Differences in effective dose estimation from dose-area product and entrance surface dose measurements in intravenous urography

被引:48
|
作者
Yakoumakis, E
Tsalafoutas, IA
Nikolaou, D
Nazos, I
Koulentianos, E
Proukakis, C
机构
[1] Univ Athens, Sch Med, Dept Med Phys, GR-11527 Athens, Greece
[2] Konstantopoulio Agia Olga Hosp, Dept Radiol, Athens 14233, Greece
[3] Konstantopoulio Agia Olga Hosp, Med Phys Unit, Athens 14233, Greece
来源
BRITISH JOURNAL OF RADIOLOGY | 2001年 / 74卷 / 884期
关键词
D O I
10.1259/bjr.74.884.740727
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this study, measurements of dose-area product (DAP) and entrance surface dose (ESD) were carried out in a sample of 25 adult patients who underwent intravenous urography (IVU). These measured quantities were used to estimate the effective dose E from the IVU examination, a quantity closely correlated to radiation risk. Estimating E involves the use of conversion coefficients that have been determined for specific X-ray views in a mathematical phantom. These are obtained under conditions which are not usually met in clinical practice. As a result, the E estimates using the two different measurable quantities can be quite different. Analysis of the calculation procedure suggests that the E estimate using the DAP measurements, in addition to being more practical, could be more accurate than using ESD measurements, as DAP is sensitive to the X-ray field size settings. Furthermore, it is shown that in the absence of the appropriate equipment, a reliable E estimate can be obtained from the ESD calculated using the exposure data for each X-ray view.
引用
收藏
页码:727 / 734
页数:8
相关论文
共 50 条
  • [31] Dose-area product in pediatric barium meal procedures
    Machado, S. O.
    Schelin, H.
    Denyak, V.
    Filipov, D.
    Bunick, A.
    Ledesma, J.
    Paschuk, S. A.
    RADIATION PHYSICS AND CHEMISTRY, 2019, 155 : 53 - 55
  • [32] Quantifying dose-creep for Skull and chest radiography using dose area product and entrance surface dose: Phantom study
    Dalah, Entesar Zawam
    RADIATION PHYSICS AND CHEMISTRY, 2020, 167 (167)
  • [33] No reduction of radiation dose following the introduction of dose-area product measurement in endoscopic retrograde cholangiopancreatography
    Kruit, Anne Sophie
    Vleggaar, Frank P.
    van Erpecum, Karel J.
    Timmerman, Annemoon M. D. E.
    Siersema, Peter D.
    Oldenburg, Bas
    EUROPEAN JOURNAL OF GASTROENTEROLOGY & HEPATOLOGY, 2015, 27 (12) : 1454 - 1458
  • [34] AN EVALUATION OF A DIAMENTOR BASED SYSTEM FOR ESTIMATION OF SPOT FILM DOSE-AREA PRODUCT VALUES
    HOSKINS, PR
    ADAM, RD
    BRITISH JOURNAL OF RADIOLOGY, 1995, 68 (814): : 1106 - 1111
  • [35] Computer assisted determination of the dose-area product in retrograde urethrography
    Merkle, E
    Aschoff, AJ
    Muller, M
    Merk, J
    Brambs, HJ
    BRITISH JOURNAL OF RADIOLOGY, 1996, 69 (819): : 262 - 265
  • [36] Dose-Area Product as a Method for Small Field Geometric QA
    Nikolic, M.
    Masterson-McGary, M.
    Toner, S.
    Kilby, W.
    Thomson, L.
    Colliander, S.
    Noll, M.
    Goggin, L.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [37] Effective dose estimation during conventional and CT urography
    Alzimami, K.
    Sulieman, A.
    Omer, E.
    Suliman, I. I.
    Alsafi, K.
    RADIATION PHYSICS AND CHEMISTRY, 2014, 104 : 154 - 157
  • [38] Estimate Eye Lens Dose by Using Personal Dose Equivalent (HP (10)) and Dose-Area Product (DAP) in Interventional Procedures
    Hu, P.
    Wang, J.
    Hu, W.
    Liu, H.
    MEDICAL PHYSICS, 2018, 45 (06) : E276 - E276
  • [39] Accuracy of a dose-area product compared to an absorbed dose to water at a point in a 2 cm diameter field
    Dufreneix, S.
    Ostrowsky, A.
    Rapp, B.
    Daures, J.
    Bordy, J. M.
    MEDICAL PHYSICS, 2016, 43 (07) : 4085 - 4092
  • [40] A patient dose survey for femoral arteriogram diagnostic radiographic examinations using a dose-area product meter
    Thwaites, JH
    Rafferty, MW
    Gray, N
    Black, J
    Stock, B
    PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (05): : 899 - 907