Analytical performance evaluation of the i-STAT Total β-human chorionic gonadotropin immunoassay

被引:14
|
作者
Sowder, Aleksandra M. [1 ]
Yarbrough, Melanie L. [2 ]
Nerenz, Robert D. [2 ]
Mitsios, John V. [2 ]
Mortensen, Rachel [3 ]
Gronowski, Ann M. [2 ,4 ]
Grenache, David G. [1 ,5 ]
机构
[1] Univ Utah, Sch Med, Dept Pathol, Salt Lake City, UT USA
[2] Washington Univ, Sch Med, Dept Pathol & Immunol, St Louis, MO USA
[3] ARUP Labs, Salt Lake City, UT 84108 USA
[4] Washington Univ, Sch Med, Dept Obstet & Gynecol, St Louis, MO 63110 USA
[5] ARUP Inst Clin & Expt Pathol, Salt Lake City, UT USA
关键词
Human chorionic gonadotropin; Quantitative; Whole blood; Point-of-care; Pregnancy test; POINT-OF-CARE; QUALITATIVE POINT; HCG DEVICES; URINE; PREGNANCY;
D O I
10.1016/j.cca.2015.04.025
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The ability to perform quantitative hCG testing in whole blood at the point-of-care is desirable. The purpose of this study was to perform an analytical validation of the Abbott i-STAT Total beta-hCG test. Methods: Whole blood, plasma, and serum samples were prepared by the addition of hCG and were used to evaluate precision, linearity, analytical sensitivity, accuracy, the high-dose hook effect, and dilution recovery. Results: Imprecision was highest with whole blood (CV = 16.0% and 6.7% at 10 and 1184 IU/l, respectively) and lowest in serum (CV = 8.1% and 4.3% at 11 and 1305 IU/l, respectively). The limits-of-quantitation were 8 and <5 IU/l for whole blood and both plasma and serum, respectively. The assay was linear between 5 and 2000 IU/l in all sample types (R-2 >= 0.998). i-STAT results agreed most closely with the Architect Total beta-hCG assay and with greater differences observed with Beckman Dxl Total beta hCG and Roche Cobas e601 hCG + beta assays (mean differences across all sample types were 9.3% and 12.3%, respectively). A high-dose hook effect was observed at concentrations >400,000 IU/l. Accuracy was achieved in samples diluted with serum but not saline. Conclusions: The i-STAT Total beta-hCG test demonstrates acceptable performance for quantifying hCG in whole blood, plasma and serum. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:165 / 170
页数:6
相关论文
共 50 条
  • [22] IMMUNOASSAY OF HUMAN CHORIONIC-GONADOTROPIN USING FLUORESCENCE POLARIZATION
    URIOS, P
    CITTANOVA, N
    JAYLE, MF
    FEBS LETTERS, 1978, 94 (01) : 54 - 58
  • [23] SPECIFIC ENZYME-IMMUNOASSAY FOR HUMAN CHORIONIC-GONADOTROPIN
    SEKIYA, T
    FURUHASHI, Y
    GOTO, S
    KASEKI, S
    TOMODA, Y
    KATO, K
    ACTA ENDOCRINOLOGICA, 1981, 97 (04): : 562 - 568
  • [24] SOL PARTICLE AGGLUTINATION IMMUNOASSAY FOR HUMAN CHORIONIC-GONADOTROPIN
    LEUVERING, JHW
    THAL, PJHM
    WAART, MVD
    SCHUURS, AHWM
    FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1980, 301 (02): : 132 - 132
  • [25] Elimination of Complement Interference in Immunoassay of Hyperglycosylated Human Chorionic Gonadotropin
    Stenman, Ulf-Hakan
    Birken, Steven
    Lempiainen, Anna
    Hotakainen, Kristina
    Alfthan, Henrik
    CLINICAL CHEMISTRY, 2011, 57 (07) : 1075 - 1077
  • [26] STRUCTURE, IMMUNOCHEMICAL PROPERTIES AND IMMUNOASSAY OF HUMAN CHORIONIC-GONADOTROPIN
    ALBERTINI, A
    GHIELMI, S
    BELLOLI, S
    RICERCA IN CLINICA E IN LABORATORIO, 1982, 12 (01): : 289 - 298
  • [27] Method evaluation for total β-human chorionic gonadotropin using urine and the ADVIA centaur
    Halldorsdottir, AM
    Carayannopoulos, MO
    Scrivner, M
    Gronowski, AM
    CLINICAL CHEMISTRY, 2003, 49 (08) : 1421 - 1422
  • [28] AN EVALUATION OF THE ABBOTT IMX TOTAL BETA HUMAN CHORIONIC-GONADOTROPIN ASSAY
    KRISHNAN, S
    WEBB, S
    CLINICAL CHEMISTRY, 1990, 36 (06) : 1184 - 1184
  • [29] EVALUATION OF A NEW CHEMI-LUMINESCENCE IMMUNOASSAY FOR THE DETERMINATION OF HUMAN CHORIONIC-GONADOTROPIN IN SERUM
    LUPPA, P
    NEUMEIER, D
    KNEDEL, M
    JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1988, 26 (11): : 705 - 713
  • [30] Two-center clinical evaluation of a new automated fluorometric immunoassay for the quantitative analysis of total βeta-human chorionic gonadotropin
    Azzazy, HME
    Romero, LF
    Hall, L
    Ruttle, D
    Christenson, RH
    CLINICAL BIOCHEMISTRY, 2003, 36 (07) : 523 - 528