Membrane based on decyl-18-crown-6 for a potassium selective sensor

被引:0
|
作者
Hopârtean, E
Pica, EM
Ana, C
Cosma, V
Hopârtean, I
机构
[1] Tech Univ Cluj Napoca, R-3400 Cluj Napoca, Romania
[2] Inst Chem, R-3400 Cluj Napoca, Romania
来源
CHEMIA ANALITYCZNA | 2001年 / 46卷 / 01期
关键词
crown ethers; potassium PVC membrane sensors; potassium potentiometric determination; ion-selective electrode; ionophores;
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Four kinds of 18-crown-6 derivatives were studied as ionophores in plasticized poly(vinyl chloride) (PVC) in order to obtain potassium selective membrane. The properties of the sensor were studied in detail. For the optimization of the selectivity toward K+ over Na+ the influence of the following factors was studied: nature and concentration of the ionophore; nature and concentration of the plasticizer and lipophilic salt addition. The potassium ion-selective membrane electrode was designed. The electrode has a fast and Nernstian response in the 1 x 10(-1)-2 x 10(-5) mol l(-1) K+, after six months of continuous measurements. The change of pH in the range 2-10.5 does not affect the response of the electrode. The internal resistance of the electrode was measured. The electrode shows a good stability and can be used in the potentiometric determination.
引用
收藏
页码:41 / 49
页数:9
相关论文
共 50 条
  • [11] Fabrication of Potassium Ion-selective Membrane Electrode as Immobilized Liquid based on Dibenzo-18-Crown-6 (DB18C6) as Ionophore
    Taheri, Khalil
    Afsari, Tahereh
    ANALYTICAL & BIOANALYTICAL ELECTROCHEMISTRY, 2014, 6 (04): : 475 - 484
  • [12] Highly Selective and Efficient Membrane Transport of Silver as AgBr-2 Ion Using K+-decyl-18-crown-6 as Carrier
    Javad Zolgharnein
    Sheida Hosseini
    Golamhassan Azimi
    Mohammad Reza Sangi
    Analytical Sciences, 2003, 19 : 871 - 875
  • [13] Highly selective and efficient membrane transport of silver as AgBr2- ion using K+-decyl-18-crown-6 as carrier
    Zolgharnein, J
    Hosseini, S
    Azimi, G
    Sangi, MR
    ANALYTICAL SCIENCES, 2003, 19 (06) : 871 - 875
  • [14] Nickel(II)-selective sensor based on dibenzo-18-crown-6 in PVC matrix
    Gupta, Vinod Kumar
    Goyal, Rajendra N.
    Agarwal, Shiva
    Kumar, Pankaj
    Bachheti, Neeta
    TALANTA, 2007, 71 (02) : 795 - 800
  • [15] PVC based monoaza-18-crown-6 membrane potentiometric sensor for the monitoring of cadmium
    Gupta, VK
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1999, 217 : U118 - U118
  • [16] Sodium ion-selective membrane electrode based on dibenzopyridino-18-crown-6
    Tavakkoli, N
    BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2004, 25 (10) : 1474 - 1476
  • [17] Thallium (I) ion Selective Membrane Electrode based on Dibenzopyridino-18-crown-6
    Zolgharnein, Javed
    Azimi, Ghulam
    Habibi, Morteza
    JOURNAL OF THE CHEMICAL SOCIETY OF PAKISTAN, 2007, 29 (05): : 487 - 491
  • [18] Polyindole modified potassium ion sensor using dibenzo-18-crown-6 mediated PVC matrix membrane
    Pandey, PC
    Prakash, R
    SENSORS AND ACTUATORS B-CHEMICAL, 1998, 46 (01) : 61 - 65
  • [19] Dicyclohexano-18-crown-6 as active material in PVC matrix membrane for the fabrication of cadmium selective potentiometric sensor
    Gupta, VK
    Chandra, S
    Mangla, R
    ELECTROCHIMICA ACTA, 2002, 47 (10) : 1579 - 1586
  • [20] (18-Crown-6)potassium chlorochromate
    Kotlyar, SA
    Zubatyuk, RI
    Shishkin, OV
    Chuprin, GN
    Kiriyak, AV
    Kamalov, GL
    ACTA CRYSTALLOGRAPHICA SECTION E-CRYSTALLOGRAPHIC COMMUNICATIONS, 2005, 61 : M293 - M295