Effects of bileaflet mechanical heart valve orientation on fluid stresses and coronary flow

被引:16
|
作者
Haya, Laura [1 ]
Tavoularis, Stavros [1 ]
机构
[1] Univ Ottawa, Dept Mech Engn, Ottawa, ON K1N 6N5, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
biological fluid dynamics; blood flow; PARTICLE IMAGE VELOCIMETRY; FREE EMBOLI FORMATION; AORTIC-VALVE; SHEAR-STRESS; ARTERY FLOW; DOWNSTREAM; TURBULENCE; FIELD; PIV;
D O I
10.1017/jfm.2016.582
中图分类号
O3 [力学];
学科分类号
08 ; 0801 ;
摘要
The effects of the orientation of a bileaflet mechanical heart valve on the viscous and turbulent stresses in the flow past it and on the flow rate in the right coronary artery were investigated in vitro in a mock circulation loop, using a fluid that matched the kinematic viscosity of blood and the refractive index of the aorta model. Measurements were made past the valve mounted in three orientations at the base of an anatomical aorta model, within physiological aortic flow conditions. At peak flow, the turbulent stresses were on average 21% higher and viscous stresses exceeding 10 Pa (namely of a level that has been associated with blood cell damage) were 30% more frequent when the valve was oriented with its plane of symmetry normal to the aorta's plane of curvature than when it was parallel to it. This was attributed to the impingement of a lateral jet on the concave wall of the aorta and to steeper velocity gradients resulting from the geometrical imbalance of the sinuses relative to the valve's central jet when the valve was in the 'normal' orientation. Very high levels of turbulent stresses were found to occur distal to the corners of the valve's lateral orifices. The bulk flow rate in the right coronary artery was highest when the valve was positioned with its central orifice aligned with the artery's opening. The coronary flow rate was directly affected by the size, orientation and time evolution of the vortex in the sinus, all of which were sensitive to the valve's orientation.
引用
收藏
页码:129 / 164
页数:36
相关论文
共 50 条
  • [41] Fluid dynamics of the CarboMedics Kinetic bileaflet prosthetic heart valve
    Baldwin, JT
    Campbell, A
    Luck, C
    Ogilvie, W
    Sauter, J
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1997, 11 (02) : 287 - 292
  • [42] A STUDY OF THE STABILITY OF ORIENTATION IN THE CARBOMEDICS BILEAFLET HEART-VALVE PROSTHESIS
    CHAMBERS, JB
    DEVERALL, PB
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1992, 6 (08) : 442 - 445
  • [43] Numerical simulation of opening process in a bileaflet mechanical heart valve under pulsatile flow condition
    Shi, YB
    Zhao, Y
    Yeo, TJH
    Hwang, NHC
    JOURNAL OF HEART VALVE DISEASE, 2003, 12 (02): : 245 - 255
  • [44] CFD simulation of a novel bileaflet mechanical heart valve prosthesis
    不详
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2005, 28 (09): : 868 - 868
  • [45] Hydrodynamics of a long-body bileaflet mechanical heart valve
    Gao, BZ
    Hosein, N
    Hwang, NHC
    ASAIO JOURNAL, 1997, 43 (05) : M396 - M401
  • [46] Durability testing of a polymeric bileaflet mechanical heart valve prosthesis
    不详
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2005, 28 (09): : 867 - 867
  • [47] An unexplained cluster of thrombosed bileaflet mechanical heart valve prostheses
    Chaudhuri, N
    Hickey, MS
    Spyt, TJ
    ANNALS OF THORACIC SURGERY, 1999, 67 (05): : 1542 - 1543
  • [48] Computational hemodynamic investigation of a new bileaflet mechanical heart valve
    Khellaf, Belkhiri
    Boussad, Boumeddane
    SIMULATION-TRANSACTIONS OF THE SOCIETY FOR MODELING AND SIMULATION INTERNATIONAL, 2020, 96 (05): : 459 - 469
  • [49] Vorticity dynamics of a bileaflet mechanical heart valve in an axisymmetric aorta
    Dasi, L. P.
    Ge, L.
    Simon, H. A.
    Sotiropoulos, F.
    Yoganathan, A. P.
    PHYSICS OF FLUIDS, 2007, 19 (06)