Analytical performance evaluation of the i-STAT Total β-human chorionic gonadotropin immunoassay

被引:14
|
作者
Sowder, Aleksandra M. [1 ]
Yarbrough, Melanie L. [2 ]
Nerenz, Robert D. [2 ]
Mitsios, John V. [2 ]
Mortensen, Rachel [3 ]
Gronowski, Ann M. [2 ,4 ]
Grenache, David G. [1 ,5 ]
机构
[1] Univ Utah, Sch Med, Dept Pathol, Salt Lake City, UT USA
[2] Washington Univ, Sch Med, Dept Pathol & Immunol, St Louis, MO USA
[3] ARUP Labs, Salt Lake City, UT 84108 USA
[4] Washington Univ, Sch Med, Dept Obstet & Gynecol, St Louis, MO 63110 USA
[5] ARUP Inst Clin & Expt Pathol, Salt Lake City, UT USA
关键词
Human chorionic gonadotropin; Quantitative; Whole blood; Point-of-care; Pregnancy test; POINT-OF-CARE; QUALITATIVE POINT; HCG DEVICES; URINE; PREGNANCY;
D O I
10.1016/j.cca.2015.04.025
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The ability to perform quantitative hCG testing in whole blood at the point-of-care is desirable. The purpose of this study was to perform an analytical validation of the Abbott i-STAT Total beta-hCG test. Methods: Whole blood, plasma, and serum samples were prepared by the addition of hCG and were used to evaluate precision, linearity, analytical sensitivity, accuracy, the high-dose hook effect, and dilution recovery. Results: Imprecision was highest with whole blood (CV = 16.0% and 6.7% at 10 and 1184 IU/l, respectively) and lowest in serum (CV = 8.1% and 4.3% at 11 and 1305 IU/l, respectively). The limits-of-quantitation were 8 and <5 IU/l for whole blood and both plasma and serum, respectively. The assay was linear between 5 and 2000 IU/l in all sample types (R-2 >= 0.998). i-STAT results agreed most closely with the Architect Total beta-hCG assay and with greater differences observed with Beckman Dxl Total beta hCG and Roche Cobas e601 hCG + beta assays (mean differences across all sample types were 9.3% and 12.3%, respectively). A high-dose hook effect was observed at concentrations >400,000 IU/l. Accuracy was achieved in samples diluted with serum but not saline. Conclusions: The i-STAT Total beta-hCG test demonstrates acceptable performance for quantifying hCG in whole blood, plasma and serum. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:165 / 170
页数:6
相关论文
共 50 条
  • [41] Analytical Performances of the Novel i-STAT Alinity Point-of-Care Analyzer
    Larcher, Romaric
    Lottelier, Maxence
    Badiou, Stephanie
    Dupuy, Anne-Marie
    Bargnoux, Anne-Sophie
    Cristol, Jean-Paul
    DIAGNOSTICS, 2023, 13 (02)
  • [42] Immunoassay sf human chorionic gonadotropin, its free subunits, and metabolites
    Cole, LA
    CLINICAL CHEMISTRY, 1997, 43 (12) : 2233 - 2243
  • [43] A SENSITIVE ENZYME-IMMUNOASSAY SPECIFIC FOR HUMAN CHORIONIC-GONADOTROPIN
    MEHTA, HC
    MACDONALD, DJ
    CLINICA CHIMICA ACTA, 1982, 121 (02) : 245 - 250
  • [44] DEVELOPMENT OF AN ENZYME-LINKED IMMUNOASSAY FOR HUMAN CHORIONIC-GONADOTROPIN
    JOSHI, UM
    RAGHAVAN, V
    SHETH, AR
    INDIAN JOURNAL OF MEDICAL RESEARCH, 1977, 65 (06) : 807 - 811
  • [45] Quantitative human chorionic gonadotropin measurement in urine using the access immunoassay
    Er, Tze-Kiong
    Tsai, Li-Yu
    Gines, Miguel Angel Ruiz
    Chen, Bai-Hsiun
    Chang, Jan-Gowth
    Jong, Yuh-Jyh
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2008, 26 (01): : 103 - 104
  • [46] HIGHLY SPECIFIC ENZYME-IMMUNOASSAY FOR HUMAN CHORIONIC-GONADOTROPIN
    IWASA, S
    KITADA, C
    YOSHIDA, I
    KONDO, K
    HORI, M
    FUJINO, M
    JOURNAL OF BIOCHEMISTRY, 1981, 89 (04): : 1091 - 1099
  • [47] OPTIMIZATION OF A SANDWICH SOL PARTICLE IMMUNOASSAY FOR HUMAN CHORIONIC-GONADOTROPIN
    LEUVERING, JHW
    THAL, PJHM
    SCHUURS, AHWM
    JOURNAL OF IMMUNOLOGICAL METHODS, 1983, 62 (02) : 175 - 184
  • [48] Human Chorionic Gonadotropin: Do You Know the Specificity of Your Immunoassay?
    Whittington, JoDell
    Wilson, Andrew
    Grenache, David
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2009, 132 (03) : 464 - 464
  • [49] QUALITATIVE ENZYME-IMMUNOASSAY FOR HUMAN CHORIONIC-GONADOTROPIN IN SERUM
    DANISCH, RJ
    HAYKAUFMAN, ML
    WADA, HG
    CLINICAL CHEMISTRY, 1983, 29 (06) : 1213 - 1214
  • [50] Analytical evaluation of i-STAT portable clinical analyzer for blood gas and electrolyte analyses in pediatric samples.
    Murthy, JN
    Hicks, JM
    Soldin, SJ
    CLINICAL CHEMISTRY, 1996, 42 (06) : 198 - 198