOTAL RESIDUAL CHLORINE ANALYZERS.
    Nutt, S.G.
    Jeffers, E.L.
    Vachon, D.T.
    Schuk, W.W.
    Journal of the Water Pollution Control Federation, 1985, 57 (09): : 948 - 954
  • [48] Temperature impact on monochloramine, free ammonia, and free chlorine indophenol methods
    Waters, Thomas E.
    Alexander, Matthew T.
    Wahman, David G.
    WATER PRACTICE AND TECHNOLOGY, 2021, 16 (01): : 135 - 145
  • [49] TESTING USABILITY METHODS IN TRANSLATION COURSES: PERSONAS AND HEURISTIC EVALUATION
    Suokas, Juho
    CURRENT TRENDS IN TRANSLATION TEACHING AND LEARNING E, 2020, 7 : 4 - 38
  • [50] Methods of usability testing in the development of eHealth applications: A scoping review
    Maramba, Inocencio
    Chatterjee, Arunangsu
    Newman, Craig
    INTERNATIONAL JOURNAL OF MEDICAL INFORMATICS, 2019, 126 : 95 - 104